Abstract

OBJECTIVES: The choice of annuloplasty device is fundamental at the time of mitral valve repair, the goal being to optimally restore the physiological 3-dimensional (3D) structure and dynamics of the mitral annulus (MA). This study evaluated MA dynamics after annuloplasty with 2 different semi-rigid devices.
METHODS: Thirty-three patients eligible for mitral valve repair were selected for annuloplasty with the Physio II ring (Edwards Lifesciences, Irving, CA, USA) (n = 17) or the Memo 3D ring (LivaNova, Saluggia, Italy) (n = 16). MA dynamics were assessed with transoesophageal 3D echocardiography intraoperatively and 1 year after repair.
RESULTS: The postoperative changes in the anteroposterior diameter {3.7% [standard deviation (SD) 2.7] vs 1.9% [SD 1.3]; P = 0.013} and in the annular height [27.7% (SD 8.7) vs 18.0% (SD 13.9); P = 0.003] were significantly larger with the Memo 3D ring during the cardiac cycle. The restoration of the saddle shape at baseline was superior with the Physio II ring, defined by a larger systolic annular height-to-commissural width ratio [15.1% (SD 2.3) vs 7.1% (SD 2.4); P < 0.001]. These observations of MA dynamics were sustained after 1 year, shown by a greater anteroposterior extension [5.1% (SD 1.0) vs 1.7% (SD 1.6); P = 0.002] and change in annular height-to-commissural width ratio [15.7% (SD 12.7) vs 3.1% (SD 3.0); P = 0.020] for the Memo 3D ring. There were no significant differences in mitral valve function between the 2 devices.
CONCLUSIONS: The MA dynamics after annuloplasty with the Physio II and Memo 3D rings demonstrated a better systolic 3D restoration of the saddle shape with the Physio II ring, whereas the saddle-shaped geometry improved significantly with the Memo 3D ring, as a dynamic phenomenon. The Memo 3D ring also showed increased anteroposterior annular mobility and folding dynamics throughout the cardiac cycle. Moreover, the observed differences in MA dynamics between both devices appeared to be sustainable 1 year after ring implantation.
© The Author(s) 2019. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  3-Dimensional echocardiography; Mitral annulus; Mitral valve repair

Year:  2019        PMID: 30770923     DOI: 10.1093/ejcts/ezz035

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  2 in total

1.  Comparison of flexible, open with semi-rigid, closed annuloplasty-rings for mitral valve repair.

Authors:  Ayse Cetinkaya; Maryam Waheed; Karin Bramlage; Oliver Johannes Liakopoulos; Mohamed Zeriouh; Stefan Hein; Peter Bramlage; Markus Schönburg; Yeong-Hoon Choi; Manfred Richter
Journal:  J Cardiothorac Surg       Date:  2021-03-20       Impact factor: 1.637

2.  Every day mitral valve reconstruction: What has changed over the last 15 years?

Authors:  Farnoosh Motazedian; Roya Ostovar; Martin Hartrumpf; Filip Schröter; Johannes M Albes
Journal:  PLoS One       Date:  2022-10-03       Impact factor: 3.752

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.