Literature DB >> 30695100

Trajectory analysis for field free line magnetic particle imaging.

Can Barış Top1, Alper Güngör1, Serhat Ilbey1, H Emre Güven1.   

Abstract

PURPOSE: Magnetic particle imaging (MPI) is a relatively new method to image the spatial distribution of magnetic nanoparticle (MNP) tracers administered to the body with high spatial and temporal resolution using an inhomogeneous magnetic field. The spatial information of the MNP's is encoded using a field free point (FFP), or a field free line (FFL), in which the magnetic field vanishes at a point, or on a line, respectively. FFL scanning has the advantage of improved sensitivity compared to FFP scanning as a result of higher signal-to-noise ratio. The trajectory traversed by the FFL or FFP is an important parameter of the MPI system and should be selected to achieve the best imaging quality in minimum scan time, while considering hardware constraints and patient safety. In this study, we analyzed the image quality of different FFL trajectories for a large field of view (FOV) using simulations, to provide a baseline information for FFL scanning MPI system design.
METHODS: We simulated a human-sized FFL scanning MPI configuration to image a circular FOV with 160 mm diameter, and compared Radial, Spiral, Uniform Spiral, Flower, and Lissajous trajectories with different trajectory densities scanned by the FFL for constant scan time. We analyzed the system matrices of the trajectories in terms of mutual coherence and homogeneity of the spatial sensitivity. We calculated the maximum electric fields induced on a homogeneous conductive body by the selection field (SF) and the focus field (FF) to compare the trajectories based on the nerve stimulation threshold. The images were obtained using the system matrix reconstruction approach with two different image reconstruction methods. In the first one, we used the conventional image reconstruction method, algebraic reconstruction technique (ART), which gives a regularized least-squares solution. In the second one, we used the state-of-the-art alternating direction method of multipliers (ADMM), which minimizes a weighted sum of the l1 -norm and the total variation (TV) of the images.
RESULTS: The Radial and Spiral trajectories resulted in a poor imaging performance at low trajectory densities due to relatively high coherency and poor sensitivity of the measurements, respectively. For ART reconstruction, the highest image quality with the lowest trajectory density was achieved with the Uniform Spiral trajectory. Uniform Spiral, Flower, and Lissajous trajectories yielded comparable performance with ADMM reconstruction. The rotating SF induced higher electric field amplitude compared to the FF. Consequently, maximum allowable gradient at the same trajectory density was greater for the Radial trajectory compared to the other trajectories.
CONCLUSIONS: For a large FOV coverage, the Uniform Spiral trajectory offers a good compromise between image quality and imaging time, taking safety and hardware limitations into account. The Radial trajectory, especially using l1 -norm and TV priors in the reconstruction, may be favorable in case the SF induced electric field is higher than that of the FF at the same frequency (e.g., relatively small FOV coverage). In general, ADMM reconstruction resulted in higher contrast and resolution compared to ART, leading to lighter requirements on the density of the trajectory.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  field free line; image reconstruction; magnetic nanoparticles; magnetic particle imaging; open MPI; peripheral nerve stimulation

Mesh:

Substances:

Year:  2019        PMID: 30695100     DOI: 10.1002/mp.13411

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

Review 1.  The Reconstruction of Magnetic Particle Imaging: Current Approaches Based on the System Matrix.

Authors:  Xiaojun Chen; Zhenqi Jiang; Xiao Han; Xiaolin Wang; Xiaoying Tang
Journal:  Diagnostics (Basel)       Date:  2021-04-26

2.  Application of Linear Gradient Magnetic Field in Arterial Profile Scanning Imaging.

Authors:  Yanjun Liu; Guoqiang Liu; Dan Yang; Bin Xu
Journal:  Sensors (Basel)       Date:  2020-08-13       Impact factor: 3.576

Review 3.  Recent developments of the reconstruction in magnetic particle imaging.

Authors:  Lin Yin; Wei Li; Yang Du; Kun Wang; Zhenyu Liu; Hui Hui; Jie Tian
Journal:  Vis Comput Ind Biomed Art       Date:  2022-10-01
  3 in total