RADIATION EXPOSURE DURING SPINE SURGERY USING C-ARM FLUOROSCOPY.

Alexandre Fogaça Cristante1, Fábio Barbieri2, Almy Anacleto Rodrigues da Silva2,3, José Claudio Dellamano4,5.   

Abstract

OBJECTIVE: To evaluate the radiation dose received by staff in spine surgeries, including those who are not considered occupationally exposed workers.
METHODS: All spinal surgeries performed in the same department during a period of 12 months were evaluated with regard to the exposure of surgeons, scrub nurses, and auxiliary personnel working in the operating room to radiation from C-arm fluoroscopy. Radiation was measured by 15 film badge dosimeters placed on the professionals' lapels, gloves, and room standardized sites. The films were analyzed in the dosimetry laboratory by collections per period.
RESULTS: During the 12 months, 81 spinal surgeries were performed by the same team, with surgical times ranging from 1 to 6 hours. The total radiation dose ranged from 0.16 mSv to 2.29 mSv depending on the dosimetry site. The most exposed site was the wrist of the main surgeon.
CONCLUSION: The results showed that in the spinal surgeries in our setting, the radiation doses are low and within legal limits. Nevertheless, constant training of professionals is essential for radiation protection of medical staff and patients. Level of evidence I/b, exploratory cohort study.

Entities:  

Keywords:  Absorption, Radiation; Radiation; Radiation Dosage; Radiation Exposure; Radiation, Ionizing; Radiography

Year:  2019        PMID: 30774530      PMCID: PMC6362693          DOI: 10.1590/1413-785220192701172722

Source DB:  PubMed          Journal:  Acta Ortop Bras        ISSN: 1413-7852            Impact factor:   0.513


INTRODUCTION

Since Wilhelm Röntgen discovered x-rays in 1895, medicine has been using its benefits for diagnoses and therapy. In recent decades, ionizing radiation has been used more frequently, with the development of new diagnostic and therapeutic techniques. High quality images that allow quick and accurate diagnoses are considered indispensable nowadays in the clinical setting. The use of fluoroscopically-guided techniques in surgery has been increasing, and images in real time are now a vital tool for minimally invasive procedures. The International Commission on Radiological Protection states: “Interventional radiology offers to medicine in all countries, no matter the stage of development, the opportunity to treat a greater range of pathologies, in more patients and at lesser cost. Interventional techniques reduce the need for expensive operating suites and extended hospital in-patient admissions. They also reduce most of the risks to the patient by the use of minimally invasive techniques and through lesser requirements for general anesthesia”. As a consequence of the increasing use of imaging in the surgical setting, exposure times, and the radiation doses involved with the use of fluoroscopy, have been increasing too. It is known that frequent exposure to ionizing radiation can cause serious health effects. Therefore, the risk to which individuals are exposed during tasks involving ionizing radiation must be carefully considered by professionals who work with radiology, radiotherapy and nuclear medicine services. - In relation to the maximum amount of ionizing radiation to be used, there are principles that must be adhered to, in order to protect the patient. However, this dosage limitation does not apply to the exposure of medical staff to radiation. As a result, this exposure can be high in medical diagnosis, when compared with other medical procedures that use ionizing radiation. In fact, imaging exams are now the procedure that most contributes to radiation exposure. But there is a dose-response relationship between exposure to low doses of radiation and mortality, with significant increases in risk observed with doses of 100 to 200 mGy or higher. Ionizing radiation is used in a large number of procedures carried out in surgery centers. One such procedure is orthopedic surgery using the C-arm fluoroscope. The C-arm is essential in spinal surgeries, to guide the correct execution of the procedure. Despite the use of modern equipment with automatic control systems to limit exposure to radiation, which automatically adjust the voltage (kV), electrical current and exposure time (mAs) to compensate for variations in thickness and density of the tissues being observed, the resulting radiation dose is significant for the patient, and also for the staff involved in the surgical procedures. , Some studies have evaluated radiation doses in medical diagnoses and surgical centers, whether for patient, for occupationally exposed individuals (OIEs), or for the whole staff involved in the procedure. The majority of these studies either simulate surgical procedures or carry out the measurements during orthopedic or other procedures, in real time. - However, there are no studies in Brazil focused on surgeons’ exposure to ionizing radiation during surgical operations to the spine.

MATERIALS AND METHODS

This experimental study was conducted in the Spine Surgery Department of a large, public, university hospital in Brazil. The protocol was approved by the local ethics committee and all the participants (surgeons, technicians, nurses and auxiliary staff) involved signed informed consent forms for the radiation dose evaluations and inclusion in the study. All the spinal surgeries performed in the department over a twelve-month period were evaluated. The same fluoroscopy equipment was used in all the procedures. Radiation film badge dosimeters were attached to fifteen standard sites, and kept in operation throughout the surgery: Standard Surgical center 1 – operating room wall Surgical center 2 – opposite wall of the operating room Below the operating room table Surgeon 1 - attached to the lapel, over the lead apron Surgeon 1 - attached to the lapel, under the lead apron Surgeon 1 - wrist Surgeon 2 - attached to the lapel, over the lead apron Surgeon 2 - attached to the lapel, under the lead apron Surgeon 2 - wrist Instrumentator (scrub nurse) - attached to the lapel Resident - attached to the lapel C-arm operator - attached to the lapel, over the lead apron C-arm operator - attached to the lapel, under the lead apron Auxiliary - in the c arm The C-arm operator and the surgeons 1 (main surgeon) and 2 (auxiliary surgeon) used two badges attached to the lapel, one over the lead cover and another underneath it, and one attached to the wrist, under the surgical glove. Care was taken to leave the movements of the hands and arms free. Figure 1 shows how the film badge dosimeters were attached to the staff members, and Figure 2 shows the operating room table and the attachment of the dosimeter to the wall.
Figure 1

Attachment of film badge dosimeters to staff member body: A: lapel; B: lapel under lead apron; C: lapel over lead apron; D: inside the surgical glove; E; under sterile surgical glove; F: position on the wrist.

Figure 2

Operating room bed, under which a film badge dosimeter was placed and the operating room wall with a dosimeter badge attached.

After each study period, the film badges were developed and analyzed in a radiology laboratory, and the following variables were recorded: date, procedure type and duration, voltage and electrical current of the C-arm, and the distances between the main beam of the C-arm and the professionals. These data were recorded on a filing card, as shown in Figure 3.
Figure 3

Filling card used to record the data.

RESULTS

During the twelve-month period, 81 spinal surgeries were performed. The mean distances from the professionals to the radiation emission site during the procedures were almost the same in all procedures (Table 1). Table 2 shows data relating to the duration of surgery, and the voltage and electric current registered in the C-arm, which varied according to the type of surgery performed.
Table 1

Medium distance from the professionals to the radiation emission site during the procedures.

ProfessionalSurgeon 1Surgeon 2InstrumentistResidentC-arm operatorAuxiliary
Distance (cm)404060100100120
Table 2

Data collected during the procedures. Periods refer to the time taken for each batch of film badges to be taken to laboratory analysis.

ProcedureDuration (h, min)Electric tension (kV)Electric current (mA)
Period A: two months
Scoliosis1h10'750,9
Scoliosis0h50'620,7
Lumbar arthrodesis2h00'850,9
Scoliosis0h50'800,4
Lumbar arthrodesis1h10'900,8
Lumbar arthrodesis1h30'861,4
Lumbar arthrodesis2h10'881,3
Period B: November, one month
Lumbar arthrodesis1h20'821,1
Lumbar arthrodesis1h35'861,3
Cervical arthrodesis2h20'802,3
Scoliosis1h10'801,1
Lumbar arthrodesis2h10'920,9
Period C: one month
Infiltration2h05'651,2
Scoliosis3h00'741,6
Lumbar arthrodesis1h30'1041,0
Lumbar arthrodesis3h00'800,9
Infiltration1h55'711,4
Lumbar scoliosis1h45'772,1
Period D: one month
Lumbar arthrodesis2h40'861,4
Lumbar arthrodesis1h50'902,4
Cervical arthrodesis1h20'751,6
Lumbar arthrodesis2h510'881,8
Period E: one month
Scoliosis3h10'
Lumbar arthrodesis2h10'702,0
Lumbar arthrodesis1h55'781,8
Period F: two months
Lumbar arthrodesis1h30'1041,0
Lumbar arthrodesis3h00'800,9
Scoliosis3h30'901,2
Lumbar arthrodesis3h00'722,0
Scoliosis3h30'
Scoliosis2h00'
Scoliosis1h30'
Period G: three months
Cervical arthrodesis5h00'804,0
Scoliosis6h40'501,3
Cervical arthrodesis1h00'804,0
Lumbar arthrodesis0h50'1040,7
Arthrodesis l1h05'1010,7
Arthrodesis l3h00'702,0
Scoliosis4h00'1040,8
Cervical arthrodesis4h00'801,6
Scoliosis4h00'883,6
Lumbar arthrodesis3h30'823,2
Lumbar arthrodesis5h00'953,2
Lumbar arthrodesis2h00'903,6
Cervical arthrodesis3h00'682,0
Scoliosis4h00'883,6
Cervical arthrodesis2h40'602,8
Cervical arthrodesis1h40'602,0
Scoliosis2h00'602,6
Lumbar arthrodesis2h00'653,0
Lumbar arthrodesis3h30'1011,6
Cervical arthrodesis5h00'713,0
Scoliosis3h00'701,4
Scoliosis4h00'1052,4
Scoliosis4h00'762,7
Lumbar arthrodesis3h30'1097,1
Cervical arthrodesis3h30'601,4
Lumbar arthrodesis3h00'703,2
Scoliosis4h00'702,0
Period H: two months
Lumbar arthrodesis3h30'482,4
Lumbar arthrodesis1h10'902,4
Lumbar arthrodesis2h30'623,4
Cervical arthrodesis4h00'652,4
Cervical arthrodesis2h00'701,6
Scoliosis2h30'845,6
Lumbar arthrodesis2h20'762,1
Cervical arthrodesis4h00'641,8
Scoliosis3h00'741,6
Lumbar arthrodesis1h30'701,6
Lumbar arthrodesis3h30'764,0
Lumbar arthrodesis4h00'701,6
Decompression5h30'762,4
Scoliosis5h00'741,6
Lumbar arthrodesis3h30'762,4
Lumbar arthrodesis3h30'823,0
Lumbar arthrodesis4h00'802,6
Scoliosis4h30'802,0
Lumbar arthrodesis4h30'802,6
Scoliosis6h00'704,5
Cervical arthrodesis3h30'814,0
Lumbar arthrodesis3h30'704,0
Lumbar arthrodesis3h30'823,0
The radiation doses recorded in the dosimeter devices over periods ranging from one to three months were grouped and analyzed in the laboratory. The results of each period are presented in Table 3. The total radiation doses varied from 0.16 mSv (under the lead apron) to 2.29 mSv (on the wrist of the main surgeon).
Table 3

Results from radiation doses measurements per period (as indicated in Table 2); periods refer to the time taken for each batch of film badges to be taken to laboratory analysis.

ProfessionalDose (mSv) per period
ABCDEFGHTotal
Standard
Surgical center 1 – operating room wall0.060.030.080.17
Surgical center 2 – operating room wall0.050.030.090.110.28
Below patient bed0.80.440.191.43
Surgeon 1 - attached to the lapel, over the lead apron0.150.130.050.080.030.050.410.291.19
Surgeon 1 - attached to the lapel, under the lead apron0.020.020.030.090.16
Surgeon 2 - attached to the lapel, over the lead apron0.020.070.050.210.060.290.120.82
Surgeon 2 - attached to the lapel, under the lead apron0.020.020.040.060.600.74
Instrumentator (scrub nurse)0.020.090.120.23
Resident0.060.020.090.100.27
C-arm operator - attached to the lapel, over the lead apron0.060.020.090.040.080.29
C-arm operator - attached to the lapel, under the lead apron0.030.040.090.16
Auxiliary0.070.130.040.24
Surgeon 1 - wrist1.430.810.052.29
Surgeon 2 - wrist0.100.350.070.020.020.030.59

Doses not mentioned are below the minimum detectable level (0.02 mSv).

Doses not mentioned are below the minimum detectable level (0.02 mSv). The last column of Table 3 shows the cumulative dose for the entire study period of about a year. The highest estimated dose was indicated for the badge dosimeters placed over the lead cover of surgeons 1 and 2, the professionals who are positioned closest to the primary beam and the patients. In such cases the dose to the professional is lower due to the protection offered by the lead apron, as indicated by the dosimeters positioned below this protective covering.

DISCUSSION AND CONCLUSIONS

Exposure to radiation can be a cause of concern in spinal surgeries, which is a complex procedure usually involving six professionals in our service and lasting from one to six hours. Handling the fluoroscope potentially exposes surgeons to a radiation dose higher than the annual recommendation in orthopaedic procedures. A study on different types of surgery using fluoroscopy has shown that the total dose in orthopaedic procedures can reach 2.92 mGy/m2, with an effective dose of 58.4 mSv. As pointed out by Anupam Mahajan et al., orthopaedic surgeons are not considered as to be workers exposed to radiation, and radiation is usually seen as an additional secondary occupational hazard for them. Still, there are non-surgeon personnel in the operating room, and these should receive monitoring and protection. An experimental study using phantom anthropometric figures and simulating the spinal surgery setting showed that scatter radiation doses decreased as distance from the patient increased during the use of C-arm fluoroscopy. The study showed that the distance between the patient and the C-arm configuration, among other factors, could reduce radiation exposure during intraoperative use of the C-arm. The position of the professional can also significantly alter the radiation exposure, and the radiation dose may differ between the first assistant, the theatre nurse and the anesthesiologist, who might be exposed to higher radiation doses than the surgeon. In our study, we measured exposure to radiation in a real setting: the operating room during spinal surgeries, with real professionals at work, and we carefully evaluated the distances and the different measurement sites. It seems that in our setting, the exposure to radiation is below the legal requirements. The wrist badge dosimeters evidenced that the radiation dose in our surgeons’ hands is also well below the maximum limit of 500 mSv per year. , , , Analysis of the radiation dose registered by the film badge dosimeters, giving the estimated radiation to which staff are exposed, showed that there were no cases in which the values were above the level stipulated as total tolerated levels for an individual (1,5 mSv). , The doses to which all professionals were exposed in this study were slightly above the natural radiation levels in the environment, but far below the maximum dose allowed by local law (20 mSv as a five-year average). , , , The results show that in this type of medical procedure, and under these study conditions, the resulting doses are low and the recommended limits were respected. Nevertheless, continual training of professionals is essential to ensure that medical staff and patients are protected against radiation.
  8 in total

1.  Measurements of surgeons' exposure to ionizing radiation dose during intraoperative use of C-arm fluoroscopy.

Authors:  Kisung Lee; Kyoung Min Lee; Moon Seok Park; Boram Lee; Dae Gyu Kwon; Chin Youb Chung
Journal:  Spine (Phila Pa 1976)       Date:  2012-06-15       Impact factor: 3.468

2.  Radiation dose to surgeons in theatre.

Authors:  B van der Merwe
Journal:  S Afr J Surg       Date:  2012-03-29       Impact factor: 0.375

3.  Occupational radiation exposure from C arm fluoroscopy during common orthopaedic surgical procedures and its prevention.

Authors:  Anupam Mahajan; Sumant Samuel; Atul K Saran; M K Mahajan; M K Mam
Journal:  J Clin Diagn Res       Date:  2015-03-01

Review 4.  Radiation dosimetry for diagnostic medical exposures.

Authors:  C J Martin
Journal:  Radiat Prot Dosimetry       Date:  2008-03-28       Impact factor: 0.972

5.  [Evaluation of radiation exposure of personnel in an orthopaedic and trauma operation theatre using the new real-time dosimetry system "dose aware"].

Authors:  M C Müller; A Strauss; R Pflugmacher; C P Nähle; P H Pennekamp; C Burger; D C Wirtz
Journal:  Z Orthop Unfall       Date:  2014-08-21       Impact factor: 0.923

6.  Cervical spine imaging using standard C-arm fluoroscopy: patient and surgeon exposure to ionizing radiation.

Authors:  Brian D Giordano; Judith F Baumhauer; Thomas L Morgan; Glenn R Rechtine
Journal:  Spine (Phila Pa 1976)       Date:  2008-08-15       Impact factor: 3.468

7.  The Radiation Exposure of Radiographer Related to the Location in C-arm Fluoroscopy-guided Pain Interventions.

Authors:  Young Jae Chang; Ah Na Kim; In Su Oh; Nam Sik Woo; Hae Kyoung Kim; Jae Hun Kim
Journal:  Korean J Pain       Date:  2014-03-28

8.  Analysis of radiation risk to patients from intra-operative use of the mobile X-ray system (C-arm).

Authors:  Yang-Sub Lee; Hae-Kag Lee; Jae-Hwan Cho; Ham-Gyum Kim
Journal:  J Res Med Sci       Date:  2015-01       Impact factor: 1.852

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.