Literature DB >> 30531495

Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO320 for public use according to the 2010 European Society of Hypertension International Protocol.

Anastasios Kollias1, Emelina Stambolliu, Konstantinos G Kyriakoulis, Stamatis S Papadatos, George S Stergiou.   

Abstract

OBJECTIVE: The aim of this study was to assess the accuracy of the single-cuff oscillometric blood pressure (BP) monitor InBody BPBIO320 developed for self-measurement by adults in public spaces according to the 2010 European Society of Hypertension International Protocol (ESH-IP). PARTICIPANTS AND METHODS: Adults were recruited to fulfill the ESH-IP requirements. The same-arm sequential BP measurement method was used.
RESULTS: A total of 38 participants were recruited and 33 were included in the analysis. The average difference between test and reference BP measurements was-0.4±5.3 (SD)/-0.5±4.6 mmHg (systolic/diastolic). The number of test-reference BP differences within 5, 10, and 15 mmHg was 73, 91, and 98 for systolic BP and 79, 93, and 99 for diastolic. The number of participants with at least two of their three BP differences within 5 mmHg was 28 for both systolic and diastolic BPs. The number of participants with none of their three BP differences within 5 mmHg was 1 for both systolic and diastolic BPs.
CONCLUSION: The InBody BPBIO320 device fulfilled the requirements of the ESH-IP validation protocol and can be recommended for clinical use in adults.

Entities:  

Mesh:

Year:  2019        PMID: 30531495     DOI: 10.1097/MBP.0000000000000359

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  2 in total

1.  Be Healthe for Your Heart: Protocol for a Pilot Randomized Controlled Trial Evaluating a Web-Based Behavioral Intervention to Improve the Cardiovascular Health of Women With a History of Preeclampsia.

Authors:  Rachael Taylor; Vanessa A Shrewsbury; Lisa Vincze; Linda Campbell; Robin Callister; Felicity Park; Tracy Schumacher; Clare Collins; Melinda Hutchesson
Journal:  Front Cardiovasc Med       Date:  2019-09-26

2.  The correlation between dietary fat quality indices and lipid profile with Atherogenic index of plasma in obese and non-obese volunteers: a cross-sectional descriptive-analytic case-control study.

Authors:  Maryam Sadat Moussavi Javardi; Zahra Madani; Ariyo Movahedi; Majid Karandish; Behnood Abbasi
Journal:  Lipids Health Dis       Date:  2020-09-26       Impact factor: 3.876

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.