In vitro testing of an intra-ventricular assist device.

Shidong Zhu1, Lin Luo1, Bibo Yang2, Xinghui Li1, Kai Ni1, Qian Zhou1, Xiaohao Wang1.   

Abstract

A novel pulsatile assist device, intra-ventricular assist device, was proposed to address various disadvantages existing in conventional pulsatile assist device, such as the large size, accessories and reduced pulsatility. The assist device was designed, fabricated and implanted into the sac from left ventricular apex in a home-designed mock circulatory system. In vitro test was carried out and results demonstrated that the response time did not vary with the heart rate, and co-pulsatiled synchronously with native heart by electrocardiograph. The key parameter, stroke volume of proposed device was precisely measured under different afterloads (60, 80, 100, and 120 mmHg), drive pressure (from 90 to 300 mmHg at 30 mmHg intervals), and heart rate (45-150 beats per minute). The measurement results revealed that the output characteristics of device, stroke volume increased with increasing drive pressure but decreased with increasing peripheral resistance, were consistent with the native heart. The proposed pump was then coupled with mock system that was set to a heart failure mode and the circulatory responses were tested. Results showed that the device improved left ventricular pressure from 106 to 158 mmHg, and stroke volume from 25.5 to 44 ml at 90 bpm.

Entities:  

Keywords:  heart failure; hemodynamics; left ventricular assist device; mock circulatory system

Mesh:

Year:  2019        PMID: 30741034     DOI: 10.1080/24699322.2018.1560099

Source DB:  PubMed          Journal:  Comput Assist Surg (Abingdon)        ISSN: 2469-9322            Impact factor:   1.787


  1 in total

1.  Finite Element Driven Design Domain Identification of a Beating Left Ventricular Simulator.

Authors:  Utku Gulbulak; Atila Ertas
Journal:  Bioengineering (Basel)       Date:  2019-09-13
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.