Minimization of the Settling Time of Variable Area Flowmeters.

Mateusz Turkowski1, Artur Szczecki2, Maciej Szudarek3.   

Abstract

In the article a differential equation describing transient behavior of variable area (VA) meters has been developed and validated experimentally for air as a measured fluid and for two float shapes-plumb bob and sphere. A modified version of simplex algorithm adapted for nonlinear constraint optimization problems was applied to minimize the settling time of VA meters in two cases. In the first case both the float and tube geometry were altered. In the second case only the float geometry was modified. The second case has been validated experimentally. The theory and experiment is in reasonable agreement (under 5% of full scale), which is satisfactory for the purposes of optimization of VA flowmeters dynamic performance. Analytical model of VA flowmeter has been proven to be a proper tool for optimization. Settling times obtained during the optimization process were several times shorter than these of commercially manufactured instruments. Overshoot has not exceeded the assumed value of 3%.

Entities:  

Keywords:  flow sensors; industry 4.0; modelling; optimization; rotameters; variable area flowmeters

Year:  2019        PMID: 30691238      PMCID: PMC6386847          DOI: 10.3390/s19030530

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


1. Introduction

Over the recent decades various versions of variable area (VA) flowmeters have grown to maturity. E.g., classic “see through” rotameters, orifice VA meters or VA flow transducers. Both the calibration procedures [1,2] and corrections for density and viscosity [3,4] have been established for a long time. Many float types have been thoroughly investigated [5]. The simplest indicating device is a transparent tube. More sophisticated flow transducers use a magnetically inert metal tube. In this case the float is provided with a permanent magnet and float position is sensed by an external magnet. Float displacement can then be converted to any standardized signal. Simple solution consists of a linear variable differential transformer [6]. Another possibility is the application of Hall sensors [7,8], magnetoresistive sensors matrix parallel to the tube [9] or capacitance transmitters [10]. In this case transducer does not exert forces on the float and therefore does not influence the VA meter dynamics. This simple and economical flowmeter continues to be one of the most widely used instruments for low and medium flow rates in numerous industry branches [11,12]. Industry 4.0, which is the current trend of automation and data exchange in manufacturing technologies including cyber-physical systems and Internet of Things, favors measurement sensors which have good dynamic performance [13], e.g., Coriolis flowmeters. However, price of Coriolis flowmeters is one or two orders of magnitude higher than price of simple VA flow transmitters. Study on dynamic performance of VA meters would allow optimization and development of simple and rugged flow transmitters that can meet contemporary needs of process automation. Up till now research on dynamics of variable area flowmeters has concerned their behavior in presence of pulsations [14,15,16]. This aspect is important in case of such pulsation sources as reciprocating pumps or other sources classified in Ref. [17]. However, for control purposes such parameters as settling time, nature of step response (periodic or aperiodic) and overshoot are vital. These subjects have not yet been studied in the context of VA meters. The paper deals with these aspects and presents optimization procedures. An analytical model describing VA meter dynamics is developed and validated. The methodology bases on registering float position by a camera. This is the only method that does not exert forces on the float. The proposed equation has been proven to work for air and for two float shapes—plumb bob and sphere. Further research is required to validate the developed equation for other fluids and other float shapes. Computational fluid dynamics may be a viable tool for this purpose. In that case simulation validation criteria described in Ref. [18] may be applied. After validation of the transient equation the article present results of optimization for two cases. In the first case both the tube and float geometry were altered, in the second case only the float geometry was modified. The second case was validated experimentally and the presented methodology has been proven to be of scientific and practical significance.

2. Materials and Methods

2.1. VA Flowmeter Characteristics in Steady State

Variable area flowmeter consists of a tapered tube and a float. The float can take various forms. Examples are traditional plumb bob (as in Figure 1a), sphere or flat disc. Let us consider two perpendicular sections of the flowmeter marked in the Figure 1b. For stationary, incompressible flow one can use Bernoulli’s law to write: where ρ is the fluid density and p1, U1, p2, U2 are respectively pressure and mean velocities in Section 2.1. and Section 2.2.
Figure 1

(a) Simplified geometry of a variable area flowmeter and the streamlines across it (not to scale); (b) Section 2.1. is where the float does not yet influence the flow. Section 2.2. is in the narrowest annular stream section, “vena contracta”. All of the marked heights are measured from the zero level. The zero level is the height at which diameter of the tube d is equal to the float diameter d

The area of the narrowest part of the stream A2 equals: where A = π/4(d2 − d and Cc is the contraction coefficient. As the velocity U1 is much smaller than U2, its square can be neglected. Therefore: this pressure difference multiplied by the area of the float A = πd2/4 results in active force directed upwards: The restoring force F is equal to the apparent weight of the float of volume V and density ρ immersed in the fluid of density ρ: where g—acceleration due to gravity. In the equilibrium state active force equals restoring force, therefore one can write: There are various methods to obtain Cc coefficient. Procedures of VA meter calculations based on d:d diameter ratio and Ruppel number are shown in Ref. [2]. Another option is to determine Cc during calibration, as proposed by Ref. [1]. In this way a flow coefficient C(Re) is obtained, instead of the contraction coefficient Cc. Although d:d is an important factor, it is taken here into account indirectly, through hydraulic diameter in Re. The flow coefficient C(Re) includes not only the effect of contraction, but also viscous friction, dynamic drag and other effects. In this case Re is defined as: where (d − d) is hydraulic diameter, U = Q/A(h)—mean velocity in the annular slot between float and tube and ν—kinematic viscosity of the fluid.

2.2. VA Behavior During Transient

During transient the active force is proportional to the velocity relative to the float. Therefore, float velocity dh/dt must be subtracted from velocity U2 given by Equation (3). What is more, moving float acts as a piston pump. It results in an additional volumetric flow rate Q′ equal to: The resultant velocity equals: taking the above into the account, mean velocity relative to the float in the Section 2.2. is: the flow rate in the Section 2.2. equals: and the active force during transient is then: taking into account dynamics, the restoring force equals: as Q can be in general a function of time Q(t), the equation of motion of the float will be therefore: It is easy to validate that in steady state (h = const.) this equation takes the form of static Equation (7). The relationship between the area A and the position of the float in the tube h (which is in fact the measure of the flow rate Q) is as follows: during transient Reynolds number equals: where: which allows to obtain the final equation: The differential Equation (19) is nonlinear and it can be only solved numerically. Runge-Kutta algorithm has been used for this purpose. There was a question of approximating function C(Re), which has been obtained experimentally. After numerous attempts with polynomial, exponential and logarithmic curves the best fit was obtained using approximating function in form of: where D and E coefficients were determined with the use of least squares method. As all the experiments were conducted with the use of atmospheric air as working medium, buoyancy forces could have been neglected. Therefore:

2.3. Validation of the Transient Equation

Seven VA flowmeters were investigated (Table 1). d and d are tube diameters at the heights corresponding to the lower and upper flow rate limits. Δh is the distance between these extreme float positions (see Figure 1a). The tubes for VA flowmeter no. 6 and 7 are the same as for no. 2 and 4, only the float was replaced.
Table 1

List of investigated VA flowmeters and their most important parameters.

VA no.Nominal Range dm3/hFloat Shapedf mmdlow mmdhigh mmΔh mmmf g D E
13–30Plumb bob3.9904.0804.3102360.0530.23690.0815
220–220Plumb bob4.0304.1304.5602230.3810.24850.0388
3100–800Plumb bob8.9109.30710.3602160.5950.21320.0502
4200–2200Plumb bob9.0209.23710.3542154.0550.17890.0929
5500–3000Plumb bob11.03011.2212.4852367.2980.14440.297
625–230Sphere3.9724.1304.5602232.540.27840.0447
7550–2400Sphere8.7319.23710.3542152.790.16800.1654
The experiments were conducted with a methodology fully developed by the authors. The step response was filmed with an analog 8 mm camera with frame rate of 50 frames per second (Figure 2a). The position of the float was later measured on the screen of a film previewer. Step change in the flow rate was generated by a fast opening of a spring operated valve.
Figure 2

(a) Test stand for registration of the step response of VA flowmeter. 1—fast opening valve, 2—flow control valve, 3—constant temperature anemometer with data acquisition system, 4—VA flowmeter, 5—camera; (b) test stand for settling time measurement. 1—fast opening valve, 2—flow control valve, 3—VA flowmeter, 4—photoelectric sensors. Photoelectric sensors triggered a timer, enabling the measurement of the settling time.

The step was not ideal, therefore it has been registered by means of hot wire anemometer [19] and later approximated by a function in form of: where Q is the flow rate in steady state. This form of excitation signal was used during the solution of the Equation (21). Exemplary signal registered by thermoanemometer and its approximation is shown in the Figure 3. Test stand for settling time measurement is presented in the Figure 2b. The movement of the float at the final phase was slow. For that reason the overshoot could have been easily assessed by visual observation.
Figure 3

Step change registered by hot wire anemometer (solid line) and its approximation (dashed line).

2.4. Optimization

For a given flow rate range and measured fluid there are three parameters that define remaining parameters of a given VA flowmeter unambiguously. Examples are d, d, Δh or m, d, Δh. A set of parameters d, (d − d Δh was chosen for the reason of simplicity of formulae to calculate the remaining parameters. A modified simplex algorithm described in the Appendix A was used for optimization. Considering the parameters of VA flowmeters manufactured commercially, constraints presented in the Table 2 were assumed.
Table 2

List of constraint functions for optimization problem.

No.Lower Bound liParameterUpper Bound ui
14 mm df 25 mm
20.05 mmdlowdf1 mm
3150 mmΔh400 mm
40.2 g/cm3m/(0.697d3f)11 g/cm3
The first three constraints are boundary constraints. As the density of float ρ does not occur explicitly in the Equation (21), the 4th constraint is referred to as inequality constraint. Volume of the float of a plumb bob type is expressed as V = 0.679d3. To determine values of coefficients D and E of the function approximating C(Re), all the experimental data concerning VA meters from Table 1 was collected. Function in a form of: was determined with the use of least squares method and was used during optimization. From the point of view of automation systems, minimum settling time and an aperiodic response (possibly with one small overshoot) is advantageous. The objective function was therefore defined as in Figure 4. Optimized VA meters should have had a minimal settling time t0.97. The overshoot should have not exceeded 3% of the float position in a steady state h.
Figure 4

Definition of the objective function (graph not in a scale).

For the each set of chosen VA parameters generated randomly by optimization algorithm described in the appendix, the values of an objective function were obtained by integration of the Equation (21). After every step of integration (ca. 0.01 s) an analysis of the step response was performed. If the float height exceeded the value of 1.03h (point a) the integration was interrupted as not fulfilling the assumptions. The results were then rejected and a new integration process was started, based on a new set of parameters generated by the optimization algorithm. If the response did not exceed the 1.03h value and in subsequent integration step the calculated h value started to decrease (point b), it was no longer possible that the float went beyond the ±0.03h limits. The value t0.97 calculated as an abscissa of the intersection of the line connecting points (t, h) and (t, h) and the horizontal line h = 0.97h was stored for further processing by the algorithm. To save time, if after some arbitrarily chosen time the response did not reach 0.97h, the integration was interrupted (point c) and results were rejected. The optimization procedure stopped when five optimization results (objective function values) remained within a specified error band.

3. Results and Discussions

3.1. Validation of the Transient Equation

The comparison of experiments with the results obtained from the Equation (21) are presented in Figure 5, Figure 6, Figure 7, Figure 8, Figure 9 and Figure 10. One can distinguish between three typical step responses. For VA no. 1 an aperiodic response was obtained (Figure 5 and Figure 6). It resembles a step response of a first order linear system with damping coefficient above 1. The settling time rises with the decreasing value of step change, which is in contrast to a typical linear system.
Figure 5

Rotameter no. 1, plumb bob. Float position h vs. time t for a step response, experimental (black solid lines) and calculated (markers).

Figure 6

Rotameter no. 1, step 10 dm3/h. Selected images captured by a camera.

Figure 7

Rotameter no. 5, plumb bob. Float position h vs. time t for a step response, experimental (black solid lines) and calculated (markers).

Figure 8

Rotameter no. 3, plumb bob. Float position h vs. time t for a step response, experimental (black solid lines) and calculated (markers).

Figure 9

Rotameter no. 7, sphere shape. Float position h vs. time t for a step response, experimental (black solid lines) and calculated (markers).

Figure 10

Rotameter no. 7. Selected images of a step response captured by a camera.

For VA no. 5 results resemble a step response of a second order linear system with a small value of damping coefficient (Figure 7). As opposed to the linear system, the oscillation frequency is higher for a lower step value. Moreover, oscillation fragments below steady-state value are shorter than the upper parts. The response of VA no. 3 resembles a step response of a linear system with damping coefficient equal to ca. 0.5, practically with one overshoot (Figure 8). In contrast to the linear system, the overshoot decreases with a decreasing step value. The response of VA no. 7 of a sphere shape (Figure 9 and Figure 10) is similar to the response of VA no. 5, but with a smaller value of damping coefficient. Uncertainty analysis is shown in the Appendix B. The sum of the experimental and theoretical uncertainties is in the worst case equal to U(h) = 18.2 mm whereas the differences between experiments and simulations were up to 11 mm. It was assumed that the mathematical model was in a satisfactory agreement with the experiments.

3.2. Optimization of VA Flowmeters

Three VA meters of a plumb bob type were optimized for nominal ranges listed in the Table 3. In studied cases the pipe length Δh was equal to the lower bound l = 150 mm and the overshoot was equal to 3% of h value. In comparison to the commercial VA meters, resultant float density was relatively low and the clearance between a float and a tube was relatively high. The settling time (0.32–0.35) s was shorter than for commercial VA meters (see Figure 5, Figure 6, Figure 7, Figure 8, Figure 9 and Figure 10 where it equals (0.9–1.5) s).
Table 3

Tube and float optimization results.

VA no.Nominal Range in dm3/hFloat Shapedf in mmdlow in mmTube TaperingΔh in mmρf in g/cm3t0.97 in s
110–100Plumb bob4.5174.8180.00311500.4370.349
250–500Plumb bob10.2110.580.00661500.2580.336
3200–2000Plumb bob20.5320.880.00881500.3550.326
Unfortunately, unit production of glass tubes for experiments is costly. Therefore the optimization was repeated for the existing tubes specified in the Table 1, and was limited to modifying float parameters. In this case optimization algorithm generated the lower range limit Q and the clearance (d − d). The remaining parameters such as upper range limit, float mass and density were then easily calculated. Additional restriction was added concerning the rangeability, namely Q/Q ≥ 8. Optimization results are presented in the Table 4. Columns 4 and 6 contain values of the clearance (d − d) and Q generated by the optimization algorithm. The calculated values of ρ and Q are presented in columns 5 and 7. Theoretical values of the settling time (column 8) are within the range of (0.341–0.372) s, which is also an improvement in comparison to the commercial meters. The experimental results are presented in columns 9–12. There is a good agreement between results of optimization and experiments. Measured settling times are equal or slightly shorter than theoretical values. The measured overshoot has never exceeded the assumed value of 3%.
Table 4

Results of float optimization for tubes specified in the Table 1.

No.Δh in mmdlow in mmdlowdf in mmρf in g/cm3Theoretical ValuesExperimental Values
Qmin in dm3/hQmax in dm3/ht0.97 in sQmin in dm3/h Qmax in dm3/ht0.97 in sOvershoot in %hss
123456789101112
12364.0800.0914.0987.5730.3728.2620.352.0
22234.1300.1061.2352.6680.3413.2640.311.9
32169.3070.3390.299373610.341463840.322.1
42159.2370.3810.252383520.345373610.332.3
523611.220.4020.252605540.356725660.361.1

4. Conclusions

Analytical model of VA flowmeters dynamics has been developed and validated experimentally. It has been proven to be a proper tool for optimization. This will make VA meters more appropriate for process automation, which is a prerequisite for the implementation of Industry 4.0 solutions. Settling times obtained during the optimization process were several times shorter than these of commercially manufactured instruments: (0.341–0.372) s instead of (0.9–1.5) s. This proves that the presented methodology is of scientific and practical significance. What is more, it has been proven that the developed equation describing VA meter dynamics is appropriate for at least two float shapes, i.e., plumb bob and sphere. The sum of the experimental and theoretical uncertainties is in the worst case 18.2 mm whereas the differences between experiments and simulations were up to 11 mm. The discrepancies can also be attributed to the numerical modeling of the system (e.g., inaccurate physical models, discretization and round-off errors) and the fact that the uncertainty analysis has been developed for the steady state case, not for transient. What is more, the approximating function C(Re) that takes float shape into account has been obtained as the best fit of experimental data. Comparing the Figure 9 with Figure 5, Figure 7and Figure 8, it can be seen that the differences between modelling and experiments are smaller for a simple sphere float shape. However, the theory and experiment are in reasonable agreement (under 5% of full scale), satisfactory for the purposes of optimization of VA flowmeters dynamic performance. In generally, the discrepancies between the experiments and calculations are smaller for larger rotameters. Further research should be conducted for various float shapes and various fluids as a measured medium. This includes both classical rotameters and variable orifice meters. The influence of forces from transmitter acting on a float should also be studied. It would also be beneficial to employ computational fluid dynamics for further optimization.
Table A1

Calculation of the sensitivity coefficients c for VA meter (3–30) dm3/h. The values in bold denote parameters that are incremented in a given row.

xi Δxi df in mmdhigh in mmdlow in mmΔh in mmm in g∙10−2ρ in kg/m3υ in m2/s·10−5Q in dm3/h C h increment in m ci
df 0.01 4.00 4.314.082365.381.2931.450.010.29050.000140.014
dh 0.013.99 4.32 4.082365.381.2931.450.010.2905−0.00791−0.791
dl 0.013.994.31 4.09 2365.381.2931.450.010.29050.008760.876
Δh13.994.314.08 237 5.381.2931.450.010.29050.00080.0008
m 0.013.994.314.08236 5.39 1.2931.450.010.2905−0.00172−0.172
ρ 0.013.994.314.082365.38 1.303 1.450.010.29050.000720.072
ν 0.013.994.314.082365.381.293 1.46 0.010.29050.000750.075
Q 0.13.994.314.082365.381.2931.45 10.1 0.29050.00797.9
C 0.013.994.314.082365.381.2931.450.01 0.3005 −0.00629−0.629
Table A2

Uncertainty budget for VA meter (3–30) dm3/h.

Input Quantity xiNominal ValueStandard Uncertainty u(xi)UnitSensitivity Coefficient ciUnit(ci·u(xi))2 in m2
df 3.99 × 10−30.0075mm0.014m/mm1.1 × 10−8
dhigh 4.31 × 10−30.0075m−0.791m/mm3.52 × 10−5
dlow 4.08 × 10−30.0075m0.876m/mm4.32 × 10−5
Δh236 × 10−30.5m0.0008m/mm1.6 × 10−7
m 5.38 × 10−30.001kg−0.172m/kg2.96 × 10−8
ρ 1.2930.0025kg/m30.072m/(kg/m3)3.24 × 10−8
ν 1.250.02m2/s × 10−5 0.075m/(m2/s × 10−5)2.25 × 10−12
Q 0.010.0000125m3/h7.9m/(m3/h)9.75 × 10−9
C 0.29050.0025-−0.629m2.47 × 10−6
Σ(ciu(xi))2 m28.11 × 10−5
  3 in total

1.  Array of Hall Effect sensors for linear positioning of a magnet independently of its strength variation. a case study: monitoring milk yield during milking in goats.

Authors:  Fernando-Juan García-Diego; Angel Sánchez-Quinche; Paloma Merello; Pedro Beltrán; Cristófol Peris
Journal:  Sensors (Basel)       Date:  2013-06-21       Impact factor: 3.576

2.  A Mathematical Model of the Thermo-Anemometric Flowmeter.

Authors:  Igor Korobiichuk; Olena Bezvesilna; Andriі Ilchenko; Valentina Shadura; Michał Nowicki; Roman Szewczyk
Journal:  Sensors (Basel)       Date:  2015-09-11       Impact factor: 3.576

3.  Model and Simulation Studies of the Method for Optimization of Dynamic Properties of Tachometric Anemometers.

Authors:  Paweł Ligęza
Journal:  Sensors (Basel)       Date:  2018-08-14       Impact factor: 3.576