Mohammad Shafiul Alam1, Mohammad Golam Kibria1, Nusrat Jahan1, Ric N Price2,3, Benedikt Ley4.
Abstract
OBJECTIVES: Spectrophotometry kits from Pointe Scientific (PS; USA) were compared to kits from Trinity Biotech (Trinity; Ireland) in 50 venous blood samples from purposively selected individuals in Bangladesh. Repeatability and inter-assay variability were assessed by Students t-test, Bland-Altman plot and Pearson correlation coefficient (r). The median glucose-6-phosphate dehydrogenase (G6PD) activity of all G6PD normal participants was calculated per assay and defined as 100% activity. Performance was calculated considering 30% and 70% cut off activities and Trinity as reference.Entities:
Keywords: G6PD; Malaria; Spectrophotometry
Mesh:
Substances:
Year: 2018 PMID: 30514365 PMCID: PMC6278131 DOI: 10.1186/s13104-018-3964-7
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Fig. 1Scatter plot mean Trinity Biotech (Trinity) results vs. mean Pointe Scientific (PS) results. Red lines from origin outwards indicate 10%, 30%, 70% and 100% cut off activity
Distribution of categorized result / test kits
| Trinity | ||||||
|---|---|---|---|---|---|---|
|
| < 10%a | 10% to < 30%a | 30% to < 70%a | > 70%a | Total | |
| < 10%b | 2 | 0 | 0 | 0 | 2 | |
| 10% to < 30%b | 1 | 5 | 0 | 0 | 5 | |
| 30% to < 70%b | 0 | 0 | 9 | 1 | 10 | |
| > 70%b | 0 | 0 | 0 | 33 | 33 | |
| Total | 2 | 5 | 9 | 34 | 50 | |
a 100% G6PD activity Trinity: 9.6 U/gHb
b 100% G6PD activity Pointe Scientific: 10.1 U/gHb