Analytical validation of a conventional cardiac troponin I assay for dogs and cats.

Rebecca Langhorn1, Jacob D Yrfelt1, Claus S Stjernegaard1, Liselotte B Christiansen2, Lisbeth H Olsen2, Lise N Nielsen1.   

Abstract

BACKGROUND: Cardiac troponins are gold-standard biomarkers of myocardial injury. There is a need for validation of assays with higher availability and lower costs in veterinary medicine.
OBJECTIVES: The primary aim of the present study was to perform an analytical validation of the IMMULITE 2000 TnI assay for use in dogs and cats. A secondary aim was to evaluate its agreement with the previously validated and sensitive Siemens ADVIA Centaur TnI-Ultra assay.
METHODS: Intra- and inter-assay variation, detection limits, the linearity under dilution, and a sample addition study (modified spike-and-recovery analysis) were investigated to assess analytical performance in 15 canine and 15 feline serum samples. Agreement between the assays was evaluated by correlation and Bland-Altman analyses including an additional 99 canine serum samples.
RESULTS: Intra-assay variation of cTnI in canine and feline serum was 3.71% and 4.68%, while inter-assay variation was 5.88% and 6.54%, respectively. The assay performed with acceptable linearity within a clinically relevant range of serum cTnI concentrations. The sample addition study revealed insufficient recovery in the range of 71.9%-81.4% for dogs and 62.6%-75.7% for cats. This was considered to be due to a negative matrix effect. A significant correlation between the assays was found, and the Bland-Altman analysis showed acceptable agreement for a wide range of concentrations, but revealed a proportional error, with the IMMULITE TnI assay consistently measuring a higher concentration than the Centaur TnI-Ultra assay. This was relevant only at high serum cTnI concentrations.
CONCLUSIONS: The IMMULITE TnI assay is considered acceptable for clinical use in dogs and cats.
© 2018 American Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  biomarker; companion animals; immunoassay; myocardial injury

Mesh:

Substances:

Year:  2018        PMID: 30536941     DOI: 10.1111/vcp.12681

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  2 in total

1.  Validation of a commercial human ELISA to measure hyaluronic acid concentration in feline plasma.

Authors:  Kaela E Shaw; Alexa M Bersenas; Shane W Bateman; Shauna L Blois; R Darren Wood
Journal:  J Vet Diagn Invest       Date:  2021-09-13       Impact factor: 1.279

2.  Serum paraoxonase 1 activity in cats: analytical validation, reference intervals, and correlation with serum amyloid A and alpha-1-acid glycoprotein.

Authors:  Gabriele Rossi; Sara Meazzi; Alessia Giordano; Saverio Paltrinieri
Journal:  J Vet Diagn Invest       Date:  2020-08-17       Impact factor: 1.279

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.