Literature DB >> 30540877

There Is No Significant Difference in Fretting and Corrosion at the Trunnion of Metal and Ceramic Heads.

Young-Hoo Kim, Jang-Won Park, Jun-Shik Kim.   

Abstract

No study has compared the prevalence of fretting and corrosion at the trunnion of cobalt-chrome (Co-Cr) and zirconia ceramic heads in the same patients. The authors sought to compare the prevalence of fretting and corrosion at the trunnion after total hip arthroplasty with a 32-mm Co-Cr or a 32-mm zirconia ceramic head-on-polyethylene bearing. Isolated revision of the ace-tabular components was performed in 78 patients (156 hips) for polyethylene wear and osteolysis of the acetabulum. Seventy-eight Co-Cr head-titanium alloy stem pairs were compared with 78 zirconia ceramic head-titanium alloy stem pairs in the same patients. Using a visual scoring technique, the 156 head-stem pairs were analyzed for the prevalence of fretting and/or corrosion at the trunnion of the stem and the prevalence of metal transfer to the inner taper of the femoral head. Mean patient age was 48 years, and there were 65 men and 13 women. There was no trunnionosis in either group. Prevalence of fretting (81% vs 83%) and corrosion (4% vs 3%) at the trunnion was not significantly different (P=.518 vs .612, respectively) between the 2 groups. The median fretting scores (1.4±1.1 vs 1.2±1.4 points) and corrosion scores (1.2±0.8 vs 1.3±0.9 points) at the trunnion were not significantly different (P=.893 vs .781, respectively) between the 2 groups. Prevalence of metal transfer to the inner taper of the femoral head (8% vs 10%) and the median metal transfer scores (1.2±0.9 vs 1.4±1.1 points) were not significantly different (P=.213 vs .729, respectively) between the 2 groups. [Orthopedics. 2019; 42(1):e99-e103.]. Copyright 2018, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30540877     DOI: 10.3928/01477447-20181206-03

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  1 in total

1.  Response to letter to the editor on "Titanium neck-titanium stem taper corrosion in a modular neck stem".

Authors:  Sumon Nandi; Ronit Shah; Carl Talmo
Journal:  Arthroplast Today       Date:  2020-03-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.