| Literature DB >> 30715042 |
Makie Terazawa1, Momoko Karita2, Shinya Kumagai3, Minoru Sasaki4.
Abstract
In this work, a respiratory sensor is studied, measuring the capacitance constructed by attached electrodes on the abdomen. Based on previous findings, that skin thickness changes caused by respiration provides the signal, the fitting condition of the electrode on the skin is stabilized using a 7-μm-thick dressing film. This film can be comfortably worn for a long time, while maintaining the electrode's position on the skin. This stabilized setup enables the detection of, not only respiration, as the cyclic capacitance change, but also of minute body volume changes over the daytime, as a change in the base line indicates the quality of the sensor signal. For this demonstration, the respiration signal is measured during the daily activity of exercise and 6-min walks.Entities:
Keywords: capacitance constructed across skin; respiratory sensor; wearable sensor
Year: 2018 PMID: 30715042 PMCID: PMC6267211 DOI: 10.3390/mi9110543
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 2.891
Figure 1(a) Skin structure. Model of the electrodes constructing the capacitance across the skin under (b) exhalation and (c) inhalation conditions. (d) The two capacitances are considered to connect in series inside the body and are measured by the outside circuit.
List of tapes tested for fixing the electrodes.
| - | Sports Tape | Silicone Gel | Dressing Film |
|---|---|---|---|
| Original use | For taping in sports | For artificial scabs | For fixing the needle of a drip infusion |
| Specification | 0.4-mm-thick tape used for under-layer and cover. | 2-mm-thick plate used for under-layer. Cover is sports tape. | 7-μm -thick films used for under-layer and cover. |
| Setup for fixing electrode |
|
|
|
| Signal evaluation | Drift is large for long time measurement. | Signal is stable but its magnitude decreases. | Signal is stable and large. |
| Judgement | bad | bad | good |
Figure 2Typical electrodes attached to the skin; conductive textiles are used.
Figure 3Typical daily capacitance change, including respiration. Since the posture generates the capacitance change, it is stabilized by lying on a mat.
List of statistical evaluations of the peak and valley values in Figure 3.
| - | Peak | Valley | ||||
|---|---|---|---|---|---|---|
| Timing | Average (pF) | Standard Deviation (pF) | Ratio | Average (pF) | Standard Deviation (pF) | Ratio |
| 10:00 a.m. | 87.15 | 1.170 | 0.0134 | 73.83 | 0.201 | 0.0027 |
| Before lunch | 80.08 | 0.372 | 0.0046 | 68.67 | 0.282 | 0.0041 |
| During lunch | 79.66 | 0.394 | 0.0049 | 69.81 | 0.319 | 0.0046 |
| After lunch | 87.16 | 0.669 | 0.0077 | 75.02 | 0.321 | 0.0043 |
| 3:00 p.m. | 76.65 | 0.523 | 0.0068 | 65.25 | 0.422 | 0.0065 |
| 4:00 p.m. | 74.10 | 0.439 | 0.0059 | 63.67 | 0.255 | 0.0040 |
Figure 4Abdomen capacitance signals during exercise. (a) Twisting the upper body left and right. (b) Bending the upper body forward and backward.
Average respiratory rate during exercise (shown in Figure 4). Twisting the upper body left and right.
| Posture | Front | Left | Front | Right | Front |
|---|---|---|---|---|---|
| Period in | 2.43–6.89 | 9.49–14.15 | 16.76–21.06 | 23.57–27.98 | 30.51–34.64 |
| Respiratory rate (cycle/min) | 26.9 | 25.7 | 27.9 | 27.2 | 29.1 |
Average respiratory rate during exercise (shown in Figure 4). Bending the upper body forward and backward.
| Posture | Front | Forward | Front | Back | Front |
|---|---|---|---|---|---|
| Period in | 2.30–5.90 | 7.92–12.59 | 15.09–18.78 | 20.93–24.50 | 27.02–31.23 |
| Respiratory rate (cycle/min) | 33.3 | 25.7 | 32.5 | 33.7 | 28.5 |
Figure 5Abdomen capacitance signals measured by the wireless unit (a) during gentle walking and (b) during harder walking for 6 min.