Literature DB >> 30775952

Transient Heat Transfer in Dental Implants for Thermal Necrosis-Aided Implant Removal: A 3D Finite Element Analysis.

Mustafa Gungormus1,2, Guzin Neda Hasanoglu Erbasar3.   

Abstract

Removal of osseointegrated but otherwise failed (mechanical failure, mispositioning, esthetics, etc) dental implants is a traumatic process resulting in loss of healthy bone and complicating the treatment process. The traumatic effects of implant removal can be reduced by weakening the implant-bone attachment. Thermal necrosis-aided implant removal has been proposed as a minimally invasive method toward this end. In this method, an electrocautery tip is contacted to the implant to increase the temperature to 47°C and generate a limited and controlled thermal necrosis at the bone-implant interface. So far, no controlled studies have been performed to investigate the optimal clinical parameters for this method. In this study, we aimed to investigate, using finite element analysis method, the optimal settings to achieve intentional thermal necrosis on 3 implant systems, at 5 W and 40 W device power and with different size tips. The temperature increase of the implants at 40 W power was very sudden (< 0.5 seconds) and as the bone reached 47°C, the implants were at unacceptable temperatures. At 5 W power, temperature increase of the implants happened at manageable durations (< 1 second). Moreover, the temperature increase was even slower with larger implants and larger tip sizes. Therefore, low power settings must be used for thermal necrosis-aided implant removal. Also, the size of the implant and the tip must be taken into consideration in deciding the duration of contact with the electrocautery tip and the implant.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30775952     DOI: 10.1563/aaid-joi-D-18-00210

Source DB:  PubMed          Journal:  J Oral Implantol        ISSN: 0160-6972            Impact factor:   1.779


  4 in total

1.  Non-Destructive Removal of Dental Implant by Using the Cryogenic Method.

Authors:  Burak Ak; Emre Gürkan Eroğlu; Abdullah Seckin Ertugrul; Ayla Batu Öztürk; Şakir Necat Yılmaz
Journal:  Medicina (Kaunas)       Date:  2022-06-25       Impact factor: 2.948

2.  Postsurgical Pain and Implant Osseointegration Failure: A Case Control Study.

Authors:  Yuli Shang; Qiuying Gao; Tina Lengas; Shu Deng
Journal:  Int J Dent       Date:  2022-07-07

3.  Evaluation of Implant Stability and Trephination Depth for Implant Removal-An In Vitro Study.

Authors:  Haya Meir; Alon Sebaoun; Perry Raz; Shifra Levartovsky; Adi Arieli; Raphel Pilo; Zafar Dor; Ilan Beitlitum
Journal:  Materials (Basel)       Date:  2022-06-13       Impact factor: 3.748

Review 4.  Reversal of Osseointegration as a Novel Perspective for the Removal of Failed Dental Implants: A Review of Five Patented Methods.

Authors:  Rolf G Winnen; Kristian Kniha; Ali Modabber; Faruk Al-Sibai; Andreas Braun; Reinhold Kneer; Frank Hölzle
Journal:  Materials (Basel)       Date:  2021-12-17       Impact factor: 3.623

  4 in total