| Literature DB >> 30718058 |
Keiko Iwakoshi1, Shoichi Tahara2, Yoko Uematsu2, Yukiko Yamajima2, Hiroyuki Miyakawa2, Kimio Monma2, Chigusa Kobayashi2, Ichiro Takano2.
Abstract
A new method for the simultaneous determination of two sweeteners (Advantame and Neotame) in processed foods using liquid chromatography (LC) with tandem mass spectrometry(MS/MS) was developed herein. Chromatographic separations were performed using an ACQUITY UPLC CSH C18 column at 40 °C via a mobile phase comprising 10-mmol/L ammonium formate and methanol. Samples were prepared via rapid dialysis using 30% methanol solution in a thermostatic shaker set at 160 rpm and 50 °C for 1 h. The matrix in the test solution had no effect on the identification and quantification of the compound without a clean-up step using solid-phase extraction (SPE). This method satisfied all validation criteria with a limit of quantification (LOQ) of 0.01 μg/g for all samples. Using this method, the amounts of Advantame and Neotame in 24 processed foods were subsequently investigated, with the results indicating their detection beyond the lower LOQ. Moreover, a multiple reaction monitoring information-dependent acquisition-enhanced product ion (MRM-IDA-EPI) method was developed and described to further enhance product-identification ability.Entities:
Keywords: Advantame; Charge-surface-modified C18; Dialysis; External solvent-calibration curve; Food; IDA; Japanese market; Liquid chromatography (LC); MRM; Neotame; Processed foods; Sweetener; Tandem mass spectrometry (MS/MS)
Mesh:
Substances:
Year: 2019 PMID: 30718058 DOI: 10.1016/j.chroma.2019.01.036
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759