Radiation Dose for Pediatric CT: Comparison of Pediatric versus Adult Imaging Facilities.

Keith J Strauss1, Elanchezhian Somasundaram1, Debapriya Sengupta1, Jennifer R Marin1, Samuel L Brady1.   

Abstract

Background The American College of Radiology Dose Index Registry for CT enables evaluation of radiation dose as a function of patient characteristics and examination type. The hypothesis of this study was that academic pediatric CT facilities have optimized CT protocols that may result in a lower and less variable radiation dose in children. Materials and Methods A retrospective study of doses (mean patient age, 12 years; age range, 0-21 years) was performed by using data from the National Radiology Data Registry (year range, 2016-2017) (n = 239 622). Three examination types were evaluated: brain without contrast enhancement, chest without contrast enhancement, and abdomen-pelvis with intravenous contrast enhancement. Three dose indexes-volume CT dose index (CTDIvol), size-specific dose estimate (SSDE), and dose-length product (DLP)-were analyzed by using six different size groups. The unequal variance t test and the F test were used to compare mean dose and variances, respectively, at academic pediatric facilities with those at other facility types for each size category. The Bonferroni-Holm correction factor was applied to account for the multiple comparisons. Results Pediatric radiation dose in academic pediatric facilities was significantly lower, with smaller variance for all brain, 42 of 54 (78%) chest, and 48 of 54 (89%) abdomen-pelvis examinations across all six size groups, three dose descriptors, and when compared with that at the other three facilities. For example, abdomen-pelvis SSDE for the 14.5-18-cm size group was 3.6, 5.4, 5.5, and 8.3 mGy, respectively, for academic pediatric, nonacademic pediatric, academic adult, and nonacademic adult facilities (SSDE mean and variance P < .001). Mean SSDE for the smallest patients in nonacademic adult facilities was 51% (6.1 vs 11.9 mGy) of the facility's adult dose. Conclusion Academic pediatric facilities use lower CT radiation dose with less variation than do nonacademic pediatric or adult facilities for all brain examinations and for the majority of chest and abdomen-pelvis examinations. © RSNA, 2019 See also the editorial by Strouse in this issue.

Entities:  

Mesh:

Year:  2019        PMID: 30720404     DOI: 10.1148/radiol.2019181753

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

Review 1.  Radiation use in diagnostic imaging in children: approaching the value of the pediatric radiology community.

Authors:  Donald P Frush; Erich Sorantin
Journal:  Pediatr Radiol       Date:  2021-03-20

2.  Deep learning image reconstruction in pediatric abdominal and chest computed tomography: a comparison of image quality and radiation dose.

Authors:  Kun Zhang; Xiang Shi; Shuang-Shuang Xie; Ji-Hang Sun; Zhuo-Heng Liu; Shuai Zhang; Jia-Yang Song; Wen Shen
Journal:  Quant Imaging Med Surg       Date:  2022-06

3.  Image quality and radiation dose of dual source high pitch computed tomography in pediatric congenital heart disease.

Authors:  Dmitrij Kravchenko; Christopher Hart; Stephan Garbe; Julian A Luetkens; Alexander Isaak; Narine Mesropyan; Mathieu Vergnat; Judith Leyens; Ulrike Attenberger; Daniel Kuetting
Journal:  Sci Rep       Date:  2022-06-15       Impact factor: 4.996

4.  Radiation Dose Management in Pediatric Brain CT According to Age and Weight as Continuous Variables.

Authors:  Yusuke Inoue; Hiroyasu Itoh; Anri Waga; Ryosuke Sasa; Kohei Mitsui
Journal:  Tomography       Date:  2022-04-01

5.  Comparing definitions of a pediatric emergency department.

Authors:  Margaret E Samuels-Kalow; Ashley F Sullivan; Krislyn M Boggs; Jingya Gao; Elizabeth R Alpern; Carlos A Camargo
Journal:  J Am Coll Emerg Physicians Open       Date:  2021-06-22
  5 in total