Mohammad Rubayet Hasan1, Hassan Al Mana2, Virginia Young3, Patrick Tang4, Eva Thomas4, Rusung Tan4, Peter Tilley5.
Abstract
Commercial multiplex assays, built on different chemistries and platforms are widely available for simultaneous detection of pathogens that cause respiratory infections. However, these tests are often difficult to implement in a resource limited setting because of high cost. In this study, we developed and validated a method for simultaneous testing of common respiratory pathogens (Respanel) by real-time PCR in a convenient, strip-tube array format. Primers and probes for sixteen PCR assays were selected from the literature or newly designed. Following optimization of individual PCR assays, strip-tube arrays were prepared by dispensing primer-probe mixes (PPM) into two sets of 8-tube strips. Nucleic acid extracts from specimens were mixed with PCR master mix, and dispensed column-wise into 2 × 8-wells of a 96-well plate. PPMs from strip-tubes were then added to the wells using a multichannel pipette for real-time PCR. Individual PCR assays were optimized using previously known specimens (n = 394) with 91%-100% concordance with culture, DFA or PCR results. Respanel was then tested in a routine manner at two different sites using specimens (n = 147) previously tested by Qiagen Resplex I&II or Fast-Track Diagnostics Respiratory Pathogens 21 assays. The sensitivity, specificity and accuracy of Respanel were 94%, 95% and 95%, respectively, against Resplex and 88%, 100% and 99%, respectively, against FTDRP21. Respanel detected more pathogens (p < 0.05) than Resplex but the rate of pathogen detection was not significantly different from FTDRP21. Respanel is a convenient and inexpensive assay that is more sensitive than Resplex and comparable to FTDRP21 for the detection of common respiratory pathogens.Entities:
Keywords: Commercial multiplex assays; Laboratory-developed tests; Real-time PCR; Respiratory pathogens; Validation
Mesh:
Year: 2018 PMID: 30578898 PMCID: PMC7113679 DOI: 10.1016/j.jviromet.2018.12.013
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.014
Fig. 1Respanel assay workflow.
Comparison of individual PCR assays included in Respanel against different reference methods.
| PCR assay | Sample description | Reference methods | Total number of samples | TP | FP | TN | FN | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Adenovirus | QCMD EQA specimens | Culture | 44 | 20 | 1 | 20 | 3 | 87 | 95 | 95 | 87 | 91 |
| Human Metapneumovirus | QCMD EQA specimens | PCR | 24 | 10 | 0 | 14 | 0 | 100 | 100 | 100 | 100 | 100 |
| Influenza A | QCMD EQA specimens | PCR | 28 | 17 | 0 | 10 | 1 | 94 | 100 | 100 | 91 | 96 |
| Influenza B | QCMD EQA specimens | DFA | 42 | 11 | 0 | 31 | 0 | 100 | 100 | 100 | 100 | 100 |
| Parainfluenza virus 1 | CAP EQA specimens | Culture | 38 | 10 | 0 | 27 | 1 | 91 | 100 | 100 | 96 | 97 |
| Parainfluenza virus 2 | CAP EQA specimens | Culture | 38 | 7 | 0 | 30 | 1 | 88 | 100 | 100 | 97 | 97 |
| Parainfluenza virus 3 | CAP EQA specimens | Culture | 33 | 12 | 0 | 19 | 2 | 86 | 100 | 100 | 90 | 94 |
| Respiratory Syncytial Virus | QCMD EQA specimens | DFA | 52 | 29 | 1 | 22 | 0 | 100 | 96 | 97 | 100 | 98 |
| Enterovirus | QCMD EQA specimens | Culture | 44 | 24 | 0 | 17 | 3 | 89 | 100 | 100 | 85 | 93 |
| Rhinovirus | QCMD EQA specimens | Culture | 44 | 12 | 0 | 32 | 0 | 100 | 100 | 100 | 100 | 100 |
| Bocavirus | Clinical samples | Resplex | 30 | 14 | 0 | 16 | 0 | 100 | 100 | 100 | 100 | 100 |
| Coronavirus 229E | Clinical samples | Resplex | 22 | 3 | 0 | 17 | 2 | 60 | 100 | 100 | 89 | 91 |
| Coronavirus OC43 | Clinical samples | Resplex | 22 | 1 | 0 | 21 | 0 | 100 | 100 | 100 | 100 | 100 |
| Coronavirus NL63 | Clinical samples | Resplex | 22 | 7 | 0 | 15 | 0 | 100 | 100 | 100 | 100 | 100 |
| Coronavirus HKU1 | Clinical samples | Resplex | 22 | 4 | 0 | 18 | 0 | 100 | 100 | 100 | 100 | 100 |
| ATCC | Culture | 52 | 14 | 0 | 38 | 0 | 100 | 100 | 100 | 100 | 100 | |
| ATCC | Culture | 36 | 9 | 0 | 27 | 0 | 100 | 100 | 100 | 100 | 100 | |
| ATCC | Culture | 52 | 3 | 0 | 49 | 0 | 100 | 100 | 100 | 100 | 100 | |
| QCMD EQA specimens | Culture | 125 | 25 | 1 | 98 | 1 | 96 | 99 | 96 | 99 | 98 |
QCMD = Quality Control for Molecular Diagnostics, CAP = College of American Pathologists, EQA = External Quality Assessment, ATCC = American Type Culture Collection, DFA = Direct Fluorescent Antibody, TP = True positive, FP = False positive, TN = True negative, FN = False negative, PPV = Positive predictive value, NPV = Negative predictive value. Sensitivity, specificity, PPV, NPV and accuracy were calculated as defined in the materials and methods (Section 2.6).
Comparison of Respanel against Resplex I & II with 34 respiratory samples.
| Organism | TP | FP | TN | FN | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|---|---|---|---|
| Adenovirus | 1 | 0 | 31 | 2 | 33 | 100 | 100 | 94 | 94 |
| Human Metapneumovirus | 1 | 0 | 33 | 0 | 100 | 100 | 100 | 100 | 100 |
| Influenza A | 2 | 6 | 26 | 0 | 100 | 81 | 25 | 100 | 82 |
| Influenza B | 0 | 0 | 34 | 0 | – | 100 | – | 100 | 100 |
| Parainfluenza Virus 1 | 1 | 0 | 33 | 0 | 100 | 100 | 100 | 100 | 100 |
| Parainfluenza Virus 2 | 2 | 1 | 31 | 0 | 100 | 97 | 67 | 100 | 97 |
| Parainfluenza Virus 3 | 0 | 3 | 31 | 0 | – | 91 | 0 | 100 | 91 |
| Respiratory Syncytial Virus | 3 | 1 | 30 | 0 | 100 | 97 | 75 | 100 | 97 |
| 12 | 6 | 15 | 1 | 92 | 71 | 67 | 94 | 79 | |
| Bocavirus | 6 | 2 | 26 | 0 | 100 | 93 | 75 | 100 | 94 |
| Coronavirus 229E | 0 | 0 | 34 | 0 | – | 100 | – | 100 | 100 |
| Coronavirus OC43 | 2 | 0 | 32 | 0 | 100 | 100 | 100 | 100 | 100 |
| Coronavirus NL63 | 1 | 1 | 32 | 0 | 100 | 97 | 50 | 100 | 97 |
| Coronavirus HKU1 | 0 | 0 | 34 | 0 | – | 100 | – | 100 | 100 |
| 4 | 1 | 29 | 0 | 100 | 97 | 80 | 100 | 97 | |
| 10 | 5 | 19 | 0 | 100 | 79 | 67 | 100 | 85 | |
| 0 | 1 | 33 | 0 | – | 97 | 0 | 100 | 97 |
TP = True positive, FP = False positive, TN = True negative, FN = False negative, PPV = Positive predictive value, NPV = Negative predictive value. Sensitivity, specificity, PPV, NPV and accuracy were calculated as defined in the materials and methods (Section 2.6).
Enterovirus and Rhinovirus results were compared together because Resplex Enterovirus and Rhinovirus assays were reported in aggregate.
Comparison of Respanel against FTDRP21 with 113 respiratory samples.
| Organism | TP | FP | TN | FN | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) |
|---|---|---|---|---|---|---|---|---|---|
| Adenovirus | 7 | 0 | 106 | 0 | 100 | 100 | 100 | 100 | 100 |
| Human Metapneumovirus | 2 | 0 | 110 | 1 | 67 | 100 | 100 | 99 | 99 |
| Influenza A | 4 | 0 | 108 | 1 | 80 | 100 | 100 | 99 | 99 |
| Influenza B | 3 | 0 | 110 | 0 | 100 | 100 | 100 | 100 | 100 |
| Parainfluenza virus 1 | 3 | 0 | 110 | 0 | 100 | 100 | 100 | 100 | 100 |
| Parainfluenza virus 2 | 2 | 0 | 111 | 0 | 100 | 100 | 100 | 100 | 100 |
| Parainfluenza virus 3 | 5 | 0 | 107 | 1 | 83 | 100 | 100 | 99 | 99 |
| Respiratory Syncytial Virus | 8 | 0 | 105 | 0 | 100 | 100 | 100 | 100 | 100 |
| Enterovirus | 3 | 1 | 108 | 1 | 75 | 99 | 75 | 99 | 98 |
| Rhinovirus | 18 | 1 | 91 | 3 | 86 | 99 | 95 | 97 | 96 |
| Bocavirus | 3 | 1 | 108 | 1 | 75 | 99 | 75 | 99 | 98 |
| Coronavirus 229E | 3 | 0 | 109 | 1 | 75 | 100 | 100 | 99 | 99 |
| Coronavirus OC43 | 3 | 0 | 110 | 0 | 100 | 100 | 100 | 100 | 100 |
| Coronavirus NL63 | 4 | 1 | 108 | 0 | 100 | 99 | 80 | 100 | 99 |
| Coronavirus HKU1 | 1 | 1 | 111 | 0 | 100 | 99 | 50 | 100 | 99 |
| 3 | 1 | 108 | 1 | 75 | 99 | 75 | 99 | 98 | |
| 0 | 0 | 113 | 0 | – | 100 | – | 100 | 100 |
TP = True positive, FP = False positive, TN = True negative, FN = False negative, PPV = Positive predictive value, NPV = Negative predictive value. Sensitivity, specificity, PPV, NPV and accuracy were calculated as defined in the materials and methods (Section 2.6).
Diagnostic performance of Respanel compared to Resplex and FTDRP21 assays.
| Statistics | Resplex I&II | FTDRP21 |
|---|---|---|
| Total number of results | 578 | 1921 |
| True positive | 45 | 72 |
| True negative | 503 | 1833 |
| False positive | 27 | 6 |
| False negative | 3 | 10 |
| Sensitivity | 94% (95% CI: 83%-99%) | 88% (95% CI: 79%-94%) |
| Specificity | 95% (95% CI: 93%-97%) | 100% (95% CI: 99%-100%) |
| Positive predictive value (PPV) | 63% (95% CI: 53%-71%) | 92% (95% CI: 84%-96%) |
| Negative predictive value (NPV) | 99% (95% CI: 98%-100%) | 99% (95% CI: 99%-100%) |
| Accuracy | 95% (95% CI: 93%-96%) | 99% (95% CI: 99%-100%) |