Lead extraction for reduction of chronic pain related to cardiovascular implantable electronic device.

Kathryn Lewis1, Mingyan Dai2, Kristen K Patton3, Yong-Mei Cha2, Travis Pollema1, Gregory K Feld1, Ulrika Birgersdotter-Green1, Victor Pretorius1.   

Abstract

AIMS: Chronic pain at the cardiovascular implantable electronic device (CIED) generator or lead insertion site that is not otherwise manageable carries a IIA indication for extraction. However, limited data exist evaluating causes of pain and outcomes of extraction in eliminating pain. A multi-centre retrospective observational study was conducted to evaluate outcomes of patients undergoing device extraction for treatment of chronic device pain. METHODS AND
RESULTS: Twenty-seven out of 2188 lead extraction candidates (1.3%) met the chronic pain IIA indication for extraction [50 ± 16 years; 14 (51%) women]. Onset, severity, triggers, and pain management were measured before and after extraction. Device type, procedure done (with/without reimplantation), and positive tissue cultures were noted. Pain was reported as constant (n = 14; 50%), intermittent (n = 13; 46%), and movement-triggered (n = 14; 50%). Average severity of pain was seven out of 10 (10 being the worst). Post-extraction, 18 (66%) received freedom from pain, including all patients with poorly formed pockets (n = 2) and subclinical infections (n = 2). Of the 18, 11 underwent reimplantation (61%) without recurrent pain. Nine still had pain (44 ± 17 years; seven women) after extraction. Eight of the nine underwent reimplantation, three on the contralateral chest wall and five ipsilaterally. Pain severity decreased (n = 5), increased (n = 1), or was unchanged (n = 3).
CONCLUSION: Chronic pain at the CIED generator site can present as chronic or movement-triggered pain, and can be due to subclinical infection or a poorly formed device pocket. Extraction relieved constant and intermittent pain in two-thirds of patients. Extraction appears less successful in eliminating pain in women who undergo subsequent reimplantation. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author(s) 2019. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  Cardiovascular implantable electronic device; Cardiovascular implantable electronic device lead extraction; Chronic pain; Implantable cardioverter-defibrillator; Pacemaker

Mesh:

Year:  2019        PMID: 30698694     DOI: 10.1093/europace/euy320

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  1 in total

1.  Pacemaker Pocket Stabilization Utilizing a Novel Envelope and a Three-Point Anchoring Technique.

Authors:  Naga Venkata K Pothineni; Ramanan Kumareswaran; Robert D Schaller
Journal:  Cureus       Date:  2021-02-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.