| Literature DB >> 30687718 |
Stepan Havranek1, Hana Alfredova1, Zdenka Fingrova1, Lucie Souckova1, Dan Wichterle1,2.
Abstract
Purpose: The acute effect of radiofrequency (RF) ablation includes local necrosis and oedema. We investigated the spatiotemporal change of atrial electrograms in the area surrounding the site of single standardized pulse of RF energy.Entities:
Keywords: atrial fibrillation; electrogram voltage; oedema; pulmonary vein isolation; radiofrequency catheter ablation
Year: 2019 PMID: 30687718 PMCID: PMC6338051 DOI: 10.3389/fcvm.2018.00190
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1(A) An example of high-density bipolar voltage remap (low-color and high fill-threshold setting) at baseline. The site of index ablation is depicted by purple tag (posteroanterior view). PV antra are tagged by white points. (B) Corresponding layout of 12 segments used for electrogram analysis. Ablation site belongs to segment #1.
Baseline demographical, clinical, and procedure characteristics.
| Males | 10 (83%) |
| Age (years) | 61 (45–67) |
| Arterial hypertension | 8 (67%) |
| Diabetes mellitus | 1 (8.3%) |
| Coronary artery disease | 1 (8.3%) |
| Indexed LA volume estimated by ECHO (ml/m2) | 36 (32–36) |
| Left ventricular ejection fraction (%) | 61 (62–65) |
| LA volume estimated by 3D electro-anatomical mapping (ml) | 118 (102–144) |
| Amiodarone | 1 (8.3%) |
| Mapping points—dense map # 1 | 113 (54–167) |
| Mapping points—dense map # 2 | 114 (69–160) |
| Mapping points—dense map # 3 | 151 (110–191) |
Values are in format median (interquartile range) or n (%). LA, left atrium.
Absolute values of bipolar atrial electrogram voltages in individual segments at baseline and two remapping periods.
| 1 | 2.33 ± 1.09 | 1.36 ± 0.66 | 1.08 ± 0.57 |
| 11 | 1.99 ± 1.12 | 0.87 ± 0.40 | 0.93 ± 0.46 |
| 12 | 2.79 ± 1.13 | 2.10 ± 1.19 | 1.74 ± 0.91 |
| 13 | 3.14 ± 1.46 | 2.49 ± 1.13 | 2.02 ± 1.00 |
| 14 | 1.95 ± 0.85 | 1.04 ± 0.46 | 0.91 ± 0.50 |
| 111 | 1.34 ± 0.89 | 0.91 ± 0.66 | 0.90 ± 0.73 |
| 112 | 1.83 ± 1.01 | 1.18 ± 0.78 | 1.30 ± 0.90 |
| 121 | 2.29 ± 1.46 | 1.95 ± 1.02 | 2.20 ± 0.80 |
| 122 | 3.50 ± 1.75 | 3.17 ± 1.57 | 2.94 ± 1.56 |
| 131 | 3.26 ± 1.61 | 3.77 ± 1.53 | 3.04 ± 1.58 |
| 132 | 3.01 ± 1.26 | 3.06 ± 1.31 | 2.36 ± 1.10 |
| 141 | 2.15 ± 0.91 | 1.50 ± 0.79 | 1.35 ± 0.69 |
| 142 | 1.60 ± 0.94 | 0.92 ± 0.51 | 0.92 ± 0.54 |
Values are in format mean ± SD. Localization of segment: see Figure .
Figure 2An example of changed bipolar atrial electrogram voltage after ablation.
Relative change in bipolar atrial electrogram voltages in individual segments between consecutive maps.
| 1 | 0.66 ± 0.24 | 0.0008 | 0.60 ± 0.38 | 0.006 | 0.86 ± 0.35 | 0.21 |
| 11 | 0.51 ± 0.26 | 0.0003 | 0.54 ± 0.23 | 0.0002 | 1.09 ± 0.26 | 0.25 |
| 12 | 0.71 ± 0.27 | 0.005 | 0.64 ± 0.26 | 0.001 | 1.10 ± 1.04 | 0.75 |
| 13 | 0.84 ± 0.36 | 0.16 | 0.70 ± 0.33 | 0.009 | 0.83 ± 0.18 | 0.007 |
| 14 | 0.62 ± 0.28 | 0.001 | 0.52 ± 0.24 | 0.00005 | 0.87 ± 0.33 | 0.21 |
| 111 | 0.60 ± 0.20 | 0.0001 | 0.70 ± 0.39 | 0.04 | 1.19 ± 0.69 | 0.39 |
| 112 | 0.71 ± 0.23 | 0.003 | 0.83 ± 0.32 | 0.12 | 1.32 ± 0.67 | 0.15 |
| 121 | 1.12 ± 1.05 | 0.77 | 0.73 ± 0.29 | 0.047 | 1.13 ± 0.36 | 0.35 |
| 122 | 0.86 ± 0.20 | 0.07 | 0.79 ± 0.19 | 0.01 | 0.90 ± 0.15 | 0.06 |
| 131 | 1.24 ± 0.32 | 0.045 | 0.92 ± 0.39 | 0.49 | 0.76 ± 0.27 | 0.01 |
| 132 | 1.05 ± 0.23 | 0.46 | 0.85 ± 0.31 | 0.13 | 0.81 ± 0.25 | 0.03 |
| 141 | 0.78 ± 0.59 | 0.25 | 0.61 ± 0.22 | 0.0001 | 0.95 ± 0.33 | 0.58 |
| 142 | 0.61 ± 0.27 | 0.0007 | 0.60 ± 0.25 | 0.0002 | 0.98 ± 0.34 | 0.84 |
Values are in format mean ± SD. p.
Figure 3Relative change of bipolar atrial electrogram voltages: < 5 min after ablation vs. baseline (left panel) and >30 min after ablation vs. baseline (right panel). Color scale represents the intensity of relative change. The central purple circle corresponds to ablation core. White dots annotate centres of each analyzed segment. See Figure 1 for segment layout. The central purple tag corresponds to ablation site. Ablation catheter is displayed in proportion to the figure scale in millimeters.
Figure 4Relative change of bipolar atrial electrogram voltages (>30 min vs. < 5 min after the ablation). Color scale represents the relative change.