| Literature DB >> 30518238 |
Christopher F Tirotta1, Richard G Lagueruela1, Danielle Madril1, Daria Salyakina2, Weize Wang2, Thomas Taylor2, Jorge Ojito3, Kathleen Kubes3, Hyunsoo Lim3, Robert Hannan3, Redmond Burke3.
Abstract
This study evaluated whether rotational thromboelastometry (ROTEM; Tem International GmbH, Munich, Germany) FIBTEM maximum clot firmness (MCF) can be used to predict plasma fibrinogen level in pediatric patients undergoing cardiac surgery. Linear regression was conducted to predict plasma fibrinogen level using FIBTEM MCF (0.05 level of significance). Scatter plot with the regression line for the model fit was created. Fifty charts were retrospectively reviewed, and 87 independent measurements of FIBTEM MCF paired with plasma fibrinogen levels were identified for analysis. Linear regression analysis suggested a significant positive linear relationship ( P < .0001) between plasma fibrinogen levels and MCF. Both MCF intercept and slope were significantly correlated with fibrinogen level ( P < .0001). The estimated regression equation (predicted fibrinogen = 78.6 + 12.4 × MCF) indicates that a 1-mm increase in MCF raises plasma fibrinogen level by an average of 12.4 mg/dL. The statistically significant positive linear relationship observed between MCF and fibrinogen levels ( P < .001) suggests that MCF can be used as a surrogate for fibrinogen level. This relationship is of clinical relevance in the calculation of patient-specific dosing of fibrinogen supplementation in this setting.Entities:
Keywords: ROTEM; RiaSTAP; cardiac surgery; fibrinogen; infants; neonates
Mesh:
Substances:
Year: 2018 PMID: 30518238 PMCID: PMC6714911 DOI: 10.1177/1076029618816382
Source DB: PubMed Journal: Clin Appl Thromb Hemost ISSN: 1076-0296 Impact factor: 2.389
Descriptive Statistics of the Plasma Fibrinogen Level and FIBTEM MCF (N = 87).
| Variable | N | Mean | Standard Deviation | Minimum | Maximum | Median |
|---|---|---|---|---|---|---|
| Plasma fibrinogen level (mg/dL) | 87 | 178.1 | 76.4 | 33.0 | 448.0 | 173.0 |
| FIBTEM MCF (mm) | 87 | 8.0 | 4.3 | 2.0 | 21.0 | 8.0 |
Figure 1.Fit plot of linear regression using FIBTEM maximum clot firmness to predict plasma fibrinogen levels (N = 87).
Results from Repeated 10-fold Cross Validation of Linear Regression Using FIBTEM MCF to Predict Plasma Fibrinogen Level.
| Parameter | Mean Estimated Correlation Coefficient | Mean SE | Mean 95% Confidence Limits |
|
|---|---|---|---|---|
| Intercept | 78.6 | 13.2 | (52.8-104.4) | <.001 |
| FIBTEM MCF (mm) | 12.4 | 1.4 | (9.6-15.2) | <.001 |
Figure 2.Repeated 10-fold cross validation results of linear regression. MCF indicates maximum clot firmness.