Abstract
INTRODUCTION: In the last 20 years glycated albumin (GA) measurement has been demonstrated to be a reliable glycation marker and recently as the most innovative one in western countries. Glycated albumin has been already adopted by some Asian countries due to its usefulness in diabetes screening. The aim of the present study was to investigate for the first time the effects of different anticoagulants on GA assay.Entities:
Keywords: anticoagulants; diabetes mellitus; preanalytical phase; serum albumin
Mesh:
Substances:
Year: 2018 PMID: 30591815 PMCID: PMC6294155 DOI: 10.11613/BM.2019.010901
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Results obtained in serum and in plasma samples collected with different anticoagulants
| Serum | 15.4 (13.2-19.1) | - | - | 6.1 (5.1-7.5) | - | - | 43.0 (39.8-45.3) | - | - |
| K3EDTA | 15.7 (13.6-19.9) | < 0.001 | + 2.8 | 6.0 (5.0-7.5) | < 0.001 | - 1.9 | 40.1 (38.0-43.0) | < 0.001 | - 5.1 |
| Li-Hep | 15.6 (13.3-19.7) | < 0.001 | + 0.9 | 6.1 (5.1-7.5) | 0.281 | + 0.4 | 42.2 (39.7-45.4) | 0.022 | - 0.7 |
| NaF-EDTA | 15.5 (13.1-19.3) | 0.403 | + 0.1 | 5.0 (4.2-6.6) | < 0.001 | - 15.1 | 35.5 (34.1-38.6) | < 0.001 | - 15.0 |
| Data are expressed as median and interquartile range (in brackets). Differences between plasma and serum were tested using paired Wilcoxon test. The mean relative bias respect to the results obtained in serum is also presented. P < 0.05 was considered statistically significant. K3EDTA - tripotassium-ethylenediaminetetraacetic acid. Li-Hep - lithium heparin. NaF - sodium-fluoride. GA - glycated albumin. | |||||||||
Figure 1Bland-Altman plots showing the relative bias between GA% concentrations as measured in tripotassium-ethylenediaminetetraacetic acid (K3EDTA, A), lithium heparin (Li-Hep, B), sodium-fluoride ethylenediaminetetraacetic acid (NaF-EDTA, C) containing tubes and serum are presented. GA - glycated albumin. The solid line represents mean bias and dashed lines represent ± 1.96 SD interval.