Image quality and dose reduction in sinus computed tomography using iterative reconstruction: a cadaver study.

Adam J Kimple1, Stanley W McClurg1, Benjamin Y Huang2, Satyan B Sreenath1, Benjamin W McClintock2, Mohamed Tomoum3, Feng-Chang Lin4, Charles S Ebert1, Brent A Senior1,5.   

Abstract

BACKGROUND: Concerns about radiation dose in computed tomography (CT) imaging have renewed interest in iterative reconstruction (IR), a technique which has the potential to produce images with less noise at lower radiation doses than traditional filtered back projection (FBP). This study aimed to assess whether application of IR could provide comparable quality sinus CT images to FBP at lower kilovolt (kV) and milliamp (mA) settings, and to establish optimal scan settings for sinus imaging. METHODOLOGY/PRINCIPAL: 30 sinus CT scans were performed on 5 cadaver heads at two kV setting and three mA settings. Each scan was reconstructed using FBP and 3 IR settings, yielding a total of 120 images series. Each image set was blinded and randomly reviewed by 3 rhinologists and 2 neuroradiologists. Using a 5-point Likert scale, 16 anatomical landmarks, were graded with respect to image quality. Data were assessed with respect to dose and IR settings using statistical analysis.
RESULTS: Higher kV and mAs settings produced significantly higher quality images for structure identification across all 16 landmarks; however, the suitability for surgery did not increase in a linear fashion and plateaued by a total radiation dose of 0.1201 mSv. IR algorithm did not provide a benefit in the overall score of scans at a fixed kV and mAs.
CONCLUSIONS: Identification of structures in sinus CT imaging significantly correlate with the kV and mA and overall dose of radiation; however, IR did not provide additional benefit in the image quality.

Entities:  

Keywords:  computed tomography; computer assisted surgery; imaging; radiology; sinus anatomy

Year:  2018        PMID: 30556057      PMCID: PMC6292212          DOI: 10.4193/RHINOL/18.015

Source DB:  PubMed          Journal:  Rhinol Online        ISSN: 2589-5613


  10 in total

Review 1.  Dosimetry for optimisation of patient protection in computed tomography.

Authors:  K A Jessen; P C Shrimpton; J Geleijns; W Panzer; G Tosi
Journal:  Appl Radiat Isot       Date:  1999-01       Impact factor: 1.513

Review 2.  International Consensus Statement on Allergy and Rhinology: Rhinosinusitis.

Authors:  Richard R Orlandi; Todd T Kingdom; Peter H Hwang; Timothy L Smith; Jeremiah A Alt; Fuad M Baroody; Pete S Batra; Manuel Bernal-Sprekelsen; Neil Bhattacharyya; Rakesh K Chandra; Alexander Chiu; Martin J Citardi; Noam A Cohen; John DelGaudio; Martin Desrosiers; Hun-Jong Dhong; Richard Douglas; Berrylin Ferguson; Wytske J Fokkens; Christos Georgalas; Andrew Goldberg; Jan Gosepath; Daniel L Hamilos; Joseph K Han; Richard Harvey; Peter Hellings; Claire Hopkins; Roger Jankowski; Amin R Javer; Robert Kern; Stilianos Kountakis; Marek L Kowalski; Andrew Lane; Donald C Lanza; Richard Lebowitz; Heung-Man Lee; Sandra Y Lin; Valerie Lund; Amber Luong; Wolf Mann; Bradley F Marple; Kevin C McMains; Ralph Metson; Robert Naclerio; Jayakar V Nayak; Nobuyoshi Otori; James N Palmer; Sanjay R Parikh; Desiderio Passali; Anju Peters; Jay Piccirillo; David M Poetker; Alkis J Psaltis; Hassan H Ramadan; Vijay R Ramakrishnan; Herbert Riechelmann; Hwan-Jung Roh; Luke Rudmik; Raymond Sacks; Rodney J Schlosser; Brent A Senior; Raj Sindwani; James A Stankiewicz; Michael Stewart; Bruce K Tan; Elina Toskala; Richard Voegels; De Yun Wang; Erik K Weitzel; Sarah Wise; Bradford A Woodworth; Peter-John Wormald; Erin D Wright; Bing Zhou; David W Kennedy
Journal:  Int Forum Allergy Rhinol       Date:  2016-02       Impact factor: 3.858

3.  Performance of iterative image reconstruction in CT of the paranasal sinuses: a phantom study.

Authors:  B Schulz; M Beeres; B Bodelle; R Bauer; F Al-Butmeh; A Thalhammer; T J Vogl; J M Kerl
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-06       Impact factor: 3.825

4.  Conclusion of increased risk of cataracts associated with CT studies of the head may not be justified.

Authors:  Mohan Doss
Journal:  AJR Am J Roentgenol       Date:  2014-04       Impact factor: 3.959

5.  Clinical consensus statement: appropriate use of computed tomography for paranasal sinus disease.

Authors:  Gavin Setzen; Berrylin J Ferguson; Joseph K Han; John S Rhee; Rebecca S Cornelius; Stuart J Froum; Grant S Gillman; Steven M Houser; Paul R Krakovitz; Ashkan Monfared; James N Palmer; Kristina W Rosbe; Michael Setzen; Milesh M Patel
Journal:  Otolaryngol Head Neck Surg       Date:  2012-10-10       Impact factor: 3.497

6.  EPOS 2012: European position paper on rhinosinusitis and nasal polyps 2012. A summary for otorhinolaryngologists.

Authors:  Wytske J Fokkens; Valerie J Lund; Joachim Mullol; Claus Bachert; Isam Alobid; Fuad Baroody; Noam Cohen; Anders Cervin; Richard Douglas; Philippe Gevaert; Christos Georgalas; Herman Goossens; Richard Harvey; Peter Hellings; Claire Hopkins; Nick Jones; Guy Joos; Livije Kalogjera; Bob Kern; Marek Kowalski; David Price; Herbert Riechelmann; Rodney Schlosser; Brent Senior; Mike Thomas; Elina Toskala; Richard Voegels; De Yun Wang; Peter John Wormald
Journal:  Rhinology       Date:  2012-03       Impact factor: 3.681

Review 7.  Chronic rhinosinusitis pathogenesis.

Authors:  Whitney W Stevens; Robert J Lee; Robert P Schleimer; Noam A Cohen
Journal:  J Allergy Clin Immunol       Date:  2015-12       Impact factor: 10.793

Review 8.  Risk of malignancy associated with head and neck CT in children: a systematic review.

Authors:  Jenny X Chen; Bart Kachniarz; Sapideh Gilani; Jennifer J Shin
Journal:  Otolaryngol Head Neck Surg       Date:  2014-07-22       Impact factor: 3.497

Review 9.  Epidemiology and economic impact of rhinosinusitis.

Authors:  Vijay K Anand
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  2004-05

10.  The risk of cataract associated with repeated head and neck CT studies: a nationwide population-based study.

Authors:  Mei-Kang Yuan; Der-Chong Tsai; Shih-Chieh Chang; Mei-Chun Yuan; Shu-Ju Chang; Huan-Wu Chen; Hsin-Bang Leu
Journal:  AJR Am J Roentgenol       Date:  2013-09       Impact factor: 3.959

  10 in total
  1 in total

1.  On the optimization of bone SPECT/CT in terms of image quality and radiation dose.

Authors:  Monika Tulik; Piotr Tulik; Teresa Kowalska
Journal:  J Appl Clin Med Phys       Date:  2020-10-27       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.