The repeatability of superficial retinal vessel density measurements in eyes with long axial length using optical coherence tomography angiography.

Mengyang Li1, Enzhong Jin1, Chongya Dong2, Chuan Zhang1, Mingwei Zhao1, Jinfeng Qu3,4.   

Abstract

BACKGROUND: To investigate the repeatability of superficial retinal vessel density measurements in healthy eyes with long axial length (AL) using optical coherence tomography angiography (OCTA).
METHODS: There were 60 eyes of 31 volunteers enrolled in this cross-sectional observational study. All subjects underwent OCTA, AL and refraction test. The enrolled eyes were divided into the long AL group (26 mm ≤ AL < 28 mm) and normal AL group (22 mm ≤ AL < 26 mm). The vessel length density (VLD), perfusion density (PD), and fovea avascular zone (FAZ) of the superficial retinal vessel were evaluated. Repeatability was assessed by intraclass correlation coefficients (ICCs) and Bland-Altman analysis. Pearson's r correlation was used to analyze the relation of AL and the absolute difference between two measurements.
RESULTS: The 3 × 3 mm scan pattern showed good repeatability with all ICCs over 0.7. For all parameters of all scan patterns, the ICCs of the normal AL group were distinctly higher than those of the long AL group; this finding was also confirmed by Bland-Altman analysis. The correlation analysis of AL and repeatability of OCTA parameters showed significant negative correlations between the ALs and repeatability of VLD in 6 × 6 mm inner ring (r2 = 0.13, p = 0.01), VLD in 6 × 6 mm outer ring (r2 = 0.09, p = 0.02) and PD in 6 × 6 mm outer ring (r2 = 0.08, p = 0.03).
CONCLUSIONS: The AL and the scanned area will both affect the repeatability of superficial retinal vessel density measurements in OCTA.

Entities:  

Keywords:  Fovea avascular zone; Myopia; Optical coherence tomography angiography; Perfusion density; Repeatability; Vessel length density

Mesh:

Year:  2018        PMID: 30558579      PMCID: PMC6297956          DOI: 10.1186/s12886-018-0992-y

Source DB:  PubMed          Journal:  BMC Ophthalmol        ISSN: 1471-2415            Impact factor:   2.209


Introduction

Optical coherence tomography angiography (OCTA) is an advanced imaging technique that allows depth resolved study of retinal microvascular networks. Compared with traditional angiography using contrast agents, OCTA can obtain more subtle macular structure details [1]. Since high myopia has become a prominent worldwide problem especially in Asia [2], understanding the alteration of retinal and choroidal vascular network in the myopia population using OCTA was of great significance. Previous studies have performed many quantitative analyses of retinal vascular network using OCTA in myopic eyes [3-5]. The superficial capillary plexus with surrounding pericytes is a component of the inner blood retina barrier, and changes in this structure play an important role in formation of macular edema in diabetic retinopathy, retinal vein occlusion, macular degeneration and other vascular diseases [6, 7]. ZEISS AngioPlex Metrix software provides automatic quantitative analysis of retinal vessels in the superficial layer, including the fovea avascular zone (FAZ), vessel length density (VLD) and perfusion density (PD), but the repeatability remains unclear. The FAZ, VLD and PD of the superficial retinal vasculature have been reported to have high repeatability but can be affected by many factors, such as scan order [8]. In our clinical practice, we have noticed the difficulty in acquiring high-quality images of some patients with high myopia. Therefore, we intend to explore the correlation between the repeatability of OCTA parameters and AL of healthy young adults for better understanding and interpretation of vascular perfusion in patients with axial myopia. In this study, we compare the repeatability of the FAZ, VLD and PD of the superficial retinal vasculature between eyes with long AL and normal AL. Parameters from 3 × 3 mm and 6 × 6 mm scans were analyzed separately.

Materials and methods

Participants

This cross-sectional study was approved by the Peking University People’s Hospital Review Board and adhered to the tenets of the Declaration of Helsinki and Health Insurance Portability and Accountability Act. Written informed consent was obtained from all individual participants included in the study. Sixty eyes of 31 healthy subjects were enrolled in this study from May 1, 2017 to October 10, 2018. The inclusion criteria were as follows: (1) age between 20 and 35 years old; (2) no history of systemic diseases; (3) no known eye disease except for refractory error; (4) no ocular media opacity, which was confirmed by slit-lamp examination (SLE); (5) best-corrected visual acuity (BCVA) no worse than 20/20; (6) no ocular surgery history, including refractory surgery. The exclusion criteria were as follows: (1) unable to acquire good quality images after repeated measurements or poor fixation; (2) refractive error greater than − 10.00 spherical diopters or − 2.00 cylindrical diopters; (3) AL longer than 28 mm. The enrolled eyes were divided into the long AL group (AL ≥ 26 mm) and normal AL group (AL < 26 mm). All subjects underwent BCVA, SLE, refraction test, AL measurement and OCTA. AL measurement was performed using IOLMaster (Zeiss 500; Carl Zeiss Meditec, Inc., Dublin, CA).

Optical coherence tomography angiography image acquisition

OCTA images were acquired using Cirrus HD-OCT model 5000 (AngioPlex software, version 10.0; Carl Zeiss Meditec, Inc) in a dark room. All measurements were performed by one operator (M.L.) under FastTrac mode. For each eye, a 3 × 3 mm scan and a 6 × 6 mm scan centered on the fovea were acquired. Each scan was repeated at least twice for repeatability analysis. Automated OCT segmentation was performed on qualified images and manual adjustment was applied when segmentation error occurs. The qualifying images were defined satisfying the following requirements: (1) signal strength (SS) ≥7, (2) no more than one blink or motion artifact, and (3) the macular fovea remained in the center of the scanned area. The difference of SS between each two repeated scans should not lager than 1. Based on these default settings, the superficial capillary plexus (SCP) en face image was segmented with an inner boundary at internal limiting membrane (ILM) and an outer boundary at the junction of inner plexiform layer (IPL) and inner nuclear layer (INL) [9].

Vascular parameters measurement

The FAZ area, VLD and PD of SCP within ETDRS 3 × 3 mm inner ring, 3 × 3 whole grid, 6 × 6 mm inner ring, 6 × 6 mm outer ring and 6 × 6 whole grid was evaluated separately (Fig. 1). FAZ area were segmented automatically and manually adjust if necessary, and then measured respectively in both 3 × 3 and 6 × 6 mm scans by AngioPlex software automatically (Fig. 1b). PD was defined as the total area of perfused vasculature per unit area in a region of evaluation (Fig. 1c). VLD was defined as the total length of skeletonized perfused vasculature per unit area in a region of evaluation (Fig. 1d).
Fig. 1

Superficial vessel plexus en face images of OCTA. a The inner ring and outer ring of the 6 × 6 mm scan pattern. b The FAZ area automatically recognized by the software on the 3 × 3 mm scan pattern; c The perfused vasculature (marked as green) automatically segmented to calculate PD; d The skeletonized vasculature (marked as red) automatically segmented to calculate VLD

Superficial vessel plexus en face images of OCTA. a The inner ring and outer ring of the 6 × 6 mm scan pattern. b The FAZ area automatically recognized by the software on the 3 × 3 mm scan pattern; c The perfused vasculature (marked as green) automatically segmented to calculate PD; d The skeletonized vasculature (marked as red) automatically segmented to calculate VLD

Statistical analysis

Intraclass correlation coefficients (ICCs) with 95% CIs were used to evaluate the repeatability of PD, VLD and FAZ between two consecutive scans of the same pattern and also between the two groups. Considering we include both eyes of subjects in our study, the ICCs were further adjusted for inter-eye correlation using a mixed-design analysis of variance model. Agreement between two measurements was also evaluated by Bland-Altman plots. Pearson’s r correlation test was used to analyze the relation of ALs and absolute differences between two measurements. The absolute difference was defined as |m1-m2| in which m1 was the first measurement result of PD, VLD or FAZ, and m2 was the second measurement. SPSS (version 23.0, IBM Corporation, Armonk, NY) was used for ICCs calculation. Bland-Altman analysis and Pearson’s r correlation test were performed using GraphPad Prism (version 6.00 for Mac, GraphPad Software, La Jolla, California, USA). p < 0.05 was considered significant.

Results

A total of 31 subjects and 62 eyes were enrolled in this study. Two eyes were excluded due to poor SS, the remaining 60 eyes were included in the analysis. The demographics of the enrolled subjects are shown in Table 1. The raw data of VLD, PD and FAZ values for 3 × 3 mm and 6 × 6 mm were shown in Additional file 1. Mann-Whitney test was used to compare the demographics of the normal AL group (n = 32) with those of the long AL group (n = 28). The mean age of all subjects was 25.61 ± 3.00 years old (range 22–35 years), and no significant difference was shown between the two groups (p > 0.05). The ALs and spherical diopter were significantly different between the two groups (p < 0.05).
Table 1

Demographic and clinical data of the enrolled subjects

VariablesValue
22 mm ≤ AL < 26 mm26 mm ≤ AL < 28 mmTotalp-value
Number322860
Mean age (±SD)25.87 (±2.73)25.32 (±3.30)25.61(±3.00)0.20
Mean AL (±SD)24.73(±0.68)26.73 (±0.52)25.66 (±1.17)< 0.01
Mean diopter (±SD)−3.27(±1.57)−6.36(±1.70)−4.71(±2.24)< 0.01

Abbreviations: SD standard deviation, AL axial length

Demographic and clinical data of the enrolled subjects Abbreviations: SD standard deviation, AL axial length The representative OCTA images of both normal and long AL group were demonstrated in Fig. 2, which showed the comparation of VLD, PD and FAZ area outline between the two groups. The ICCs and 95% confidence intervals of VLD, PD and FAZ are shown in Table 2. There were very subtle changes of ICCs after adjusted for inter-eye correlation using a mixed-design analysis of variance model (ICC in Table 2). Therefore, we believed that the involvement of both eyes will not affect the repeatability test in this study. For all parameters of all scan patterns, the ICC of the normal AL group was much higher than that of the long AL group, with the biggest difference of VLD in 6 × 6 inner ring and smallest difference of PD in 3 × 3 inner ring and FAZ area in 6 × 6 scan. In normal AL group, the repeatability was very good (ICC > 0.9) in VLD and PD in 3 × 3 inner ring and FAZ area in 3 × 3 mm scan; was good (ICC > 0.7) in VLD and PD in 6 × 6 inner ring; was moderate (ICC > 0.4) in VLD and PD in 6 × 6 outer ring and FAZ area in 6 × 6 mm scan.. In long AL group, no parameters has ICC larger than 0.9, the repeatability was good in VLD and PD in 3 × 3 inner ring and FAZ area in 3 × 3 mm scan; was moderate in VLD and PD in 6 × 6 inner and outer ring and FAZ area in 6 × 6 mm scan. For VLD and PD in inner ring and FAZ area, the repeatability was higher in 3 × 3 mm scan than in 6 × 6 mm scan.
Fig. 2

The representative images of OCTA images of normal and long axial length eyes showed the comparation of VLD, PD and FAZ area between the two groups. a and b showed two repeated OCTA 3 × 3 mm SCP en face image and b-scans of a 25-year-old male with an axial length of 24.51 mm. c and d showed two repeat 3 × 3 mm SCP en face images of a 35-year-old male with an axial length of 27.34 mm. The yellow area was the automatically segmented foveal avascular zone. The upper and right numbers were VLD and PD in the inner ring respectively. e and f showed two repeated OCTA 6 × 6 mm scans of a 26-year-old female with an axial length of 23.58 mm; g and h showed 6 × 6 mm scans of a 35-year-old male with an axial length of 27.27 mm. The yellow area was the automatically segmented foveal avascular zone. The top, upper middle, lower middle and bottom numbers represent VLD in the outer ring, VLD in the inner ring, PD in the inner ring and PD in the outer ring. Noticed the differences in all metrics between two repeated scans and lower repeatability in long axial eyes

Table 2

Statistical analysis results: ICC of VLD, PD and FAZ

Scan patterns and metrics22 mm ≤ AL < 26 mm26 mm ≤ AL < 28 mmTotal
ICCICCadjICCICCadjICCICCadj
VLD3 × 3 inner0.90(0.81–0.95)0.910.78(0.56–0.88)0.750.83(0.74–0.90)0.83
6 × 6 inner0.76(0.56–0.87)0.760.41(0.05–0.67)0.410.54(0.31–0.70)0.54
6 × 6 outer0.61(0.34–0.79)0.610.43(0.08–0.69)0.430.50(0.28–0.66)0.50
PD3 × 3 inner0.90(0.81–0.95)0.900.80(0.63–0.90)0.810.86(0.77–0.91)0.85
6 × 6 inner0.71(0.49–0.85)0.710.43(0.07–0.68)0.430.52(0.31–0.68)0.52
6 × 6 outer0.55(0.26–0.75)0.540.40(0.04–0.67)0.400.46(0.23–0.64)0.46
FAZ3 × 30.96(0.92–0.98)0.960.86(0.73–0.94)0.800.92(0.87–0.95)0.92
6 × 60.69(0.46–0.84)0.690.62(0.34–0.80)0.600.65(0.48–0.77)0.67

ICCs of VLD, PD and FAZ in two groups measured by 3 × 3 mm and 6 × 6 mm scan patterns are presented in Table 2. The columns of ICC showed the ICCs adjusted for inter-eye correlation using a mixed-design analysis of variance model

Abbreviations: VLD vessel length density, PD perfusion density, FAZ fovea avascular zone, AL axial length, ICC adjusted ICC

The representative images of OCTA images of normal and long axial length eyes showed the comparation of VLD, PD and FAZ area between the two groups. a and b showed two repeated OCTA 3 × 3 mm SCP en face image and b-scans of a 25-year-old male with an axial length of 24.51 mm. c and d showed two repeat 3 × 3 mm SCP en face images of a 35-year-old male with an axial length of 27.34 mm. The yellow area was the automatically segmented foveal avascular zone. The upper and right numbers were VLD and PD in the inner ring respectively. e and f showed two repeated OCTA 6 × 6 mm scans of a 26-year-old female with an axial length of 23.58 mm; g and h showed 6 × 6 mm scans of a 35-year-old male with an axial length of 27.27 mm. The yellow area was the automatically segmented foveal avascular zone. The top, upper middle, lower middle and bottom numbers represent VLD in the outer ring, VLD in the inner ring, PD in the inner ring and PD in the outer ring. Noticed the differences in all metrics between two repeated scans and lower repeatability in long axial eyes Statistical analysis results: ICC of VLD, PD and FAZ ICCs of VLD, PD and FAZ in two groups measured by 3 × 3 mm and 6 × 6 mm scan patterns are presented in Table 2. The columns of ICC showed the ICCs adjusted for inter-eye correlation using a mixed-design analysis of variance model Abbreviations: VLD vessel length density, PD perfusion density, FAZ fovea avascular zone, AL axial length, ICC adjusted ICC Bland-Altman analysis of VLD, PD and FAZ is shown in Fig. 3. The mean differences in VLD, PD and FAZ area are depicted by black horizontal lines in the graph. Two green dashed horizontal lines indicate the mean difference ± 1.96 SD, which are also referred to as limit agreements (LAs), representing the range of values for the differences between the two measurements that could be expected 95% of the time. All of the values from the long AL group are marked as red dots.
Fig. 3

Bland-Altman analysis of VLD, PD and FAZ in two different scan patterns. All of the values from the long AL group are marked as red dots, and values from the normal AL group are marked as empty circles. a Bland-Altman analysis of VLD in 3 × 3 mm inner ring. b VLD in 6 × 6 mm inner ring. c VLD in 6 × 6 mm outer ring. d, e, f showed Bland-Altman analysis of PD. d Bland-Altman analysis of PD in 3 × 3 mm inner ring. e PD in 6 × 6 mm inner ring. f PD in 6 × 6 mm outer ring. g Bland-Altman analysis of FAZ in 3 × 3 mm scan. h Bland-Altman analysis of FAZ in 6 × 6 mm scan

Bland-Altman analysis of VLD, PD and FAZ in two different scan patterns. All of the values from the long AL group are marked as red dots, and values from the normal AL group are marked as empty circles. a Bland-Altman analysis of VLD in 3 × 3 mm inner ring. b VLD in 6 × 6 mm inner ring. c VLD in 6 × 6 mm outer ring. d, e, f showed Bland-Altman analysis of PD. d Bland-Altman analysis of PD in 3 × 3 mm inner ring. e PD in 6 × 6 mm inner ring. f PD in 6 × 6 mm outer ring. g Bland-Altman analysis of FAZ in 3 × 3 mm scan. h Bland-Altman analysis of FAZ in 6 × 6 mm scan For VLD, the range of LA was smallest in the 3 × 3 mm inner ring (lower and upper bounds of − 1.84 and 2.20 mm− 1) and biggest in the 6 × 6 inner ring (lower and upper bounds of − 3.17 and 2.59 mm− 1). The number of eyes beyond LAs of VLD in the 3 × 3 mm inner ring was 4 eyes in long AL group and 1 eye in normal AL group. The number of eyes beyond LAs of VLD in the 6 × 6 mm inner ring was 6 eyes in long AL group and no eye in normal AL group. The number of eyes beyond LAs of VLD in the 6 × 6 mm outer ring was 4 eyes in long AL group and 1 eye in normal AL group. For PD, the range of LA was smallest in the 3 × 3 mm inner ring (lower and upper bounds of − 0.030 and 0.039 mm− 1) and biggest in the 6 × 6 inner ring (lower and upper bounds of − 0.090 and 0.086 mm− 1). The number of eyes beyond LAs of PD in the 3 × 3 mm inner ring was 1 eye in each group. The number of eyes beyond LAs of PD in the 6 × 6 mm inner ring was 1 eye in long AL group and no eye in normal AL group. The number of eyes beyond LAs of PD in the 6 × 6 mm outer ring was 3 eyes in long AL group and 1 eye in normal AL group. For FAZ, the range of LA was smaller in the 3 × 3 mm scan (lower and upper bounds of − 0.069 and 0.062 mm− 1) and biggest in the 6 × 6 mm scan (lower and upper bounds of − 0.175 and 0.141 mm− 1). The number of eyes beyond LAs of PD in the 3 × 3 mm scan was 3 eyes in long AL group and 1 eye in normal AL group. The number of eyes beyond LAs of PD in the 6 × 6 mm scan was 2 eyes in long AL group and no eye in normal AL group. The correlation between AL and the absolute difference of two measurements is shown in Fig. 4. The scatter plots and fitting lines indicated significant negative correlations between the ALs and repeatability of VLD in 6 × 6 mm inner ring (r2 = 0.13, p = 0.01), VLD in 6 × 6 mm outer ring (r2 = 0.09, p = 0.02) and PD in 6 × 6 mm outer ring (r2 = 0.08, p = 0.03). Comparing to VLD, the repeatability of PD was less affected by AL in both 3 × 3 mm and 6 × 6 mm scans. There was also borderline statistically significant negative correlation between VLD in 6 × 6 mm outer ring (r2 = 0.06, p = 0.07), FAZ area in 6 × 6 scan (r2 = 0.06, p = 0.06) and 3 × 3 scan (r2 = 0.05, p = 0.07).
Fig. 4

Relationship between AL and the absolute difference between two constant measurements in PD, VLD and FAZ area. a AL and absolute difference in PD. b AL and absolute difference in VLD. c AL and absolute difference in FAZ area

Relationship between AL and the absolute difference between two constant measurements in PD, VLD and FAZ area. a AL and absolute difference in PD. b AL and absolute difference in VLD. c AL and absolute difference in FAZ area

Discussion

Myopia is one of the most prevalent eye disorders and is estimated to affect 1.5 billion people worldwide. Previously, with the development of OCTA, many researchers have focused on quantification of macular vascular density using OCTA. Changes in the superficial and deep retinal microvasculature as well as choroidal capillaries have been found in high myopia patients [4, 10–12]. These changes could be explained by longer axial length and thinner choroid but the repeatability of these measurements should be considered before we draw any conclusion. Unfortunately, most of previous studies did not report their repeatability of quantification parameters. The variation of the use of build-in automatic algorism or custom algorism and the application of magnification correction in different studies makes it difficult to interpret the result from different studies. To the best of our knowledge, this study is the first to explore the repeatability of the FAZ, VLD and PD of SCP between eyes with long AL and normal. Many previous studies revealed the good repeatability of Optovue Avanti OCTA [13-16] and other devices [17-19] in normal AL eyes. For the ZEISS device, researchers also reported good repeatability of FAZ in eyes with spherical refraction within − 3.00D [20]. However, their work was based on manual segmentation and algorism using ImageJ, which inevitably introduces human error into the evaluation. Another study using built-in AngioPlex software for automatic quantitative analysis report excellent repeatability in normal eyes, with a greater ICC for the 3 × 3 mm scan pattern [8]. The ICCs of vessel density and FAZ area in those studies were remarkably higher than the repeatability of them in all subjects in our results, probably because the average AL in our study (25.66 mm) was higher than theirs. For all parameters of all scan patterns, the ICCs of the normal AL group were distinctly higher than those of the long AL group; this finding was also confirmed by Bland-Altman analysis. The correlation analysis of AL and repeatability of OCTA parameters showed significant or borderline significant negative correlations between the ALs and repeatability of almost all parameters except PD in 3 × 3 mm inner ring. This may be explained by the attenuation of vascular signal in long AL eyes or real reduction of superficial macular microvascular complex in long AL eyes. With the progression of myopia, the elongation of the eyeball stretches the retinal tissue, some very narrow microvascular may become undetectable now and then in the repeat scans. In general, quantitative analysis of vessel density of the inner ring of the 3 × 3 mm scan pattern had the best repeatability, with an ICC of 0.90 and 0.91 for VLD and PD, respectively. The ICC of FAZ area in 3 × 3 mm scan was the highest (0.96) among all parameters. These indicate that the scan area also has influence on the repeatability of the parameters we studied. From the Bland-Altman analysis of VLD, PD and FAZ, we can notice that the distribution of VLD in the 6 × 6 mm inner ring moved to the smaller side of X axial and become more disperse comparing to 3 × 3 mm inner ring. A possible explanation is the decreased resolution of microvascular network in large scanned area. Thus, cautions should be raised to interpret quantitative measurement from different scan area. For the evaluation of FAZ, we recommend the 3 × 3 mm scan for better repeatability. Our study has the following limitations: relative small sample size; all of the images were obtained from one device to avoid variability but the result was also limited to certain device; we limited our subjects to healthy young adults with an age range of 22–35 years to decrease the effect of age and potential age-related vascular diseases, particularly in FAZ area [21, 22]. Thus, further studies are needed to reveal the effect of AL in older population.

Conclusions

In conclusion, The AL and the scanned area will both affect the repeatability of superficial retinal vessel density measurements in OCTA. The elongation of AL lowered the repeatability of superficial retinal vessel density and FAZ measurements in OCTA. For the same parameter, 3 × 3 mm scan pattern has better repeatability than that of 6 × 6 mm scan pattern. The 6 × 6 mm scan pattern was more affected by AL. Cautions should be raised when comparing microvascular parameters from different scan pattern or from patients with different AL. Table S1. Raw data of OCTA measurements. This file includes VLD, PD and FAZ values for 3 × 3 mm and 6 × 6 mm scans of all the subjects in this study. (XLSX 32 kb)
  8 in total

1.  Corneal densitometry patterns in Descemet membrane endothelial keratoplasty and Descemet stripping automated keratoplasty.

Authors:  Jorge Peraza-Nieves; José-María Sánchez-González; Carlos Rocha-de-Lossada; Rahul Rachwani-Anil; Miriam Sánchez-Valera; Davide Borroni; Josep Torras-Sanvicens
Journal:  Int Ophthalmol       Date:  2021-03-24       Impact factor: 2.031

2.  Predicting Axial Length From Choroidal Thickness on Optical Coherence Tomography Images With Machine Learning Based Algorithms.

Authors:  Hao-Chun Lu; Hsin-Yi Chen; Chien-Jung Huang; Pao-Hsien Chu; Lung-Sheng Wu; Chia-Ying Tsai
Journal:  Front Med (Lausanne)       Date:  2022-06-28

3.  Factors Affecting Repeatability of Assessment of the Retinal Microvasculature Using Optical Coherence Tomography Angiography in Healthy Subjects.

Authors:  Taek Hoon Lee; Hyung Bin Lim; Ki Yup Nam; Kyeungmin Kim; Jung Yeul Kim
Journal:  Sci Rep       Date:  2019-11-08       Impact factor: 4.379

4.  Choroidal blood perfusion as a potential "rapid predictive index" for myopia development and progression.

Authors:  Xiangtian Zhou; Cong Ye; Xiaoyan Wang; Weihe Zhou; Peter Reinach; Jia Qu
Journal:  Eye Vis (Lond)       Date:  2021-01-04

5.  Multi-Center Repeatability of Macular Capillary Perfusion Density Using Optical Coherence Tomography Angiography.

Authors:  Hong Jiang; Ava-Gaye Simms; Shokufeh Sadaghiani; Yu You Jiang; Jessica I W Morgan; Geoffrey K Aguirre; Pradip M Pattany; John A Detre; Jianhua Wang
Journal:  Clin Ophthalmol       Date:  2022-09-17

6.  Compartmental analysis of three-dimensional choroidal vascularity and thickness of myopic eyes in young adults using SS-OCTA.

Authors:  Huan Luo; Jinfu Sun; Lan Chen; Dandan Ke; Zheng Zhong; Xi Cheng; Huimin Yu; Xufang Sun
Journal:  Front Physiol       Date:  2022-09-07       Impact factor: 4.755

7.  Repeatability of binarization thresholding methods for optical coherence tomography angiography image quantification.

Authors:  Nihaal Mehta; Phillip X Braun; Isaac Gendelman; A Yasin Alibhai; Malvika Arya; Jay S Duker; Nadia K Waheed
Journal:  Sci Rep       Date:  2020-09-21       Impact factor: 4.379

8.  Correlations among peripapillary vasculature, macular superficial capillaries, and eye structure in healthy and myopic eyes of Chinese young adults (STROBE).

Authors:  Yuxia Guo; Yunlei Pang; YanJie Kang; Xiaoguang Zhang; Han Zhang; Guisen Zhang; Lei Liu
Journal:  Medicine (Baltimore)       Date:  2020-09-11       Impact factor: 1.817

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.