Janet Gifford1, Moeen Saleem2.
Abstract
Entities:
Keywords: Electromagnetic interference; Electrosurgery; Micra leadless pacemaker; Oversensing; Sensing integrity counter
Year: 2018 PMID: 30581734 PMCID: PMC6301881 DOI: 10.1016/j.hrcr.2018.08.011
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Procedures and outcomes
| Procedure | Type of electrosurgery or biopsy | Mode | Sensitivity | SIC | Heart rate > 200 bpm | Device reset | Reported pacing inhibition |
|---|---|---|---|---|---|---|---|
| Hysterectomy, salpingectomy, and repair of enterocele | Monopolar | VVIR | 2.0 mV | SIC 0 | No | No | No |
| Foot ulcer debridement and toe amputation | Monopolar | VVI | 2.0 mV | SIC 0 | No | No | No |
| Neck exploration and mediastinal drainage | Monopolar | VVI | 2.0 mV | SIC 0 | No | No | No |
| Cystoscopy, bladder biopsy, and fulguration | Fulguration and biopsy | VVI | 1.5 mV | SIC 0 | No | No | No |
| EGD with biopsy | Biopsy | VVI | 1.5 mV | SIC 0 | No | No | No |
| Exploratory laparotomy and colon resection | Monopolar | VVI | 1.5 mV | SIC 0 | No | No | No |
| Repair hip fracture | Monopolar | VVI | 2.0 mV | SIC 0 | No | No | No |
EGD = esophagogastroduodenoscopy; SIC = sensing integrity counter.
On heart rate histogram.