Georg A Weidlich1, Fred Hacker2, David Bellezza3, Patrick Maguire4, Edward A Gardner5.
Abstract
The work described here compared the available technical solutions for the treatment of ventricular tachycardia with stereotactic body radiation therapy. Due to the complexity of target motion during cardiac and pulmonary motion as well as the several proximate radio-sensitive structures of the tracheobronchial tree and esophagogastrointestinal tract, four potential candidates for this treatment were identified: Accuray CyberKnife (Accuray Incorporated, Sunnyvale, California, United States), Varian TrueBeam (Varian Medical Systems, Palo Alto, California, USA), Elekta Infinity (Elekta, Stockholm, Sweden), and Varian Edge (Varian Medical Systems, Palo Alto, California, USA). All four treatment modalities were evaluated for their ability to deliver a conformal, homogeneous dose to most of the target volume, to spare nearby and distant critical and sensitive anatomical structures as well as for treatment efficiency. It was found that conventional linear accelerator technology was superior in their ability to spare distant critical structures and deliver treatments efficiently while the CyberKnife showed superiority in sparing nearby critical structures more aggressively by creating larger dose gradients at the periphery of the target volume. Both treatment modalities were similar in their ability to cover the entire target with the prescription dose, conform that dose to the target volume, and deliver a homogeneous dose.Entities:
Keywords: cardiac radiosurgery; cyberknife™; elekta infinity; varian edge; varian truebeam; ventricular tachycardia
Year: 2018 PMID: 30555760 PMCID: PMC6294279 DOI: 10.7759/cureus.3445
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Figure 1CardioPlan™ patient contouring image demonstrating the yellow target created by cardiologists, on either side of the proximal coronary arteries. Contours are sent to the treatment planning system via a DICOM file.
DICOM: digital imaging and communications in medicine
Figure 2Accuray CyberKnife Dose Distribution with PTV and prescription isodose line in orange
Accuray Incorporated, Sunnyvale, California, United States
PTV: planning target volume
Figure 3Varian TrueBeam dose distribution with PTV in orange and prescription isodose line in yellow
Varian Medical Systems, Palo Alto, California, USA
PTV: planning target volume
Summary of optimization objectives for CyberKnife, Elekta, and Varian linear accelerator plans
CyberKnife: Accuray Incorporated, Sunnyvale, California, United States; Elekta: Elekta, Stockholm, Sweden; Varian: Varian Medical Systems, Palo Alto, California, USA PTV: planning target volume; RT: right; LT: left; CA: coronary artery
| Structure | Criteria | Dose Objective (Gy) |
| PTV | Min Dose | 25.0 |
| PTV | Min DVH D95 | 26.5 |
| PTV | Max Dose | 32.0 |
| RT Coronary Artery | Max Dose | 19.0 |
| Ramus Coronary Artery | Max Dose | 19.0 |
| LT Anterior Descending CA | Max Dose | 19.0 |
| Mitral Valve | Max Dose | 19.0 |
| Pericardium | Max Dose | 19.0 |
| LT Main CA | Max Dose | 19.0 |
| Circumflex | Max Dose | 19.0 |
| LT Bronchus | Max Dose | 20.0 |
| Defibrillator | Max Dose | 2.0 |
Comparison of treatment parameters for Accuray CyberKnife, Varian TrueBeam with the ITV approach and gating, TrueBeam, Elekta no margin plans, and Varian Edge with ITV, gating, and no margin plans. The most desirable values/characteristics are in bold
CyberKnife: Accuray Incorporated, Sunnyvale, California, United States; Elekta: Elekta, Stockholm, Sweden; Varian: Varian Medical Systems, Palo Alto, California, USA ITV: internal target volume; PTV: planning target volume; CTV: clinical target volume; VMAT: volumetric arc therapy; CT: computed tomography; MLC: multi-leaf collimator; CBCT: cone-beam CT; CCC: collapsed cone convolution; AAA: analytical anisotropic algorithm
| Criteria/ Technology | Accuray CyberKnife w/ fixed collimator | Accuray CyberKnife w/ MLC | Varian TrueBeam (ITV) | Varian TrueBeam (Gated) | Varian TrueBeam (no expansion) | Elekta Infinity (no exp) | Varian Edge (ITV) | Varian Edge (Gated) | Varian Edge (no expansion) |
| Rx Dose (Gy) | 25 | 25 | 25 | 25 | 25 | 25 | 25 | 25 | 25 |
| Energy (MV) | 6 | 6 | 6 | 6 | 6 | 6 | 6 FFF | 6 FFF | 6 FFF |
| Dose Rate (MU/min) | 1000 | 1000 | 600 | 600 | 600 | 600 | 1400 | 1400 | 1400 |
| Maximum Dose (Gy) | 31.25 | 28.74 | 34.71 | 34.15 | 32.12 | 30.74 | 32.47 | 32.09 | 31.39 |
| CT slice spacing (mm) | 1.25 | 1.25 | 1.25 | 1.25 | 1.25 | 1.25 | 1.25 | 1.25 | 1.25 |
| CTV/PTV (ccm) | 19.80/32.55 | 19.69/31.79 | 19.80/63.25 | 19.80/51.45 | 19.80/32.55 | 19.80/32.55 | 19.4/59.4 | 19.4/50.0 | 19.40/31.5 |
| Rx Isodose Line (%) | 80 | 87 | 74 | 74 | 80 | 89 | 77 | 78 | 80 |
| PTV coverage (%) | 95.7 | 95.8 | 91.0 | 92.6 | 93.2 | 93.1 | 95 | 95 | 95 |
| CI/nCI | 1.64/1.72 | 1.37/1.44 | 1.59/1.72 | 1.56/1.69 | 1.44/1.57 | 1.31/1.43 | 0.99/1.10 | 0.98/1.08 | 0.97/1.07 |
| Treatment Modality | Non-isocentric, non-coplanar, robotic | Non-isocentric, non-coplanar, robotic | VMAT | VMAT | VMAT | VMAT | VMAT | VMAT | VMAT |
| Tracking/Gating | Synchrony Tracking | Synchrony Tracking | Neither | Gating | Neither | Neither | Neither | Gating | Neither |
| Collimator | Fixed Collimator | 2-3mm MLC | 5mm MLC | 5 mm MLC | 5 mm MLC | 5 mm MLC | 2.5 mm MLC | 2.5 mm MLC | 2.5 mm MLC |
| Beam Direction | Full-Path, ~2 Pi solid angle | Full-Path, ~2 Pi solid angle | Coplanar, continuous arc, 115 segments | Coplanar, continuous arc, 115 segments | Coplanar, continuous arc, 115 segments | Coplanar, continuous arc, 115 segments | Coplanar, 3 continuous 358 degree arcs | Coplanar, 3 continuous 358 degree arcs | Coplanar, 3 continuous 358 degree arcs |
| Image Guidance | Orthogonal, planar, oblique, kV, intra-fraction | Orthogonal, planar, oblique, kV, intra-fraction | CBCT, pre-fraction | CBCT pre-fraction | CBCT, pre-fraction | CBCT, pre-fraction | CBCT, pre-fraction | CBCT, pre-fraction | CBCT, pre-fraction |
| Monitor Units | 25187.8 | 11725.6 | 6767.4 | 6704.8 | 6400.7 | 7598.2 | 11511 | 12054 | 11790 |
| Estimated Tx Time (w/o gating) (min) | 95 | 34 | 9.0 | 29.7 | 9.5 | 11.3 | 8.2 | 8.6 | 8.4 |
| Dose Calc Algorithm | Raytracing | Pencil Beam | CCC | CCC | CCC | CCC | AAA | AAA | AAA |
Dosimetric comparison of Accuray CyberKnife, Varian TrueBeam with ITV and gating approach, TrueBeam, Elekta no margin plans, Varian Edge with ITV, gating, and no margin plans – most desirable values are in bold. Dose maximum was reported as D0.001 ccm.
CyberKnife: Accuray Incorporated, Sunnyvale, California, United States; Elekta: Elekta, Stockholm, Sweden; Varian: Varian Medical Systems, Palo Alto, California, USA ITV: internal target volume; PTV: planning target volume; CTV: clinical target volume; CT: computed tomography; MLC: multi-leaf collimator; RCA: right coronary artery; CA: coronary artery; LAD: Left anterior descending
| Anatomical Structure | Accuray CyberKnife w/ cones | Accuray CyberKnife w/ MLC | Varian TrueBeam (ITV) | Varian TrueBeam (Gated) | Varian TrueBeam (No expansion) | Elekta Infinity (No expansion) | Varian Edge (ITV) | Varian Edge (Gated) | Varian Edge (No expansion) |
| PTV mean (Gy) | 28.31 | 27.18 | 30.41 | 30.26 | 28.85 | 27.85 | 28.17 | 27.84 | 27.22 |
| RCA max (Gy) | 12.24 | 2.98 | 7.15 | 7.27 | 6.11 | 8.91 | 9.61 | 9.37 | 8.03 |
| Ramus CA max (Gy) | 21.32 | 28.12 | 25.56 | 24.59 | 22.10 | 21.76 | 24.88 | 24.22 | 20.23 |
| LAD CA max (Gy) | 21.23 | 21.88 | 24.86 | 23.71 | 22.09 | 20.93 | 24.08 | 22.76 | 18.95 |
| L Main CA max (Gy) | 14.63 | 13.61 | 22.78 | 21.65 | 19.57 | 18.11 | 14.79 | 13.86 | 11.77 |
| Circ max (Gy) | 21.35 | 23.18 | 24.18 | 23.11 | 21.07 | 21.06 | 24.26 | 23.08 | 20.36 |
| Mitral Valve max (Gy) | 19.26 | 22.18 | 29.62 | 28.39 | 25.46 | 24.07 | 25.86 | 25.11 | 21.62 |
| Pericardium max (Gy) | 31.25 | 28.74 | 34.71 | 34.15 | 32.05 | 30.71 | 32.36 | 31.90 | 31.21 |
| Full myocar CTV mean (Gy) | 31.25 | 27.36 | 31.32 | 31.15 | 32.05 | 30.71 | 29.09 | 28.59 | 27.71 |
| Esophagus max (Gy) | 5.21 | 4.14 | 10.48 | 9.62 | 8.02 | 10.89 | 4.47 | 4.16 | 3.53 |
| Left bronchus max (Gy) | 11.21 | 12.69 | 19.40 | 18.17 | 15.44 | 16.54 | 13.64 | 12.42 | 10.80 |
| Heart max (Gy) | 31.25 | 28.74 | 34.71 | 34.15 | 32.04 | 30.71 | 32.36 | 31.90 | 31.21 |
| Heart mean (Gy) | 6.61 | 5.25 | 9.08 | 8.14 | 6.58 | 6.01 | 5.77 | 5.20 | 4.07 |
| Total Lung mean (Gy) | 1.62 | 1.45 | 2.51 | 2.16 | 1.79 | 1.62 | 1.69 | 1.53 | 1.19 |
| Cord max (Gy) | 2.50 | 2.10 | 5.91 | 6.50 | 4.91 | 7.59 | 2.40 | 2.12 | 1.96 |
| Defib max (Gy) | 1.05 | 0.09 | 0.51 | 0.43 | 0.36 | 0.48 | 0.30 | 0.27 | 0.21 |