Literature DB >> 30523946

Monte Carlo-based SPECT reconstruction within the SIMIND framework.

Johan Gustafsson1, Gustav Brolin, Michael Ljungberg.   

Abstract

This paper presents the development and validation of a Monte Carlo-based singe photon emission computed tomography reconstruction program for parallel-hole collimation contained within the SIMIND Monte Carlo framework. The Monte Carlo code is used as an accurate forward-projector and is combined with a simplified back-projector to perform iterative tomographic reconstruction using the Maximum Likelihood Expectation Maximization and Ordered Subsets Expectation Maximization algorithms, together forming a program called SIMREC. The Monte Carlo simulation transforms the estimated source distribution directly from activity to counts in its projections. Hence, the reconstructed image is expressed in activity without reference to an external calibration. The program is tested using phantom measurements of spheres filled with 99mTc, 177Lu and 131I placed in air and centrally and peripherally in a water-filled elliptical phantom. The feasibility of applying the reconstruction to patients is also demonstrated for a range of radiopharmaceuticals. The deviation in total activity in the spheres ranged between  -4.1% and 6.2% compared with the activity determined when preparing the phantom. The SIMREC program was found to be accurate with respect to activity estimation and to reconstruct visually acceptable images within a few hours when applied to patient examples.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30523946     DOI: 10.1088/1361-6560/aaf0f1

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Deep-Learning Generation of Synthetic Intermediate Projections Improves 177Lu SPECT Images Reconstructed with Sparsely Acquired Projections.

Authors:  Tobias Rydén; Martijn Van Essen; Ida Marin; Johanna Svensson; Peter Bernhardt
Journal:  J Nucl Med       Date:  2020-08-28       Impact factor: 10.057

2.  Monte Carlo modelling of a compact CZT-based gamma camera with application to 177Lu imaging.

Authors:  Daniel Roth; Erik Larsson; Michael Ljungberg; Katarina Sjögreen Gleisner
Journal:  EJNMMI Phys       Date:  2022-05-08

3.  Establishment of a clinical SPECT/CT protocol for imaging of 161Tb.

Authors:  I Marin; T Rydèn; M Van Essen; J Svensson; N Gracheva; U Köster; J R Zeevaart; N P van der Meulen; C Müller; P Bernhardt
Journal:  EJNMMI Phys       Date:  2020-07-01
  3 in total