Localization accuracy of two electromagnetic tracking systems in prostate cancer radiotherapy: A comparison with fiducial marker based kilovoltage imaging.

A Vanhanen1, H Syrén2, M Kapanen3.   

Abstract

The aim of this study was to evaluate the localization accuracy of electromagnetic (EM) tracking systems RayPilot (Micropos Medical AB) and Calypso (Varian Medical Systems) in prostate cancer radiotherapy. The accuracy was assessed by comparing couch shifts obtained with the EM methods to the couch shifts determined by simultaneous fiducial marker (FM) based orthogonal kilovoltage (kV) imaging. Agreement between the methods was compared using Bland-Altman analysis. Interfractional positional stability of the FMs, RayPilot transmitters and Calypso transponders was investigated. 582 fractions from 22 RayPilot patients and 335 fractions from 26 Calypso patients were analyzed. Mean (± standard deviation (SD)) differences between RayPilot and kV imaging were 0.3 ± 2.2, -2.2 ± 2.4 and -0.0 ± 1.0 mm in anterior-posterior (AP), superior-inferior (SI) and left-right (LR) directions, respectively. Corresponding 95% limits of agreement (LOA) were ±4.3, ±4.7 and ±2.1 mm around the mean. Mean (±SD) differences between Calypso and kV imaging were -0.2 ± 0.6, 0.1 ± 0.5 and -0.1 ± 0.4 mm in AP, SI and LR directions, respectively, and corresponding LOAs were ±1.3, ±1.0 and ±0.8 mm around the mean. FMs and transponders were stable: SD of intermarker and intertransponder distances was 0.5 mm. Transmitters were unstable: mean caudal transmitter shift of 1.8 ± 2.0 mm was observed. Results indicate that the localization accuracy of the Calypso is comparable to kV imaging of fiducials and the methods could be used interchangeably. The localization accuracy of the RayPilot is affected by transmitter instability and the positioning of the patient should be verified by other setup techniques. The study is part of clinical trial NCT02319239.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Calypso; Fiducial markers; Interfraction motion; Intrafraction motion; Prostate; Radiotherapy; RayPilot; kV-imaging

Mesh:

Substances:

Year:  2018        PMID: 30527084     DOI: 10.1016/j.ejmp.2018.11.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Utilizing the TrueBeam Advanced Imaging Package to monitor intrafraction motion with periodic kV imaging and automatic marker detection during VMAT prostate treatments.

Authors:  Mark C Korpics; Michelle Rokni; Michael Degnan; Bulent Aydogan; Stanley L Liauw; Gage Redler
Journal:  J Appl Clin Med Phys       Date:  2020-01-24       Impact factor: 2.102

Review 2.  Target motion management in breast cancer radiation therapy.

Authors:  Elham Piruzan; Naser Vosoughi; Seied Rabi Mahdavi; Leila Khalafi; Hojjat Mahani
Journal:  Radiol Oncol       Date:  2021-10-08       Impact factor: 2.991

3.  Triggered kV Imaging During Spine SBRT for Intrafraction Motion Management.

Authors:  Jihye Koo; Louis Nardella; Michael Degnan; Jacqueline Andreozzi; Hsiang-Hsuan M Yu; Jose Penagaricano; Peter A S Johnstone; Daniel Oliver; Kamran Ahmed; Stephen A Rosenberg; Evan Wuthrick; Roberto Diaz; Vladimir Feygelman; Kujtim Latifi; Eduardo G Moros; Gage Redler
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

4.  Intrafraction Prostate Motion Management During Dose-Escalated Linac-Based Stereotactic Body Radiation Therapy.

Authors:  Denis Panizza; Valeria Faccenda; Raffaella Lucchini; Martina Camilla Daniotti; Sara Trivellato; Paolo Caricato; Valerio Pisoni; Elena De Ponti; Stefano Arcangeli
Journal:  Front Oncol       Date:  2022-04-07       Impact factor: 5.738

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.