Optimal pressing strength and time for capillary refilling time.

Rui Kawaguchi1, Taka-Aki Nakada2, Taku Oshima1, Masayoshi Shinozaki3, Toshiya Nakaguchi3, Hideaki Haneishi3, Shigeto Oda1.   

Abstract

Entities:  

Mesh:

Year:  2019        PMID: 30621748      PMCID: PMC6323707          DOI: 10.1186/s13054-018-2295-3

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


× No keyword cloud information.
Substantial investigations on the microcirculation have highlighted the importance of the capillary refilling time (CRT), which is widely used in clinical settings [1-3]. However, CRT measurement conditions, including pressing strength and time, are inconsistent (e.g., pressing strength: light, moderate, or firm; pressing time: 3 s, 5 s, or until the capillary bed visually blanches) [2-5]. CRT is conventionally measured visually after manual compression, which limits the accuracy of measurement and precludes the precise determination of how long and strong the nail bed should be pressed. Therefore, we developed a novel device which can be adjusted for pressing strength and time to precisely measure CRT using an electric actuator and strength and color sensors. CRT was measured in age- and sex-matched healthy adults (n = 31) using the developed device under conditions of pressing strength of 1, 3, 5, and 7 N, and a pressing time 1, 2, 3, 4, 5, and 6 s (Additional file 1: Methods and supplemental data). There was a significant difference in CRT for pressing strength but not for pressing time (two-way analysis of variance (ANOVA): strength P < 0.001, time P = 0.97) (Fig. 1). There were significant differences in CRT between pressing strengths of 1 and 3 N (P < 0.01) but no significant differences among 3, 5, and 7 N (P = 0.16). Thus, 3–7 N appears to be the optimal strength range for CRT measurement. A strength of 3 N is that which is needed to lift an object weighing 300 g (e.g., two smartphones) using fingers; the examinees would feel a 300-g load on their fingers. To identify sufficient pressing time, we further analyzed the plots of the readings from the color sensor during nail bed compression. The measured values change rapidly in the initial part of compression (rapid phase) until the inflection point, where the slope of color change reduces (slow phase) (Fig. 2a). The longest period of the rapid phase was 1.999 s among all the study subjects (minimum 0.1022 s, 95% 1.149 s) (Fig. 2b). The pressure release during the rapid phase may destabilize this measurement. Thus, a pressing time of 2 s is a good threshold to obtain stable CRT measurements for all subjects.
Fig. 1

Capillary refilling time (CRT). CRT was measured in 31 age- and sex-matched healthy adults under the conditions with pressing strength 1, 3, 5, and 7 N, and pressing time 1, 2, 3, 4, 5, and 6 s in technical quintuplicate. Error bars indicate standard errors

Fig. 2

Analysis on pressing time. a A schema of curve plot of the readings by the color sensor. The value sensing by color sensor rapidly changes in the initial part of the pressing nail bed (rapid phase) until the inflection point and then the slope of color change become slow (slow phase). b Histogram of inflection times (n = 744 (31 subjects × 24 conditions)). Gray lines indicate 5% and 95% lines. The red line indicates 50% time line. CRT capillary refilling time

Capillary refilling time (CRT). CRT was measured in 31 age- and sex-matched healthy adults under the conditions with pressing strength 1, 3, 5, and 7 N, and pressing time 1, 2, 3, 4, 5, and 6 s in technical quintuplicate. Error bars indicate standard errors Analysis on pressing time. a A schema of curve plot of the readings by the color sensor. The value sensing by color sensor rapidly changes in the initial part of the pressing nail bed (rapid phase) until the inflection point and then the slope of color change become slow (slow phase). b Histogram of inflection times (n = 744 (31 subjects × 24 conditions)). Gray lines indicate 5% and 95% lines. The red line indicates 50% time line. CRT capillary refilling time In the study aimed at identifying the standard pressing strength and CRT, pressing the nail bed with 3–7 N for 2 s appears to be optimal. Further development of portable CRT measurement devices fulfilling these conditions may contribute to achieving precise CRT measurements to monitor the microcirculation in clinical settings. Methods and supplemental data. Figure E1. Picture of developed device. The upper image is the uncovered condition to show the internal structure. The lower image is the covered condition in which we measured CRT of study subjects. 1) Stepper motor. 2) Uni-axial stage. 3) Color sensor, light source and indenter. 5) Industrial camera. 6) Emergency stop switch. 7) Cover for hiding press principle. Size 16 cm × 20 cm × 23 cm. Figure E2. Principle of the device. Figure E3. Definition of CRT. Table E1. Characteristics of subjects and data. (DOCX 29524 kb)
  8 in total

1.  Basics of changes in hemodynamic monitoring in sepsis care.

Authors:  Jae Chol Yoon; Won Young Kim
Journal:  J Thorac Dis       Date:  2019-10       Impact factor: 2.895

Review 2.  Endothelial Activation and Microcirculatory Disorders in Sepsis.

Authors:  Lisa Raia; Lara Zafrani
Journal:  Front Med (Lausanne)       Date:  2022-06-03

3.  Interpretation or misinterpretation of clinical trials on septic shock: about the ANDROMEDA-SHOCK trial.

Authors:  François Dépret; Maxime Coutrot
Journal:  Ann Transl Med       Date:  2020-06

4.  Does training level affect the accuracy of visual assessment of capillary refill time?

Authors:  Koichiro Shinozaki; Lee S Jacobson; Kota Saeki; Naoki Kobayashi; Steve Weisner; Julianne M Falotico; Timmy Li; Junhwan Kim; Joshua W Lampe; Lance B Becker
Journal:  Crit Care       Date:  2019-05-06       Impact factor: 9.097

5.  Feedback function for capillary refilling time measurement device.

Authors:  Masayoshi Shinozaki; Taka-Aki Nakada; Rui Kawaguchi; Yuichiro Yoshimura; Toshiya Nakaguchi; Hideaki Haneishi; Shigeto Oda
Journal:  Crit Care       Date:  2019-09-03       Impact factor: 9.097

Review 6.  Advances in the Approaches Using Peripheral Perfusion for Monitoring Hemodynamic Status.

Authors:  Julianne M Falotico; Koichiro Shinozaki; Kota Saeki; Lance B Becker
Journal:  Front Med (Lausanne)       Date:  2020-12-07

7.  Comparison of point-of-care peripheral perfusion assessment using pulse oximetry sensor with manual capillary refill time: clinical pilot study in the emergency department.

Authors:  Koichiro Shinozaki; Lee S Jacobson; Kota Saeki; Hideaki Hirahara; Naoki Kobayashi; Steve Weisner; Julianne M Falotico; Timmy Li; Junhwan Kim; Lance B Becker
Journal:  J Intensive Care       Date:  2019-11-27

8.  Pulse oximetry-based capillary refilling evaluation predicts postoperative outcomes in liver transplantation: a prospective observational cohort study.

Authors:  Miyuki Yamamoto; Kent Doi; Naoki Hayase; Toshifumi Asada; Nobuhisa Akamatsu; Junichi Kaneko; Kiyoshi Hasegawa; Naoto Morimura
Journal:  BMC Anesthesiol       Date:  2020-09-29       Impact factor: 2.217

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.