Reliability of, and Agreement Between, two Breath-by-Breath Indirect Calorimeters at Varying Levels of Inspiratory Oxygen.

Mathias Krogh Poulsen1, Lars Pilegaard Thomsen1, Søren Kjaergaard2, Stephen Edward Rees1, Dan Stieper Karbing1.   

Abstract

BACKGROUND: Indirect calorimetry (IC) is considered the accurate way of measuring energy expenditure (EE). IC devices often apply the Haldane transformation, introducing errors at inspiratory oxygen fraction (FiO2 ) >60%. The aim was to assess measurement reliability and agreement between an unevaluated IC (device 2) (Beacon Caresystem, Mermaid Care A/S, Noerresundby, Denmark) not using Haldane transformation and an IC that does (device 1) (Ecovx, GE, Helsinki, Finland) at varying FiO2 .
METHODS: Twenty healthy male subjects participated, with 16 completing the study (33 ± 9 years, 83.3 ± 16 kg, 1.83 ± 0.08 m). Subjects were mechanically ventilated in pressure support (3cmH2 O; positive end-expiratory pressure: 3cmH2 O) at FiO2 of 21%, 50%, 85%, and 21% for 15 minutes at each FiO2 . Mean EE, oxygen consumption (VO2 ), and CO2 production (VCO2 ) were compared within and between devices across FiO2 levels.
RESULTS: Device 2 showed within-device EE significant differences at 21% vs 50% FiO2 and device 1 for VCO2 at 50% vs. 85% FiO2 . For all variables, both devices showed reliable measurements at 21% and 50% FiO2 , but at 85%, FiO2 bias and limits of agreement increased. Between devices, there were significant differences for EE at both 21% and 85% FiO2 for VO2 and for VCO2 at 85% FiO2 .
CONCLUSION: Both systems measured EE, VO2 , and VCO2 at 21%-85% FiO2 reliably but with bias at 85% FiO2 . The devices were in agreement at 21% and 50% FiO2 , but further studies need to confirm accuracy at high FiO2 .
© 2019 American Society for Parenteral and Enteral Nutrition.

Entities:  

Keywords:  bias; indirect calorimetry; mechanical ventilation; reproducibility of results; resting energy expenditure

Mesh:

Year:  2019        PMID: 30714634     DOI: 10.1002/ncp.10250

Source DB:  PubMed          Journal:  Nutr Clin Pract        ISSN: 0884-5336            Impact factor:   3.080


  2 in total

1.  External Validation with Accuracy Confounders of VCO2-Derived Predicted Energy Expenditure Compared to Resting Energy Expenditure Measured by Indirect Calorimetry in Mechanically Ventilated Children.

Authors:  Panagiotis Briassoulis; Stavroula Ilia; Efrossini Briassouli; George Briassoulis
Journal:  Nutrients       Date:  2022-10-10       Impact factor: 6.706

2.  External Validation of Equations to Estimate Resting Energy Expenditure in Critically Ill Children and Adolescents with and without Malnutrition: A Cross-Sectional Study.

Authors:  George Briassoulis; Efrossini Briassouli; Stavroula Ilia; Panagiotis Briassoulis
Journal:  Nutrients       Date:  2022-10-06       Impact factor: 6.706

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.