Design and preliminary biomechanical analysis of a novel motion preservation device for lumbar spinal disease after vertebral corpectomy.

Jiantao Liu1, Xijing He2, Zhengchao Gao2, Binbin Niu2, Dongbo Lv1, Yanzheng Gao3.   

Abstract

OBJECTIVE: To design a novel prosthesis, a movable artificial lumbar complex (MALC), for non-fusion reconstruction after lumbar subtotal corpectomy and to evaluate the stability, range of motion and load-bearing strength in the human cadaveric lumbar spine.
METHODS: Biomechanical tests were performed on lumbar spine specimens from 15 healthy cadavers which were divided in three groups: non-fusion, fusion and intact group. The range of motion (ROM), stability and load-bearing strength were measured.
RESULTS: The prosthesis was composed of three parts: the upper and lower artificial lumbar discs and the middle artificial vertebra. Both the MALC and titanium mesh cage re-established vertebral height, and no spinal cord compression or prosthesis dislocation was observed at the operative level. Regarding stability, there was no significant difference in all directions between the intact group and non-fusion group (P > 0.05). Segment movements of the specimens in the non-fusion group revealed significantly decreased T12-L1 ROM and significantly increased L1-2 and L2-3 ROM in flexion/extension and lateral bending compared with those in the fusion group (P < 0.05). Regarding load-bearing strength, when the lumbar vertebra was ruptured, there was no damage to the MALC and titanium mesh cage, but the maximum load in the non-fusion group was larger (P > 0.05).
CONCLUSIONS: Compared with titanium cages, the MALC prosthesis not only restored the vertebral height and effectively preserved segment movements without any abnormal gain of mobility in adjacent inter-vertebral spaces but also bore the lumbar load and reduced the local stress load of adjacent vertebral endplates.

Entities:  

Keywords:  Artificial vertebral body; Biomechanical phenomena; Non-fusion; Prostheses and implants; Spine

Mesh:

Year:  2019        PMID: 30747259     DOI: 10.1007/s00402-018-03106-2

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  2 in total

1.  Application of 3-dimensional printing technology combined with guide plates for thoracic spinal tuberculosis.

Authors:  Yifan Wang; Shiyuan Shi; Qi Zheng; Yanghui Jin; Yingjie Dai
Journal:  Medicine (Baltimore)       Date:  2021-02-12       Impact factor: 1.817

2.  Cervical non-fusion using biomimetic artificial disc and vertebra complex: technical innovation and biomechanics analysis.

Authors:  Jialiang Li; Pengrong OuYang; Xijing He; Xinyu Wei; Zhongwei Sun; Hui Dong; Zhijing Wen; Yibin Wang; Pengzhen Gu; Teng Lu; Ning Liu; Haopeng Li
Journal:  J Orthop Surg Res       Date:  2022-02-23       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.