Thomas J Buell1, Shay Bess2, Ming Xu3, Frank J Schwab4, Virginie Lafage4, Christopher P Ames5, Christopher I Shaffrey1, Justin S Smith1.
Abstract
OBJECTIVEProximal junctional kyphosis (PJK) is, in part, due to altered segmental biomechanics at the junction of rigid instrumented spine and relatively hypermobile non-instrumented adjacent segments. Proper application of posteriorly anchored polyethylene tethers (i.e., optimal configuration and tension) may mitigate adjacent-segment stress and help prevent PJK. The purpose of this study was to investigate the impact of different tether configurations and tensioning (preloading) on junctional range-of-motion (ROM) and other biomechanical indices for PJK in long instrumented spine constructs.METHODSUsing a validated finite element model of a T7-L5 spine segment, testing was performed on intact spine, a multilevel posterior screw-rod construct (PS construct; T11-L5) without tether, and 15 PS constructs with different tether configurations that varied according to 1) proximal tether fixation of upper instrumented vertebra +1 (UIV+1) and/or UIV+2; 2) distal tether fixation to UIV, to UIV-1, or to rods; and 3) use of a loop (single proximal fixation) or weave (UIV and/or UIV+1 fixation in addition to UIV+1 and/or UIV+2 proximal attachment) of the tether. Segmental ROM, intradiscal pressure (IDP), inter- and supraspinous ligament (ISL/SSL) forces, and screw loads were assessed under variable tether preload.RESULTSPS construct junctional ROM increased abruptly from 10% (T11-12) to 99% (T10-11) of baseline. After tethers were grouped by most cranial proximal fixation (UIV+1 vs UIV+2) and use of loop versus weave, UIV+2 Loop and/or Weave most effectively dampened junctional ROM and adjacent-segment stress. Different distal fixation and use of loop versus weave had minimal effect. The mean segmental ROM at T11-12, T10-11, and T9-10, respectively, was 6%, 40%, and 99% for UIV+1 Loop; 6%, 44%, and 99% for UIV+1 Weave; 5%, 23%, and 26% for UIV+2 Loop; and 5%, 24%, and 31% for UIV+2 Weave.Tethers shared loads with posterior ligaments; consequently, increasing tether preload tension reduced ISL/SSL forces, but screw loads increased. Further attenuation of junctional ROM and IDP reversed above approximately 100 N tether preload, suggesting diminished benefit for biomechanical PJK prophylaxis at higher preload tensioning.CONCLUSIONSIn this study, finite element analysis demonstrated UIV+2 Loop and/or Weave tether configurations most effectively mitigated adjacent-segment stress in long instrumented spine constructs. Tether preload dampened ligament forces at the expense of screw loads, and an inflection point (approximately 100 N) was demonstrated above which junctional ROM and IDP worsened (i.e., avoid over-tightening tethers). Results suggest tether configuration and tension influence PJK biomechanics and further clinical research is warranted.Entities:
Keywords: ASD = adult spinal deformity; IDP = intradiscal pressure; ISL = interspinous ligament; PJK = proximal junctional kyphosis; PS = multilevel posterior screw–rod; ROM = range of motion; SSL = supraspinous ligament; UIV = upper instrumented vertebra; adjacent-segment disease; adult spinal deformity; complication; finite element analysis; proximal junctional failure; proximal junctional kyphosis; scoliosis; spine surgery; tether
Year: 2019 PMID: 30738393 DOI: 10.3171/2018.10.SPINE18429
Source DB: PubMed Journal: J Neurosurg Spine ISSN: 1547-5646