Development and evaluation of an arterial spin-labeling digital reference object for quality control and comparison of data analysis applications.

Alexander G Antolak1, Edward F Jackson.   

Abstract

Longitudinal assessment of quantitative imaging biomarkers (QIBs) requires a comprehensive quality control (QC) program to minimize bias and variance in measurement results. In addition, the availability of data analysis software from multiple vendors emphasizes the need for a means of quantitatively comparing the computed QIB measures produced by the applications. The purpose of this work is to describe a digital reference object (DRO) that has been developed for the evaluation of arterial spin-labeling (ASL) measurement results. The ASL DRO is a synthetic data set consisting of 10  ×  10 voxel square blocks with a range of ASL control image signal-to-noise ratio (SNRControl), blood flow (BF), and proton density (PD) image SNR values (SNRControl:1-100, BF:10-210 ml/100 g min-1, SNRPD:10-100). A pseudo-continuous ASL sequence was simulated with acquisition parameters and modeled signal intensities defined according to those typically associated with clinically-acquired ASL images. ASL parameters were estimated using the commercially-available nordicICE software package (NordicNeuroLab, Inc, Milwaukee, WI). Percent bias measures and Bland-Altman analyses demonstrated decreased bias and variance with increasing SNRControl and BF values. Excellent agreement with reference values was seen for all BF values above an SNRControl of 5 (concordance correlation coefficient greater than 0.92 for all SNRPD values). The ASL DRO developed in this work allows for the evaluation of software bias and variance across physiologically-meaningful BF and SNRControl values. Such studies are essential to the transition of quantitative ASL-based BF measurements into widespread clinical research applications, and ultimately, routine clinical care.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30540982     DOI: 10.1088/1361-6560/aaf83b

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  An Anthropomorphic Digital Reference Object (DRO) for Simulation and Analysis of Breast DCE MRI Techniques.

Authors:  Leah Henze Bancroft; James Holmes; Ryan Bosca-Harasim; Jacob Johnson; Pingni Wang; Frank Korosec; Walter Block; Roberta Strigel
Journal:  Tomography       Date:  2022-04-02

2.  Synthetic Arterial Spin Labeling MRI of the Kidneys for Evaluation of Data Processing Pipeline.

Authors:  Irène Brumer; Dominik F Bauer; Lothar R Schad; Frank G Zöllner
Journal:  Diagnostics (Basel)       Date:  2022-07-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.