Literature DB >> 30640309

Potential Deviation Factors Affecting Stereolithographic Surgical Guides: A Systematic Review.

Diego Antonio Sigcho López1, Iván García2, Gustavo Da Silva Salomao3, Dalva Cruz Laganá4.   

Abstract

OBJECTIVE: The aim of this study was to review potential deviation factors in stereolithographic surgical guides for dental implantology, warnings, and limitations of the system.
METHODS: An electronic search in databases EMBASE, the Cochrane Library, and PubMed were conducted by 3 researchers to collect information on the accuracy of static computer-guided implant placement to summarize and analyze the overall accuracy. The latter included a search for correlations between factors such as support (teeth/mucosa/bone), number of templates, use of fixation pins, jaw, template production, guiding system, and guided implant placement in articles related to guided surgery with stereolithographic static systems. Studies published between 2012 and 2017 were reviewed.
RESULTS: From 761 identified articles, a total of 24 articles were reviewed, which included 2767 dental implants. Data from studies analysis had shown a mean deviation of 3.08 degrees in angular position, 1.14 at the entry point, and 1.46 at apex position. Involved deviation factors were related to planning, laboratory, and surgical phases.
CONCLUSION: Guided surgery may have a limited precision as technique, which surgeons need to be aware in the planning process. This review suggests some security measures in guided surgery process.

Mesh:

Substances:

Year:  2019        PMID: 30640309     DOI: 10.1097/ID.0000000000000853

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  4 in total

1.  Comparison of Different Types of Static Computer-Guided Implant Surgery in Varying Bone Inclinations.

Authors:  Pisut Thangwarawut; Pokpong Amornvit; Dinesh Rokaya; Sirichai Kiattavorncharoen
Journal:  Materials (Basel)       Date:  2022-04-20       Impact factor: 3.748

2.  Computer-assisted, template-guided immediate implant placement and loading in the mandible: a case report.

Authors:  Thomas Spielau; Uli Hauschild; Joannis Katsoulis
Journal:  BMC Oral Health       Date:  2019-04-11       Impact factor: 2.757

3.  Does the macro design of an implant affect the accuracy of template-guided implantation? A prospective clinical study.

Authors:  Sigmar Schnutenhaus; Cornelia Edelmann; Heike Rudolph
Journal:  Int J Implant Dent       Date:  2021-04-26

4.  Stereolithographic Surgical Guide with a Combination of Tooth and Bone Support: Accuracy of Guided Implant Surgery in Distal Extension Situation.

Authors:  Chia-Cheng Lin; Masahiro Ishikawa; Takeo Maida; Hsin-Chung Cheng; Keng-Liang Ou; Takashi Nezu; Kazuhiko Endo
Journal:  J Clin Med       Date:  2020-03-05       Impact factor: 4.241

  4 in total