| Literature DB >> 30675932 |
Eric C Spivey1,2, Josiah C McMillen1,3, Daniel J Ryan1,3, Jeffrey M Spraggins1,3,4, Richard M Caprioli1,3,4,5,6.
Abstract
A transmission geometry optical configuration allows for smaller laser spot size to facilitate high-resolution matrix-assisted laser/desorption ionization (MALDI) mass spectrometry. This increase in spatial resolution (ie, smaller laser spot size) is often associated with a decrease in analyte signal. MALDI-2 is a post-ionization technique, which irradiates ions and neutrals generated in the initial MALDI plume with a second orthogonal laser pulse, and has been shown to improve sensitivity. Herein, we have modified a commercial Orbitrap mass spectrometer to incorporate a transmission geometry MALDI source with MALDI-2 capabilities to improve sensitivity at higher spatial resolutions.Entities:
Keywords: MALDI; MALDI-2; high spatial resolution; transmission geometry
Year: 2019 PMID: 30675932 DOI: 10.1002/jms.4335
Source DB: PubMed Journal: J Mass Spectrom ISSN: 1076-5174 Impact factor: 1.982