BACKGROUND: A project for the standardization of HbA2 was launched by the IFCC back in 2004.Entities:
Keywords: HbA2; standardization; traceability; β-thalassemia
Year: 2018 PMID: 30574041 PMCID: PMC6295589
Source DB: PubMed Journal: EJIFCC ISSN: 1650-3414
Performance characteristics of different methods for HbA2 testing
| Method | Between-run reproducibility | Bias versus consensus mean | r | |
|---|---|---|---|---|
| Slope | Intercept | |||
| Bio-Rad D10 | 1.7 | 0.978 to 1.018 | 0.07 to 0.24 | 0.9981 |
| Bio-Rad Variant II β-thalassemia | 0.9 | 0.866 to 0.905 | 0.20 to 0.38 | 0.9977 |
| Bio-Rad Variant II dual kit | 1.7 | 1.026 to 1.047 | -0.23 to -0.14 | 0.9995 |
| Menarini HA8160 | 2.3 | 0.790 to 0.826 | 0.56 to 0.72 | 0.9976 |
| Trinity Premier High Resolution | 6.6 | 1.012 to 1.087 | -0.08 to 0.24 | 0.9941 |
| Trinity Premier Quick Scan | 1.4 | 1.115 to 1.156 | -0.42 to -0.24 | 0.9985 |
| Sebia Capillars 2 FP | 1.9 | 0.898 to 0.946 | -0.19 to 0.02 | 0.9969 |
| Tosoh G8 | 1.9 | 1.150 to 1.180 | -0.63 to -0.50 | 0.9992 |
Data were derived from reference 3. Reproducibility is expressed as CV%.
For slope and intercept, the 95% confidence intervals are reported.
Figure 1IFCC reference system and traceability chain for HbA2 measurement