Literature DB >> 30527292

Improved visualization of hypodense liver lesions in virtual monoenergetic images from spectral detector CT: Proof of concept in a 3D-printed phantom and evaluation in 74 patients.

N Große Hokamp1, V C Obmann2, R Kessner2, K R Laukamp2, T Persigehl3, S Haneder3, N Ramaiya4.   

Abstract

OBJECTIVES: The well-known boost of iodine associated-attenuation in low-keV virtual monoenergetic images (VMI_low) is frequently used to improve visualization of lesions and structures taking up contrast media. This study aimed to evaluate this concept in reverse. Hence to investigate if increased attenuation within the liver allows for improved visualization of little or not-enhancing lesions.
METHODS: A 3D-printed phantom mimicking the shape of a human liver exhibiting a lesion in its center was designed and printed. Both, parenchyma- and lesion-mimic were filled with different solutions exhibiting 80/100/120HU and 0/15/40/60HU, respectively. Further, a total of 74 contrast-enhanced studies performed on a spectral detector CT scanner (SDCT) were included in this retrospective study. Patients had MRI or follow-up proven cysts and/or hypodense metastases. VMI of 40-200 keV as well as conventional images (CI) were reconstructed. ROI were placed in lesion and parenchyma(-mimics) on CI and transferred to VMI. Signal- and contrast-to-noise ratio were calculated (S-/CNR). Further, two radiologists independently evaluated image quality. Data was statistically assessed using ANOVA or Wilcoxon-test.
RESULTS: In phantoms, S/CNR was significantly higher in VMI_low. The cyst-mimic in highly attenuating parenchyma-mimic on CI yielded a CNR of 6.4 ± 0.8; using VMI_40 keV, mildly hypodense lesion-mimic in poorly attenuating parenchyma-mimic exhibited a similar CNR (5.8 ± 0.9; p ≤ 0.05). The same tendency was observed in patients (cyst in CI/metastasis in VMI_40 keV: 4.4 ± 1.2/3.9 ± 1.8; p ≤ 0.05). Qualitative analysis indicated a benefit of VMI_40 keV (p ≤ 0.05).
CONCLUSIONS: VMI_low from SDCT allow for an improved visualization of hypodense focal liver lesions exploiting the concept of contrast blooming in reverse.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  3D-printing; Dual energy computed tomography; Liver lesions; Spectral detector computed tomography; Virtual monoenergetic images

Mesh:

Substances:

Year:  2018        PMID: 30527292     DOI: 10.1016/j.ejrad.2018.11.001

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  7 in total

1.  Optimal Conspicuity of Liver Metastases in Virtual Monochromatic Imaging Reconstructions on a Novel Photon-Counting Detector CT-Effect of keV Settings and BMI.

Authors:  Stefanie Bette; Josua A Decker; Franziska M Braun; Judith Becker; Mark Haerting; Thomas Haeckel; Michael Gebhard; Franka Risch; Piotr Woźnicki; Christian Scheurig-Muenkler; Thomas J Kroencke; Florian Schwarz
Journal:  Diagnostics (Basel)       Date:  2022-05-14

Review 2.  Spectral detector CT applications in advanced liver imaging.

Authors:  Noor Fatima Majeed; Marta Braschi Amirfarzan; Christoph Wald; Jeremy R Wortman
Journal:  Br J Radiol       Date:  2021-05-28       Impact factor: 3.629

3.  Value of virtual monochromatic spectral image of dual-layer spectral detector CT with noise reduction algorithm for image quality improvement in obese simulated body phantom.

Authors:  Hyo-Jin Kang; Jeong Min Lee; Sang Min Lee; Hyun Kyung Yang; Ri Hyeon Kim; Ju Gang Nam; Aruna Karnawat; Joon Koo Han
Journal:  BMC Med Imaging       Date:  2019-08-28       Impact factor: 1.930

4.  Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study.

Authors:  Jasmin A Holz; Hatem Alkadhi; Kai R Laukamp; Simon Lennartz; Carola Heneweer; Michael Püsken; Thorsten Persigehl; David Maintz; Nils Große Hokamp
Journal:  Sci Rep       Date:  2020-12-09       Impact factor: 4.379

5.  Virtual monoenergetic images from spectral detector computed tomography facilitate washout assessment in arterially hyper-enhancing liver lesions.

Authors:  R P Reimer; N Große Hokamp; A Fehrmann Efferoth; A Krauskopf; D Zopfs; J R Kröger; T Persigehl; D Maintz; A C Bunck
Journal:  Eur Radiol       Date:  2020-11-12       Impact factor: 5.315

6.  Value of spectral detector computed tomography for the early assessment of technique efficacy after microwave ablation of hepatocellular carcinoma.

Authors:  Robert Peter Reimer; Nils Große Hokamp; Julius Niehoff; David Zopfs; Simon Lennartz; Mariam Heidar; Roger Wahba; Dirk Stippel; David Maintz; Daniel Pinto Dos Santos; Christian Wybranski
Journal:  PLoS One       Date:  2021-06-15       Impact factor: 3.240

7.  Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma.

Authors:  Yasunori Nagayama; Shota Tanoue; Taihei Inoue; Seitaro Oda; Takeshi Nakaura; Daisuke Utsunomiya; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2019-07-16       Impact factor: 5.315

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.