Characterizing a novel scintillating glass for application to megavoltage cone-beam computed tomography.

Yue-Houng Hu1, Daniel Shedlock2, Adam Wang2, Joerg Rottmann1, Paul Baturin2, Marios Myronakis1, Pascal Huber3, Rony Fueglistaller3, Mengying Shi4, Daniel Morf3, Josh Star-Lack2, Ross I Berbeco1.   

Abstract

PURPOSE: The purpose of this study was to evaluate the performance of a prototype electric portal imaging device (EPID) with a high detective quantum efficiency (DQE) scintillator, LKH-5. Specifically, image quality in context of both planar and megavoltage (MV) cone-beam computed tomography (CBCT) is analyzed.
METHODS: Planar image quality in terms of modulation transfer function (MTF), noise power spectrum (NPS), and DQE are measured and compared to an existing EPID (AS-1200) using the 6 MV beamline for a Varian TrueBeam linac. Imager performance is contextualized for three-dimensional (3D), MV-CBCT performance by measuring imager lag and analyzing the expected degradation of the DQE as a function of dose. Finally, comparisons between reconstructed images of the Catphan phantom in terms of qualitative quality and signal-difference-to-noise ratio (SDNR) are made for 6 MV images using both conventional and LKH-5 EPIDs as well as for the kilovoltage (kV) on-board imager (OBI).
RESULTS: Analysis of the NPS reveals linearity at all measured doses using the prototype LKH-5 detector. While the first zero of the MTF is much lower for the LKH-5 detector than the conventional EPID (0.6 cycles/mm vs 1.6 cycles/mm), the normalized NPS (NNPS) multiplied by total quanta (qNNPS) of the LKH-5 detector is roughly a factor of seven to eight times lower, yielding a DQE(0) of approximately 8%. First, second, and third frame lag were measured at approximately 23%, 5%, and 1%, respectively, although no noticeable image artifacts were apparent in reconstructed volumes. Analysis of low-dose performance reveals that DQE(0) remains at 80% of its maximum value at a dose as low as 7.5 × 10-6  MU. For a 400 projection technique, this represents a total scan dose of 0.0030 MU, suggesting that if imaging doses are increased to a value typical of kV-CBCT scans (~2.7 cGy), the LKH-5 detector will retain quantum noise limited performance. Finally, comparing Catphan scans, the prototype detector exhibits much lower image noise than the conventional EPID, resulting in improved small object representation. Furthermore, SDNR of H2 O and polystyrene cylinders improved from -1.95 and 2.94 to -15 and 18.7, respectively.
CONCLUSIONS: Imaging performance of the prototype LKH-5 detector was measured and analyzed for both planar and 3D contexts. Improving noise transfer of the detector results in concurrent improvement of DQE(0). For 3D imaging, temporal characteristics were adequate for artifact-free performance and at relevant doses, the detector retained quantum noise limited performance. Although quantitative MTF measurements suggest poorer resolution, small object representation of the prototype imager is qualitatively improved over the conventional detector due to the measured reduction in noise.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  zzm321990DQEzzm321990; zzm321990EPIDzzm321990; zzm321990MTFzzm321990; zzm321990NPSzzm321990; MV-CBCT; portal imaging; scintillator

Mesh:

Year:  2019        PMID: 30586163     DOI: 10.1002/mp.13355

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Frequency-dependent optimal weighting approach for megavoltage multilayer imagers.

Authors:  Ingrid Valencia Lozano; Mengying Shi; Marios Myronakis; Paul Baturin; Rony Fueglistaller; Pascal Huber; Mathias Lehmann; Daniel Morf; Dianne Ferguson; Matthew W Jacobson; Thomas Harris; Ross I Berbeco; Christopher L Williams
Journal:  Phys Med Biol       Date:  2021-04-16       Impact factor: 4.174

2.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174

3.  GPU-accelerated Monte Carlo simulation of MV-CBCT.

Authors:  Mengying Shi; Marios Myronakis; Matthew Jacobson; Dianne Ferguson; Christopher Williams; Mathias Lehmann; Paul Baturin; Pascal Huber; Rony Fueglistaller; Ingrid Valencia Lozano; Thomas Harris; Daniel Morf; Ross I Berbeco
Journal:  Phys Med Biol       Date:  2020-12-02       Impact factor: 4.174

4.  A rapid, accurate image simulation strategy for mega-voltage cone-beam computed tomography.

Authors:  Mengying Shi; Marios Myronakis; Matthew Jacobson; Mathias Lehmann; Dianne Ferguson; Paul Baturin; Pascal Huber; Rony Fueglistaller; Thomas Harris; Ingrid Valencia Lozano; Christopher Williams; Daniel Morf; Ross I Berbeco
Journal:  Phys Med Biol       Date:  2020-07-06       Impact factor: 4.174

Review 5.  Technical Principles of Dual-Energy Cone Beam Computed Tomography and Clinical Applications for Radiation Therapy.

Authors:  Shailaja Sajja; Young Lee; Markus Eriksson; Håkan Nordström; Arjun Sahgal; Masoud Hashemi; James G Mainprize; Mark Ruschin
Journal:  Adv Radiat Oncol       Date:  2019-07-30
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.