| Literature DB >> 30561284 |
Nancy L McDaniel1, Wendy Novicoff2, Brian Gunnell3, David Cattell Gordon3.
Abstract
Background and Objective: Research demonstrates that telemedicine is effective in pediatric settings but little is published to validate the quality of the data acquired by remote peripheral examination devices to accurately inform clinical decision-making. Introduction: The primary aim was to compare a novel Food and Drug Administration (FDA)-cleared multifunctional remote examination device (Tyto) with other stand-alone digital examination devices. The secondary aim was to ascertain whether either device produced images or sounds better able to provide clinical information to clinicians caring for children. Materials andEntities:
Keywords: e-health; pediatrics; telehealth; telemedicine
Year: 2018 PMID: 30561284 PMCID: PMC6918850 DOI: 10.1089/tmj.2018.0214
Source DB: PubMed Journal: Telemed J E Health ISSN: 1530-5627 Impact factor: 3.536
Statistical Summary: Image and Sound Quality Comparisons
| ASSESSMENT | ANALYSIS | RECORDED DATAPOINTS | |
|---|---|---|---|
| Overall image/sound quality | χ2 (Chi-squared) | Tyto 1,381 | 0.0001 |
| UVA[ | |||
| Proportions without a diagnosis | Proportional | Tyto 0.53 | 0.001 |
| UVA 0.48 | |||
| Image/sound quality mean score (SD) | Two sample | Tyto 2.80 (1.05) | 0.0001 |
| UVA 3.39 (0.94) | |||
| Image quality—ear | χ2 (Chi-squared) | Tyto 696 | 0.0001 |
| UVA 797 | |||
| Sound quality—heart | χ2 (Chi-squared) | Tyto 350 | 0.0001 |
| UVA 381 | |||
| Sound quality—lung | χ2 (Chi-squared) | Tyto 335 | 0.0001 |
| UVA 353 | |||
| Image/sound quality—inter-rater reliability | Cronbach alpha/intraclass correlation coefficient | 0.84–0.99 | |
| Image/sound quality—intra-rater reliability | Cronbach alpha/intraclass correlation coefficient | 0.90–0.99 | |
Auditory comparisons were made with the One Digital Stethoscope (Thinklabs Medical LLC) and imaging comparisons were made with the Horus HD Digital Scope System (JEDME Instrument Co). Both are listed as standard of care UVA devices used in clinical practice.
SD, standard deviation.
Statistical Summary: Diagnostic Comparisons
| ASSESSMENT | ANALYSIS | RECORDED DATAPOINTS | |
|---|---|---|---|
| Overall diagnosis | χ2 (Chi-squared) | Tyto 1,378 | 0.0001 |
| UVA 1,498 | |||
| Proportions without a diagnosis | Proportional | Tyto 0.56 | 0.001 |
| UVA 0.48 | |||
| Diagnostic mean score | Two sample | Tyto 1.66 | 0.0001 |
| UVA 1.95 | |||
| Diagnosis—ear | χ2 (Chi-squared) | Tyto 694 | 0.0001 |
| UVA 774 | |||
| Diagnosis—heart | χ2 (Chi-squared) | Tyto 350 | 0.0001 |
| UVA 372 | |||
| Diagnosis—lung | χ2 (Chi-squared) | Tyto 334 | 0.004 |
| UVA 352 | |||
| Diagnosis—inter-rater reliability | Cronbach alpha/intraclass correlation coefficient | 0.85–0.99 | |
| Diagnosis—intra-rater reliability | Cronbach alpha/intraclass correlation coefficient | 0.89–0.99 | |