Literature DB >> 30661115

Infrared photographs with a selective wavelength filter to diagnose small-angle esotropia in young children.

Yung Ju Yoo1, Hee Kyung Yang2, Jong-Mo Seo3, Kwang Gi Kim4, Jeong-Min Hwang5.   

Abstract

PURPOSE: To validate the effectiveness of the selective wavelength filter analysis with infrared photographs for diagnosing small-angle esotropia in children under age 4.
METHODS: This is a retrospective, observational, case-control study. A total of 83 esotropes with an esodeviation of ≤ 16 prism diopters (PD) and 75 orthotropic controls under 4 years of age were included. Full-face infrared photographs were taken with a selective wavelength filter in front of either eye. The angles of esodeviation on photographs were measured with the three-dimensional Strabismus Photo Analyzer. The alternate prism and cover test or the Krimsky test were repeatedly performed to measure ocular alignment.
RESULTS: The testability of infrared photographs using selective wavelength filters in children under 4 years of age was 85.6%. The mean angle of esodeviation was 11.3 ± 4.0 PD by manual measurements and 11.5 ± 4.4 PD by the infrared photograph analysis. Manual measurements and the infrared photograph analysis showed a strong positive correlation (R = 0.815, P < 0.001). The sensitivity and specificity of the infrared photograph analysis for detecting small-angle esotropia were 95.2% and 77.9%, respectively, with a cutoff value of 4.0 PD.
CONCLUSIONS: The automated infrared photograph analysis was simple and effective for diagnosing small-angle esotropia in young children.

Entities:  

Keywords:  Alignment; Esotropia; Image analysis; Selective wavelength filter

Mesh:

Year:  2019        PMID: 30661115     DOI: 10.1007/s00417-019-04240-2

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  3 in total

1.  Morphometry of the Oculomotor Nerve in Duane's Retraction Syndrome.

Authors:  Min Seok Kang; Hee Kyung Yang; Jounghan Kim; Jae Hyoung Kim; Jeong-Min Hwang
Journal:  J Clin Med       Date:  2020-06-24       Impact factor: 4.241

2.  An artificial intelligence platform for the diagnosis and surgical planning of strabismus using corneal light-reflection photos.

Authors:  Keli Mao; Yahan Yang; Chong Guo; Yi Zhu; Chuan Chen; Jingchang Chen; Li Liu; Lifei Chen; Zijun Mo; Bingsen Lin; Xinliang Zhang; Sijin Li; Xiaoming Lin; Haotian Lin
Journal:  Ann Transl Med       Date:  2021-03

3.  The reliability of the angle of deviation measurement from the Photo-Hirschberg tests and Krimsky tests.

Authors:  S Tengtrisorn; A Tungsattayathitthan; S Na Phatthalung; P Singha; N Rattanalert; S Bhurachokviwat; S Chouyjan
Journal:  PLoS One       Date:  2021-12-01       Impact factor: 3.240

  3 in total