Finite element analysis of peri-implant bone volume affected by stresses around Morse taper implants: effects of implant positioning to the bone crest.

J Paulo Macedo1,2, Jorge Pereira1,2, João Faria3, Júlio C M Souza3, J Luis Alves3, José López-López2, Bruno Henriques3,4.   

Abstract

OBJECTIVES: The purpose of the present study was to evaluate the distribution and magnitude of stresses through the bone tissue surrounding Morse taper dental implants at different positioning relative to the bone crest.
MATERIALS AND METHODS: A mandibular bone model was obtained from a computed tomography scan. A three-dimensional (3D) model of Morse taper implant-abutment systems placed at the bone crest (equicrestal) and 2 mm bellow the bone crest (subcrestal) were assessed by finite element analysis (FEA). FEA was carried out on axial and oblique (45°) loading at 150 N relatively to the central axis of the implant. The von Mises stresses were analysed considering magnitude and volume of affected peri-implant bone.
RESULTS: On vertical loading, maximum von Mises stresses were recorded at 6-7 MPa for trabecular bone while values ranging from 73 up to 118 MPa were recorded for cortical bone. On oblique loading at the equiquestral or subcrestal positioning, the maximum von Mises stresses ranged from 15 to 21 MPa for trabecular bone while values at 150 MPa were recorded for the cortical bone. On vertical loading, >99.9vol.% cortical bone volume was subjected to a maximum of 2 MPa while von Mises stress values at 15 MPa were recorded for trabecular bone. On oblique loading, >99.9vol.% trabecular bone volume was subjected to maximum stress values at 5 MPa, while von Mises stress values at 35 MPa were recorded for >99.4vol.% cortical bone.
CONCLUSIONS: Bone volume-based stress analysis revealed that most of the bone volume (>99% by vol) was subjected to significantly lower stress values around Morse taper implants placed at equicrestal or subcrestal positioning. Such analysis is commentary to the ordinary biomechanical assessment of dental implants concerning the stress distribution through peri-implant sites.

Entities:  

Keywords:  Bone resorption; FEA; Morse taper implant; implant design; marginal bone loss; placement depth

Mesh:

Substances:

Year:  2018        PMID: 30693810     DOI: 10.1080/10255842.2018.1507025

Source DB:  PubMed          Journal:  Comput Methods Biomech Biomed Engin        ISSN: 1025-5842            Impact factor:   1.763


  3 in total

1.  Radiological implications of crestal and subcrestal implant placement in posterior areas. A cone-beam computed tomography study.

Authors:  Hilario Pellicer-Chover; Julio Rojo-Sanchís; Miguel Peñarrocha-Diago; José Viña-Almunia; David Peñarrocha-Oltra; Maria Peñarrocha-Diago
Journal:  J Clin Exp Dent       Date:  2020-09-01

2.  Biomechanical Evaluation of Stress Distribution in Equicrestal and Sub-crestally Placed, Platform-Switched Morse Taper Dental Implants in D3 Bone: Finite Element Analysis.

Authors:  Yashaswini Ellendula; Anam Chandra Sekar; Sandeep Nalla; Ram B Basany; Kunchala Sailasri; Ashwini Thandu
Journal:  Cureus       Date:  2022-04-29

3.  New Dental Implant with 3D Shock Absorbers and Tooth-Like Mobility-Prototype Development, Finite Element Analysis (FEA), and Mechanical Testing.

Authors:  Avram Manea; Grigore Baciut; Mihaela Baciut; Dumitru Pop; Dan Sorin Comsa; Ovidiu Buiga; Veronica Trombitas; Horatiu Colosi; Ileana Mitre; Roxana Bordea; Marius Manole; Manuela Lenghel; Simion Bran; Florin Onisor
Journal:  Materials (Basel)       Date:  2019-10-21       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.