Analytical Assessment of the New Roche Cobas t 711 Fully Automated Coagulation Analyzer.

Giuseppe Lippi1, Gian Luca Salvagno1, Matteo Gelati1, Giovanni Poli1, Davide Giavarina2, Emmanuel J Favaloro3.   

Abstract

This study aimed to provide a preliminary evaluation of the analytical performance of the new Roche COBAS T 711: fully automated coagulation analyzer, which uses both liquid and lyophilized reagent cassettes. The analytical assessment included analysis of imprecision and linearity of prothrombin time (PT), activated partial thromboplastin time (APTT), and fibrinogen on COBAS T 711: analyzer. Test results of 120 routine plasma samples were also compared with those obtained using two other coagulation analyzers (Instrumentation Laboratory ACL TOP 700 and Stago STA-R MAX). The accuracy, imprecision, and comparability of manual and automatic lyophilized material resuspension were also evaluated using 200 routine plasma samples. Overall, automatic resuspension was found to be more precise than, and equally accurate as, manual reconstitution, with coefficient of variations (CV%) three- to sixfold lower compared with manual reconstitution. The analytical imprecision was found to be excellent, as attested by total CV% of 0.7% for PT, 1.7 to 1.8% for APTT, and 1.9 to 3.2% for fibrinogen. Linearity was excellent over a clinically significant range of PT, APTT, and fibrinogen values, displaying correlation coefficients comprised between 0.994 and 0.999. Methods comparison studies revealed that results of PT, APTT, and fibrinogen on COBAS T 711: are globally aligned with those obtained using identical plasma samples on IL ACL TOP 700 and Stago STA-R MAX, displaying correlation coefficients of 0.97 for PT, 0.81 and 0.88 for APTT, 0.90 and 0.94 for fibrinogen, respectively. In conclusion, the results of this preliminary evaluation demonstrate that PT, APTT, and fibrinogen on COBAS T 711: coagulation analyzer displays excellent performance for routine use in clinical laboratories. Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Entities:  

Mesh:

Year:  2019        PMID: 30699441     DOI: 10.1055/s-0038-1676578

Source DB:  PubMed          Journal:  Semin Thromb Hemost        ISSN: 0094-6176            Impact factor:   4.180


  4 in total

1.  Assessment and establishment of a reference interval for Roche Cobas t 711 coagulation analyzer for a hospital in China.

Authors:  Chi Zhu; Lili Sun; Haowei Li; Zhenghua Dong; Siyuan Yu; Xiaoming Zhao; Ji Yang; Wenjuan Wu
Journal:  Pract Lab Med       Date:  2022-02-17

2.  Performance of 6 routine coagulation assays on the new Roche Cobas t711 analyzer.

Authors:  Marlies Oostendorp; Roefke L Noorman; J Dinant Nijenhuis; Jacques B de Kok
Journal:  Pract Lab Med       Date:  2019-11-09

3.  System performance evaluation of the cobas t 711 and cobas t 511 coagulation analyzers in routine laboratory settings.

Authors:  Steve Kitchen; Moniek de Maat; Michael Nagler; Robert Jones; Anna Lowe; John Burden; Kai Groneberg; Gergely Rozsnyai
Journal:  Blood Coagul Fibrinolysis       Date:  2020-10       Impact factor: 1.061

4.  Multicenter performance evaluation and reference range determination of a new one-stage factor VIII assay.

Authors:  Anna E Lowe; Robert Jones; Steve Kitchen; Ulrich Geisen; Gergely Rozsnyai; Petra Jilma; Peter Quehenberger
Journal:  J Clin Lab Anal       Date:  2022-03-11       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.