Tomotaka Ohshima1, Shigeru Miyachi1, Naoki Matsuo2, Reo Kawaguchi2, Yoko Kato3, Masakazu Takayasu2.
Abstract
We previously reported using a microguidewire with a curved, pigtail-shaped tip as an anchor for navigating a microcatheter. However, we considered that the microcatheter could follow even if the tip of wire did not advance so far distally. The present study assessed the usefulness of a modified version of this pigtail-shaped microguidewire for navigating microcatheters to difficult vasculatures during neurointerventions. The microguidewire was repeatedly evaluated using a silicon tube vascular model. We also employed this method as the first choice for some patients. In both the in vitro tests and clinical use, the round curved tip could be used to position the microguidewire at the entry of a sharply angled vessel or aneurysmal neck, and the following catheter could then be advanced gently and easily. This technique allowed navigation of the microcatheter without inserting the guidewire into the distal vasculature.Entities:
Keywords: catheterization; endovascular treatment; intracranial aneurysms; microguidewire; shaping
Mesh:
Year: 2018 PMID: 30587869 PMCID: PMC6295425 DOI: 10.18999/nagjms.80.4.551