Literature DB >> 30614137

Improving the detection specificity of endogenous MRI for reactive oxygen species (ROS).

Rong-Wen Tain1,2, Alessandro M Scotti1,3, Kejia Cai1,3.   

Abstract

BACKGROUND: The detection of tissue reactive oxygen species (ROS) using endogenous MRI methods has great potential applications in research and the clinic. We recently demonstrated that ROS produce a significant T1 -shortening effect. However, T1 or T1 -weighted contrast is not specific, as there are many other factors that alter tissue T1 .
PURPOSE: To investigate whether the presence of ROS alters tissue environmental conditions such as the proton exchange rate (K ex ) to improve the detection specificity of endogenous ROS MRI. STUDY TYPE: Prospective. SUBJECTS/PHANTOM: The ROS-producing phantoms consisted of fresh egg white treated with H2 O2 and healthy mice injected with pro-oxidative rotenone. FIELD STRENGTH/SEQUENCE: T1 mapping was performed based on fast spin-echo sequence and K ex was evaluated using chemical exchange saturation transfer (CEST) MRI with varied saturation power (QUESP) on a 9.4 T animal scanner. ASSESSMENT: Phantom experiments were conducted to evaluate the overall K ex of CEST-expressing metabolites in fresh egg white treated with H2 O2 of various concentrations (0, 0.025, 0.05, 0.1, and 0.25 v/v%). The egg white phantom continuously produced ROS for more than 3 hours. Various experiments were performed to rule out potential contributing factors to the observed K ex changes. In addition, in vivo MRI study was conducted with a well-established rotenone-exposed mouse model. STATISTICAL TESTS: Student's t-test.
RESULTS: Egg white phantoms treated with H2 O2 of various concentrations showed a 26-85% increase in K ex compared with controls. In addition, the K ex of egg white is negligibly affected by other potential confounding factors, including paramagnetic contrast agents (<11%), oxygen (2.3%), and iron oxidation (<10%). Changes in temperature (<1°C) and pH (ΔpH <0.1) in H2 O2 -treated egg white were also negligible. Results from the in vivo rotenone study were consistent with the phantom studies by showing reduced T1 relaxation time (6%) and increased K ex (9%) in rotenone-treated mice. DATA
CONCLUSION: We demonstrate that the specificity of endogenous ROS MRI can be improved with the aid of proton exchange rate mapping. LEVEL OF EVIDENCE: 2 Technical Efficacy Stage: 2 J. Magn. Reson. Imaging 2019;50:583-591.
© 2019 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  CEST MRI; T1 relaxation time; free radicals; oxidative stress; proton exchange rate; reactive oxygen species

Year:  2019        PMID: 30614137     DOI: 10.1002/jmri.26629

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  Novel proton exchange rate MRI presents unique contrast in brains of ischemic stroke patients.

Authors:  Zhenxiong Wang; Mehran Shaghaghi; Shun Zhang; Guiling Zhang; Yiran Zhou; Di Wu; Zhuoli Zhang; Wenzhen Zhu; Kejia Cai
Journal:  J Neurosci Methods       Date:  2020-09-05       Impact factor: 2.390

2.  In vivo quantification of proton exchange rate in healthy human brains with omega plot.

Authors:  Mehran Shaghaghi; Weiwei Chen; Alessandro Scotti; Haiqi Ye; Yan Zhang; Wenzhen Zhu; Kejia Cai
Journal:  Quant Imaging Med Surg       Date:  2019-10

3.  Proton exchange rate of chemical exchange saturation transfer MRI constructed from direct saturation-removed omega plots to improve the assessment of patients with ischemic stroke.

Authors:  Yiran Zhou; Zhenxiong Wang; Mehran Shaghaghi; Guiling Zhang; Hongquan Zhu; Di Wu; Jun Lu; Su Yan; Shun Zhang; Li Li; Jia Li; Kejia Cai; Wenzhen Zhu
Journal:  Quant Imaging Med Surg       Date:  2022-10

4.  In Vivo Proton Exchange Rate (kex ) MRI for the Characterization of Multiple Sclerosis Lesions in Patients.

Authors:  Haiqi Ye; Mehran Shaghaghi; Qianlan Chen; Yan Zhang; Sarah E Lutz; Weiwei Chen; Kejia Cai
Journal:  J Magn Reson Imaging       Date:  2020-09-24       Impact factor: 4.813

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.