'Relationship between thermal dose and cell death for "rapid" ablative and "slow" hyperthermic heating'.

Petros X E Mouratidis1, Ian Rivens1, John Civale1, Richard Symonds-Tayler1, Gail Ter Haar1.   

Abstract

AIM: Thermal isoeffective dose (TID) has not been convincingly validated for application to predict biological effects from rapid thermal ablation (e.g., using >55 °C). This study compares the classical method of quantifying TID (derived from hyperthermia data) with a temperature-adjusted method based on the Arrhenius model for predicting cell survival in vitro, after either 'rapid' ablative or 'slow' hyperthermic exposures.
METHODS: MTT assay viability data was obtained from two human colon cancer cell lines, (HCT116, HT29), subjected to a range of TIDs (120-720 CEM43) using a thermal cycler for hyperthermic (>2 minutes, <50 °C) treatments, or a novel pre-heated water bath based technique for ablative exposures (<10 seconds, >55 °C). TID was initially estimated using a constant RCEM>43°C=0.5, and subsequently using RCEM(T), derived from temperature dependent cell survival (injury rate) Arrhenius analysis.
RESULTS: 'Slow' and 'rapid' exposures resulted in cell survival and significant regrowth (both cell lines) 10 days post-treatment for 240 CEM43 (RCEM>43°C=0.5), while 340-550 CEM43 (RCEM>43°C =0.5) delivered using 'rapid' exposures showed 12 ± 6% viability and 'slow' exposures resulted in undetectable viability. Arrhenius analysis of experimental data (activation energy ΔE = 5.78 ± 0.04 × 105 J mole-1, frequency factor A = 3.27 ± 11 × 1091 sec-1) yielded RCEM=0.42 * e0.0041*T which better-predicted cell survival than using R CEM> 43°C=0.5.
CONCLUSIONS: TID calculated using an RCEM(T) informed by Arrhenius kinetic parameters provided a more consistent, heating strategy independent, predictor of cell viability, improving dosimetry of ablative thermal exposures. Cell viability was only undetectable above 305 ± 10 CEM43 using this revised measure.
© 2019 The Author(s). Published with license by Taylor & Francis Group, LLC This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Entities:  

Keywords:  Arrhenius; Thermal dose; heat; hyperthermia; thermal ablation

Mesh:

Year:  2019        PMID: 30700171     DOI: 10.1080/02656736.2018.1558289

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  4 in total

1.  Temperature processing and distribution in larynx thermal inhalation injury with analogy to human airway cells: a mechanism of protection.

Authors:  Huihao Jiang; Xiaocheng Zhou; Guoan Zhang
Journal:  Am J Transl Res       Date:  2022-06-15       Impact factor: 3.940

2.  Design and Characterization of an RF Applicator for In Vitro Tests of Electromagnetic Hyperthermia.

Authors:  Riccardo Ferrero; Ioannis Androulakis; Luca Martino; Robin Nadar; Gerard C van Rhoon; Alessandra Manzin
Journal:  Sensors (Basel)       Date:  2022-05-10       Impact factor: 3.847

3.  Hybrid Magnetic Nanovectors Promote Selective Glioblastoma Cell Death through a Combined Effect of Lysosomal Membrane Permeabilization and Chemotherapy.

Authors:  Carlotta Pucci; Daniele De Pasquale; Attilio Marino; Chiara Martinelli; Simone Lauciello; Gianni Ciofani
Journal:  ACS Appl Mater Interfaces       Date:  2020-06-08       Impact factor: 9.229

Review 4.  Latest Advances in the Use of Therapeutic Focused Ultrasound in the Treatment of Pancreatic Cancer.

Authors:  Petros X E Mouratidis; Gail Ter Haar
Journal:  Cancers (Basel)       Date:  2022-01-27       Impact factor: 6.575

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.