Literature DB >> 30665177

Audiological outcomes in patients with congenital aural atresia implanted with transcutaneous active bone conduction hearing implant.

M E Zernotti1, M M Chiaraviglio2, S B Mauricio2, P A Tabernero3, M Zernotti3, M F Di Gregorio3.   

Abstract

OBJECTIVES: The objective of this study is to evaluate the safety and efficacy of the transcutaneous Bone Conduction Implant, the Bonebridge, in patients with congenital aural atresia.
METHODS: Audiometry, speech recognition test and free field audiometry were performed. Word recognition scores and speech perception was evaluated using Spanish phonetically-balanced disyllables word list.
RESULTS: Fourteen subjects were implanted with the Bonebridge (seven bilateral placements). The study cohort comprised seven males and seven females aged from 3 to 17 years (mean age 9.76yrs). All patients accepted and benefited from the implanted Bonebridge system. The pre-operative PTA4 was 66.4 dB (64.2-68.6, 95-%CI) and improved after activation to 19.2 dB (16.9-21.5, 95%CI), resulting in a mean functional gain of 47,2 dB. Regarding speech discrimination, the pre-operative outcomes of the disyllabic measurements were 34.3% and for monosyllables 27.4%. Following activation the speech discrimination improved to 98.6% and 97.9%, respectively. No infections or adverse device related effects occurred in patient group.
CONCLUSION: We have concluded that the Bonebridge implant is an innovative solution for patients with conductive or mixed hearing loss and unilateral loss suffering from congenital atresia. Different surgical techniques may be used for implant placement, based on the patient's anatomy. Studies show improved functional gain, better speech perception, and lower rates of percutaneous complications associated with this implant.
Copyright © 2019 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Atresiaplasty; BONEBRIDGE; Congenital aural atresia; Prevalence of atresia; Pure conductive hearing loss; Transcutaneous active bone conduction hearing implant

Mesh:

Year:  2019        PMID: 30665177     DOI: 10.1016/j.ijporl.2019.01.016

Source DB:  PubMed          Journal:  Int J Pediatr Otorhinolaryngol        ISSN: 0165-5876            Impact factor:   1.675


  8 in total

1.  Complications of Transcutaneous Protheses - A Systematic Review of Publications Over the Past 10 Years.

Authors:  Marcos Antonio de Souza; Santiago Luis Vallejos Riart; Stephanie Rugeri de Souza; Rubens de Brito; Ricardo Ferreira Bento
Journal:  Int Arch Otorhinolaryngol       Date:  2022-02-04

2.  Two Bonebridge bone conduction hearing implant generations: audiological benefit and quality of hearing in children.

Authors:  Soňa Šikolová; Milan Urík; Dagmar Hošnová; Vít Kruntorád; Michal Bartoš; Oldřich Motyka; Petr Jabandžiev
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-09-08       Impact factor: 3.236

3.  Speech performance and subjective satisfaction of middle ear implant in congenital aural atresia.

Authors:  Farid Alzhrani; Salman F Alhabib; Medhat Yousef
Journal:  Acta Otorhinolaryngol Ital       Date:  2022-04       Impact factor: 2.618

4.  Patient-reported long-term benefit with an active transcutaneous bone-conduction device.

Authors:  Julia Hundertpfund; Jens Eduard Meyer; Attila Óvári
Journal:  PLoS One       Date:  2020-11-02       Impact factor: 3.240

5.  Audiological results and subjective benefit of an active transcutaneous bone-conduction device in patients with congenital aural atresia.

Authors:  Veronika Volgger; Inge Teresa Schießler; Joachim Müller; Florian Schrötzlmair; Marlene Pollotzek; John Martin Hempel
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-06-26       Impact factor: 3.236

6.  Evaluation of the Bonebridge BCI 602 active bone conductive implant in adults: efficacy and stability of audiological, surgical, and functional outcomes.

Authors:  Katarzyna B Cywka; Piotr H Skarzynski; Bartlomiej Krol; Stavros Hatzopoulos; Henryk Skarzynski
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-02-19       Impact factor: 3.236

7.  Medical, Technical and Audiological Outcomes of Hearing Rehabilitation with the Bonebridge Transcutaneous Bone-Conduction Implant: A Single-Center Experience.

Authors:  Faris F Brkic; Dominik Riss; Katharina Scheuba; Christoph Arnoldner; Wolfgang Gstöttner; Wolf-Dieter Baumgartner; Erich Vyskocil
Journal:  J Clin Med       Date:  2019-10-03       Impact factor: 4.241

8.  The Bonebridge BCI 602 Active Transcutaneous Bone Conduction Implant in Children: Objective and Subjective Benefits.

Authors:  Katarzyna B Cywka; Henryk Skarżyński; Bartłomiej Król; Piotr H Skarżyński
Journal:  J Clin Med       Date:  2021-12-16       Impact factor: 4.241

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.