Literature DB >> 30629299

Multicenter CT phantoms public dataset for radiomics reproducibility tests.

Petros Kalendralis1, Alberto Traverso1, Zhenwei Shi1, Ivan Zhovannik1,2, René Monshouwer2, Martijn P A Starmans3,4, Stefan Klein3,4, Elisabeth Pfaehler5, Ronald Boellaard6, Andre Dekker1, Leonard Wee1.   

Abstract

PURPOSE: The aim of this paper is to describe a public, open-access, computed tomography (CT) phantom image set acquired at three centers and collected especially for radiomics reproducibility research. The dataset is useful to test radiomic features reproducibility with respect to various parameters, such as acquisition settings, scanners, and reconstruction algorithms. ACQUISITION AND VALIDATION
METHODS: Three phantoms were scanned in three independent institutions. Images of the following phantoms were acquired: Catphan 700 and COPDGene Phantom II (Phantom Laboratory, Greenwich, NY, USA), and the Triple modality 3D Abdominal Phantom (CIRS, Norfolk, VA, USA). Data were collected at three Dutch medical centers: MAASTRO Clinic (Maastricht, NL), Radboud University Medical Center (Nijmegen, NL), and University Medical Center Groningen (Groningen, NL) with scanners from two different manufacturers Siemens Healthcare and Philips Healthcare. The following acquisition parameter were varied in the phantom scans: slice thickness, reconstruction kernels, and tube current. DATA FORMAT AND USAGE NOTES: We made the dataset publically available on the Dutch instance of "Extensible Neuroimaging Archive Toolkit-XNAT" (https://xnat.bmia.nl). The dataset is freely available and reusable with attribution (Creative Commons 3.0 license). POTENTIAL APPLICATIONS: Our goal was to provide a findable, open-access, annotated, and reusable CT phantom dataset for radiomics reproducibility studies. Reproducibility testing and harmonization are fundamental requirements for wide generalizability of radiomics-based clinical prediction models. It is highly desirable to include only reproducible features into models, to be more assured of external validity across hitherto unseen contexts. In this view, phantom data from different centers represent a valuable source of information to exclude CT radiomic features that may already be unstable with respect to simplified structures and tightly controlled scan settings. The intended extension of our shared dataset is to include other modalities and phantoms with more realistic lesion simulations.
© 2019 The Authors Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2019        PMID: 30629299      PMCID: PMC6849778          DOI: 10.1002/mp.13385

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


Introduction

Computer‐aided analysis of clinical radiological images offers a data‐at‐large‐scale approach toward personalized medicine1 wherein tumor phenotype may be inferred using images of the entire tumor instead of selective sample biopsies. On the premise that phenotypic variability affects clinical outcome,2 medical imaging offers an efficient and noninvasive method to determine prognosis. This approach has immense potential to support clinical decision‐making in the personalized medicine paradigm,3 that is, which would be a superior choice of treatment for a given person. Studies in the active field of image‐derived markers (i.e., “radiomics”) strongly suggest that tomographic images do indeed embed more prognostic information than may be seen by an unassisted human eye.4, 5, 6, 7, 8 In order to be widely generalizable and have meaningful clinical use, it is essential that reproducibility of features can be tested in phantoms,9, 10 in addition to validating models in human subjects across different settings and multiple independent institutions.11, 12, 13 Studies have shown that feature reproducibility may be affected by differences in image acquisition parameters, such as slice thickness and reconstruction algorithm.14, 15, 16, 17 Since clinical image acquisition protocols are one of the major sources of variation among different hospitals, phantoms allow testing, comparison, and harmonization of radiomic features in similar vein to diagnostic imaging quality assurance. We hypothesize that even simplified phantoms allow us to test for radiomic features that may already become unstable even under tightly constrained conditions. In this data publication, we offer computed tomography (CT) scans of simple phantoms across three Dutch academic medical centers for open access. We chose to start with CT since this modality is readily available in many centers and is a workhorse imaging modality for radiotherapy intervention planning. In many clinics, CT scanners are mature technology with well‐established protocols for calibration, quality assurance, and routine maintenance.

Acquisition and validation methods

Phantoms

Catphan 700

To obtain a baseline for overall CT scanner performance, we scanned a Catphan 700 phantom (Phantom Laboratory, Greenwich, NY, USA) that had been designed specifically for routine quality assurance on CT scanners. It is only suitable for use in CT, and contains test modules for contrast, geometric accuracy, and spatial resolution.18, 19

COPDGene Phantom II

The COPDGene Phantom II (Phantom Laboratory, Greenwich, NY, USA) was designed for thoracic CT quality assurance in prospective clinical trials (specifically asthma and chronic obstructive pulmonary disorder) with guidance from the Quantitative Image Biomarker Alliance Technical Committee. We used the CCT162 version, which included the standard version CTP698 with two additional supports and acrylic end‐plates for stabilization of the phantom during the scanning. An outer polyurethane ring simulated tissue attenuation while an internal oval body (15 cm × 25 cm) simulated lung attenuation. The inner oval held a number of cylindrical cavities for foam, acrylic, and water,20, 21 as well as a number of internal structures simulating different‐sized bronchi.

Triple modality 3D Abdominal Phantom

A 3D multimodality Abdominal Phantom (CIRS, Norfolk, Virginia, USA) measuring 26 cm × 12.5 cm × 19 cm22 was designed to be used for liver biopsy training under guidance by CT, magnetic resonance imaging, or ultrasonography. We scanned Model 057A that simulated the abdomen of a small adult. The materials encased within the phantom represented the liver, portal vein, kidneys, bottom of the lungs, abdominal aorta, vena cava, lumbar spine, and six lowest ribs.

Image acquisition

The images used in our study were acquired using three different CT scanners at independent Dutch centers: MAASTRO Clinic (Maastricht), Radboud University Medical Center (Nijmegen) and University Medical Center Groningen (Groningen). The standard clinical operating procedures for thoracic and abdominal radiotherapy planning CT scans at each of the three centers were used to generate a baseline scan of each phantom. These baseline parameters are stated in Tables 1 and 2, for the Phantom Laboratory and CIRS phantoms, respectively.
Table 1

CT scanner details and image acquisition parameters for baseline scans of the Catphan 700 and COPDGene Phantom II in each of the participating clinics

ParametersDICOM tags MAASTRO Clinic (MAAS) Radboud University Medical Center (RADB)University Medical Center Groningen (UMCG)
Catphan 700/COPDGene Phantom II baseline scan parameters
Manufacturer(0008, 0070)SiemensPhillipsSiemens
Model(0008, 1090)Biograph 40Brilliance Big BoreBiograph 64
Software Version(0018, 1020)syngo CT 2006A3.6.6VG60A
Slice thickness (mm)(0018, 0050)333
TUBE VOLTAGE (KV)(0018, 0060)12012080
Reconstruction diameter (mm)(0018, 1100)500255239
Tube current (mA)(0018, 1151)39134149
EXPOSURE (mAs)(0018, 1152)2412453
Convolution kernel(0018, 1210)B31fBI30f
Rows(0028, 0010)5121024512
Columns(0028, 0011)5121024512
Pixel spacing(0028, 0030)0.980.250.46
Bits stored(0028, 0101)121212
High bit(0028, 0102)111111
Rescale offset(0028, 1052)−1024−1024−1024
Rescale slope(0028, 1053)111
Table 2

CT scanner details and image acquisition parameters for baseline scans of the multimodality CIRS Abdominal Phantom in each of the participating clinics

ParametersDICOM tagsMAASTRO Clinic (MAAS)Radboud University Medical Center (RADB)University Medical Center Groningen (UMCG)
Triple modality 3D Abdominal Phantom baseline scan parameters
Manufacturer(0008, 0070)SiemensPhillipsSiemens
Model(0008, 1090)Biograph 40Brilliance Big BoreBiograph 64
Software Version(0018,1020)syngo CT 2006A3.6.6VG60A
Manufacturer(0008, 0070)SiemensPhillipsSiemens
TUBE VOLTAGE (KV)(0018, 0060)12012080
Reconstruction diameter (mm)(0018, 1100)500255239
Tube current (mA)(0018, 1151)11819018
EXPOSURE (mAs)(0018, 1152)731759
Convolution kernel(0018, 1210)B30fBI30f
Rows(0028, 0010)512512512
Columns(0028, 0011)512512512
Pixel spacing(0028, 0030)0.980.750.59
Bits stored(0028, 0101)121212
High bit(0028, 0102)111111
Rescale offset(0028, 1052)−1024−1024−1024
Rescale slope(0028, 1053)111
CT scanner details and image acquisition parameters for baseline scans of the Catphan 700 and COPDGene Phantom II in each of the participating clinics CT scanner details and image acquisition parameters for baseline scans of the multimodality CIRS Abdominal Phantom in each of the participating clinics We subsequently applied perturbations to imaging settings of the baseline scan. We adjusted the following parameters strictly one at a time and saved each scan: slice thickness (1, 3, and 5 mm), reconstruction kernels (between three and five settings depending on the scanner), and current‐exposure product (50, 150, and 300 mAs). The individual setting for each scan is given in Tables 3 and 4, for the Phantom Laboratory and CIRS phantoms, respectively.
Table 3

The individual scan settings for the Catphan 700 and COPD II phantoms from the participating different Dutch clinics

SubjectInstitutionSlice thickness (mm)Voltage (kvp)Current (mA)Exposure (mAs)Convolution kernel
Collection: series 1 — Catphan 700 and COPD II individual subject scan settings
CatPhan‐01‐MAASMAASTRO31203924B31f
CatPhan‐01‐RADBRadboud3120134124B
CatPhan‐01‐UMCGGroningen380165.558.5I30f
COPD‐001‐MAASMAASTRO312013080.5B31f
COPD‐001‐RADBRadboud3120210194B
COPD‐001‐UMCGGroningen312019168I30f
COPD‐002‐MAASMAASTRO1120112.569.5B31f
COPD‐002‐RADBRadboud1120210194B
COPD‐002‐UMCGGroningen112020573I30f
COPD‐003‐MAASMAASTRO5120106.566B31f
COPD‐003‐RADBRadboud5120210194B
COPD‐003‐UMCGGroningen512019569I30f
COPD‐004‐MAASMAASTRO31209156B31f
COPD‐004‐RADBRadboud31205450B
COPD‐004‐UMCGGroningen312014050I30f
COPD‐005‐MAASMAASTRO31208050B31f
COPD‐005‐RADBRadboud3120108100B
COPD‐005‐UMCGGroningen3120280100I30f
COPD‐006‐MAASMAASTRO312013080.5B41f
COPD‐006‐RADBRadboud3120325300B
COPD‐006‐UMCGGroningen3120660300I30f
COPD‐007‐MAASMAASTRO312013080.5B41f
COPD‐007‐RADBRadboud3120210194A
COPD‐007‐UMCGGroningen3100230104I40f
COPD‐008‐MAASMAASTRO312013080.5B75f
COPD‐008‐RADBRadboud3120210194C
COPD‐008‐UMCGGroningen3100231104I44f
COPD‐009‐MAASMAASTRO312013080.5B60f
COPD‐009‐RADBRadboud3120210194E
COPD‐009‐UMCGGroningen3100236107I49f
COPD‐010‐MAASMAASTRO312013080.5B80f
COPD‐010‐RADBRadboud3120210194L
COPD‐010‐UMCGGroningen3100232105I50f
COPD‐011‐UMCGGroningen3100238108I70f
COPD‐012‐UMCGGroningen3100236107B30f
Table 4

The individual settings of the Triple modality 3D Abdominal Phantoms from the three participating Dutch clinics

SubjectInstitutionSlice thickness (mm)Voltage (kvp)Current (mA)Exposure (mAs)Convolution kernel
Collection: series 2 — CIRS multimodality phantom individual subject scan settings
CIRS‐AB‐001‐MAASMAASTRO312011873B30f
CIRS‐AB‐001‐RADBRadboud3120190175B
CIRS‐AB‐001‐UMCGGroningen310010050I30f
CIRS‐AB‐002‐MAASMAASTRO112013383B30f
CIRS‐AB‐002‐RADBRadboud1120190175B
CIRS‐AB‐002‐UMCGGroningen11009547I30f
CIRS‐AB‐003‐MAASMAASTRO512013685B30f
CIRS‐AB‐003‐RADBRadboud5120190175B
CIRS‐AB‐003‐UMCGGroningen51009849I30f
CIRS‐AB‐004a‐UMCGGroningen312010050I30f
CIRS‐AB‐004b‐UMCGGroningen312010050I30f
CIRS‐AB‐004‐MAASMAASTRO312014188B30f
CIRS‐AB‐004‐RADBRadboud31205450B
CIRS‐AB‐005a‐UMCGGroningen3120200100I30f
CIRS‐AB‐005b‐UMCGGroningen3120200100I30f
CIRS‐AB‐005‐MAASMAASTRO112013785B30f
CIRS‐AB‐005‐RADBRadboud3120108100B
CIRS‐AB‐006‐MAASMAASTRO5120137.585.5B30f
CIRS‐AB‐006‐RADBRadboud3120325300B
CIRS‐AB‐006‐UMCGGroningen3120600300I30f
CIRS‐AB‐007‐RADBRadboud3120190175A
CIRS‐AB‐007‐UMCGGroningen31009849I40f
CIRS‐AB‐008‐RADBRadboud3120190175C
CIRS‐AB‐008‐UMCGGroningen31009849I44f
CIRS‐AB‐009‐RADBRadboud3120190175D
CIRS‐AB‐009‐UMCGGroningen31009648I49f
CIRS‐AB‐010‐UMCGGroningen31009748I50f
CIRS‐AB‐011‐UMCGGroningen31009849I70f
CIRS‐AB‐012‐UMCGGroningen31009748B30f
The individual scan settings for the Catphan 700 and COPD II phantoms from the participating different Dutch clinics The individual settings of the Triple modality 3D Abdominal Phantoms from the three participating Dutch clinics

Image annotations

CatPhan 700 images were only used for image quality assessment of the baseline scans between participating centers, therefore, no annotations were added to the scans. Regions of interest (ROIs) on the COPDGene and Abdominal Phantoms were manually delineated in MIRADA DBx (version 1.2.0.59, Mirada Medical, Oxford, United Kingdom). In the COPD phantom, we delineated four distinct spherical ROIs within two of the insert cavities. In the multimodality phantom, we delineated two different ROIs corresponding to two of the simulated liver lesions, one large and one small (as shown in Fig. 1). The delineations were performed by one medical physicist at MAASTRO Clinic. All images and annotations were then exported as Digital Imaging and Communications in Medicine (DICOM)‐Radiotherapy (RT) objects.
Figure 1

The delineated spherical ROIs within two of the inserts cavities for the COPD and Triple modality 3D Abdominal Phantoms are presented in (a) and (b), respectively.

The delineated spherical ROIs within two of the inserts cavities for the COPD and Triple modality 3D Abdominal Phantoms are presented in (a) and (b), respectively.

Data format and usage notes

Our scans are made open access via an instance of the Extensible Neuroimaging Archive Toolkit (XNAT) hosted within Dutch national research infrastructure (TraIT, http://www.ctmm-trait.nl).23 XNAT is an open source platform for imaging‐based research and clinical investigations, which manages access to different datasets compartmentalized into separate projects (i.e., collections). Within each collection, XNAT permits browsing of individual cases. The platform supports direct uploading of DICOM images and DICOM‐RT objects (plan, structure set, and dose grid) with http file transfer.24 Studies in XNAT can be queried and retrieved by means of an API (Application Programming Interface) in the Python programming language by installing the xnat library (https://pypi.org/project/xnat/). The Phantom Laboratory COPD phantom images have been uploaded to the XNAT collection STWSTRATEGY‐Phantom_Series1: (https://xnat.bmia.nl/data/projects/stwstrategyps1). The CIRS multimodality Abdominal Phantom images have been uploaded to the XNAT collection STWSTRATEGY‐Phantom_Series2: (https://xnat.bmia.nl/data/projects/stwstrategyps2). The Phantom Laboratories Catphan 700 phantom images have been uploaded to the XNAT collection STW‐STRATEGY‐Phantom_Series3: (https://xnat.bmia.nl/data/projects/stwstrategyps3). In each of the above collections, the subject identifier matches exactly the names shown in the leftmost column of Tables 3 and 4. DICOM‐formatted images and the annotations as DICOM‐RTStruct objects are nested under the subject level. A python script for downloading an entire collection is available here: (https://github.com/maastroclinic/XNAT-collections-download-script). The images of the Catphan 700 quality assurance phantom from each center were analyzed online on the quality assurance tests webpage of the ImageOwl company (https://catphanqa.imageowl.com/). Clinical lung CT imaging protocols were used as the reference baseline for radiomics studies, rather than the vendors’ service scan setting. The ImageOWL vendor service provides a detailed quality assurance analysis along with the implementation of linearity and sensitometry plots, noise measurements, and the spatial linearity. All quality assurance test parameters were within tolerance for the clinical lung scan settings used as the reference. The quality assurance reports can be found in Data S1.

Discussion

We have made publically available multicenter phantom CT scans to support investigations in radiomics repeatability and reproducibility, specifically to identify features that may be unstable with respect to image acquisition settings in simplified geometry. Radiomics reproducibility may be investigated as a function of: scanner manufacturer/scanner type, slice thickness, tube current (i.e., signal to noise ratio), and reconstruction algorithms. We invite the radiomics community to make use of our dataset for research by extracting radiomic features with their own processing pipelines and comparing the results with other investigators. We also invite the community to contact us in order to share the results of their computations. For the next steps, we intend to host the computed features set from the open source library pyradiomics v2 (https://github.com/Radiomics/pyradiomics)25 as well as the associated DICOM image metadata on a public open‐access website (http://www.radiomics.org). This is a fundamental step toward improving benchmarking and standardization of the radiomics field of study. This is in support of valuable harmonization projects such as the IBSI (Image Biomarker Standardization Initiative).26 The features and metadata will be made available as linked Resource Descriptor Format (RDF) objects labeled with a dedicated radiomic‐specific semantic web ontology (https://bioportal.bioontology.org/ontologies/RO), such that the data can be queried through the SPARQL language. To assist the radiomics community with data sharing, a standard tabular template and conversion script to RDF will also be provided at http://www.radiomics.org. A number of key limitations in the data must be noted at the present time. First, as explicitly declared by the phantom manufacturers, the phantoms used in this study had not been designed with the specific aim of simulating standard radiomic features. It is presently not fully understood exactly what should be used as a canonical set of imaging features. Secondly, we posit that the so‐called “test lesions” within the current phantoms represent oversimplified geometries and relatively uniformly dense material. Complex texture patterns and shape features are not well represented in such simple phantoms. However, these phantoms do present a preliminary opportunity for investigating reproducibility of radiomic features, thus we may be able to test for certain features that already unstable in simplified conditions. We would assert that a feature that is not reproducible in such a constrained setting might be unlikely to be highly reproducible in multi‐institutional human studies. To improve on the current situation, the dataset might be expanded by scans of more phantoms that contain more realistic tumor‐mimicking inserts. These may prove to be more suitable for selecting stable features for inclusion in radiomic investigations. One example of a public phantom dataset which is available on “The Cancer Imaging Archive‐TCIA” is the Credence Cartridge Radiomics (CCR) Phantom (https://wiki.cancerimagingarchive.net/display/Public/Credence+Cartridge+Radiomics+Phantom+CT+Scans). The CCR phantom collection has a similar goal as our study, the investigation of the reproducibility of radiomic features. There is a significant factor that differentiates the CCR phantom public dataset from our phantoms public collections. The structure of the CCR phantom which includes ten cartridges, each with a unique texture, addresses only the question of repeatability and reproducibility of textural features. Lastly, while we have started with CT as the most commonly available imaging modality in our field, we intend to expand this collection to include positron emission tomography (PET) and magnetic resonance imaging (MRI). In addition to making available multicenter and multimodality phantoms for radiomics reproducibility studies, future work in this field should make publicly accessible DICOM metadata and image preprocessing steps, so as to make radiomics studies as findable, accessible, interoperable, reusable (FAIR) as possible. To this end, image metadata needs to be linked to the features using publicly available Semantic DICOM (SEDI) ontology27 and the Radiomics ontology needs to extended to cover image preprocessing.

Conclusion

We offer a publicly accessible multicenter CT phantom dataset with carefully controlled image acquisition parameters to support reproducibility research in the field of radiomics. The dataset is hosted in a well‐established and publicly funded XNAT instance. The data are shared under a Creative Commons Attribution 3.0 License (free to browse, download, and use at no cost for scientific and educational purposes). The dataset is offered to the radiomics community to compare simple features extracted with different software pipelines as well as to identify features that may not be stable with respect to image acquisition conditions even under highly simplified conditions. Our unique contribution to the field is to investigate the robustness of each radiomic feature with respect to different scanning acquisition parameters.

Conflict of interest

The authors declare no conflict of interests pertaining to the above scientific work. Data S1: Supplementary material with all the information used for the analysis of the scans of the quality assurance phantom Catphan 700. Click here for additional data file.
  21 in total

1.  DicomBrowser: software for viewing and modifying DICOM metadata.

Authors:  Kevin A Archie; Daniel S Marcus
Journal:  J Digit Imaging       Date:  2012-10       Impact factor: 4.056

2.  Quantitative radiomics: impact of stochastic effects on textural feature analysis implies the need for standards.

Authors:  Matthew J Nyflot; Fei Yang; Darrin Byrd; Stephen R Bowen; George A Sandison; Paul E Kinahan
Journal:  J Med Imaging (Bellingham)       Date:  2015-08-05

3.  The Extensible Neuroimaging Archive Toolkit: an informatics platform for managing, exploring, and sharing neuroimaging data.

Authors:  Daniel S Marcus; Timothy R Olsen; Mohana Ramaratnam; Randy L Buckner
Journal:  Neuroinformatics       Date:  2007

Review 4.  Radiogenomics predicting tumor responses to radiotherapy in lung cancer.

Authors:  Amit K Das; Marcus H Bell; Chaitanya S Nirodi; Michael D Story; John D Minna
Journal:  Semin Radiat Oncol       Date:  2010-07       Impact factor: 5.934

5.  Reproducibility and Prognosis of Quantitative Features Extracted from CT Images.

Authors:  Yoganand Balagurunathan; Yuhua Gu; Hua Wang; Virendra Kumar; Olya Grove; Sam Hawkins; Jongphil Kim; Dmitry B Goldgof; Lawrence O Hall; Robert A Gatenby; Robert J Gillies
Journal:  Transl Oncol       Date:  2014-02-01       Impact factor: 4.243

Review 6.  Quantifying tumour heterogeneity in 18F-FDG PET/CT imaging by texture analysis.

Authors:  Sugama Chicklore; Vicky Goh; Musib Siddique; Arunabha Roy; Paul K Marsden; Gary J R Cook
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-10-13       Impact factor: 9.236

7.  Influence of CT acquisition and reconstruction parameters on radiomic feature reproducibility.

Authors:  Abhishek Midya; Jayasree Chakraborty; Mithat Gönen; Richard K G Do; Amber L Simpson
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-15

Review 8.  Transparent reporting of a multivariable prediction model for individual prognosis or diagnosis (TRIPOD): the TRIPOD statement.

Authors:  Gary S Collins; Johannes B Reitsma; Douglas G Altman; Karel G M Moons
Journal:  BMJ       Date:  2015-01-07

Review 9.  Imaging and cancer: a review.

Authors:  Leonard Fass
Journal:  Mol Oncol       Date:  2008-05-10       Impact factor: 7.449

10.  How to use CT texture analysis for prognostication of non-small cell lung cancer.

Authors:  Kenneth A Miles
Journal:  Cancer Imaging       Date:  2016-04-11       Impact factor: 3.909

View more
  4 in total

1.  Harmonization of radiomic feature distributions: impact on classification of hepatic tissue in CT imaging.

Authors:  Hubert Beaumont; Antoine Iannessi; Anne-Sophie Bertrand; Jean Michel Cucchi; Olivier Lucidarme
Journal:  Eur Radiol       Date:  2021-01-18       Impact factor: 5.315

Review 2.  Radiomics: A Primer on Processing Workflow and Analysis.

Authors:  Emily Avery; Pina C Sanelli; Mariam Aboian; Seyedmehdi Payabvash
Journal:  Semin Ultrasound CT MR       Date:  2022-02-12       Impact factor: 1.641

3.  Technical Note: Virtual phantom analyses for preprocessing evaluation and detection of a robust feature set for MRI-radiomics of the brain.

Authors:  Marco Bologna; Valentina Corino; Luca Mainardi
Journal:  Med Phys       Date:  2019-10-08       Impact factor: 4.071

4.  XNAT-PIC: Extending XNAT to Preclinical Imaging Centers.

Authors:  Sara Zullino; Alessandro Paglialonga; Walter Dastrù; Dario Livio Longo; Silvio Aime
Journal:  J Digit Imaging       Date:  2022-03-18       Impact factor: 4.903

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.