Quantitative analysis of radiation dosage and image quality between digital breast tomosynthesis (DBT) with two-dimensional synthetic mammography and full-field digital mammography (FFDM).

Yoonmi Choi1, Ok-Hee Woo2, Hye-Sun Shin1, Kyu Ran Cho1, Bo Kyoung Seo1, Ga-Young Choi1.   

Abstract

PURPOSE: Currently in diagnostic setting for breast cancer, FFDM and DBT are performed conjunctively. However, performing two imaging modalities may increase radiation exposure by double. Two-dimensional reconstructed images created from DBT with 2DSM, has a potential to replace conventional FFDM in concerning both radiation dosage and image quality. With increasing concerns for individual radiation exposure, studies analyzing radiation dosage in breast imaging modalities are needed. This study compared radiation dosage and image quality between DBT + 2DSM versus FFDM. METHODS AND MATERIALS: 374 patients (mean age 52 years) who underwent both DBT and FFDM were retrospectively reviewed. Radiation dosage data were obtained by radiation dosage scoring and monitoring program Radimetrics (Bayer HealthCare, Whippany, NJ). Entrance dose and mean glandular doses in each breast were obtained for both modalities. To compare image quality of DBT + 2DSM and FFDM, a 5-point scoring system for lesion clarity was assessed. The parameters of radiation dosage (entrance dose, mean glandular dose) and image quality (lesion clarity scoring) were compared.
RESULTS: For entrance dose, DBT had lower mean dosage (14.8 mGy) compared with FFDM (21.8 mGy, p-value < 0.0001). Mean glandular doses for both breasts were lower in DBT (Left 1.74, Right 2.1) compared with FFDM (Left 2.85, Right 2.74, p-value < 0.0001). Lesion clarity score was higher in DBT with 2DSM (mean score 4.03) compared with FFDM (3.82, p-value < 0.0001).
CONCLUSION: DBT showed lower radiation entrance dose and mean glandular doses to both breasts compared with FFDM. DBT + 2DSM had better image quality than FFDM, suggesting that DBT with 2DSM has potential as an alternative to FFDM.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  2-Dimensional synthetic mammography; Digital breast tomosynthesis; Full-field digital mammography; Image quality; Radiation dose; Radimetrics

Mesh:

Year:  2019        PMID: 30703693     DOI: 10.1016/j.clinimag.2019.01.014

Source DB:  PubMed          Journal:  Clin Imaging        ISSN: 0899-7071            Impact factor:   1.605


  1 in total

1.  Breast Radiation Exposure of 3D Digital Breast Tomosynthesis Compared to Full-Field Digital Mammography in a Clinical Follow-Up Setting.

Authors:  Marcel Opitz; Sebastian Zensen; Katharina Breuckmann; Denise Bos; Michael Forsting; Oliver Hoffmann; Martin Stuschke; Axel Wetter; Nika Guberina
Journal:  Diagnostics (Basel)       Date:  2022-02-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.