Literature DB >> 30506961

Ability to detect endodontic complications using three different cone beam computed tomography units with and without artefact reduction modes: an ex vivo study.

C Koç1, K Kamburoğlu2, G Sönmez2, F Yılmaz3, O Gülen4, S Karahan5.   

Abstract

AIM: To assess observer performance in detecting endodontic complications using three different cone beam computed tomography (CBCT) units with and without the application of artefact reduction modes.
METHODOLOGY: The study involved 40 freshly extracted human mandibular teeth (n = 10 per group) and divided randomly into four endodontic complication groups. Group 1) Instrument fracture; Group 2) Strip perforation; Group 3) Canal underfilling; and Group 4) Canal overfilling. Images of each tooth were obtained using three different CBCT units offering artefact reduction algorithms: the ProMax 3D Max, the Pax Flex 3D and the Dentri S. Four observers evaluated the images for the presence/absence of the four simulated endodontic complications. Weighted kappa coefficients and intra-class correlation coefficients (ICCs) were calculated to reveal the intra- and inter-observer agreement for each imaging mode, respectively. Receiver operating characteristic (ROC) analysis was used to evaluate the observers' performance. DeLong tests were used to compare the results for each image mode and observer using a significance level of α = 0.05.
RESULTS: In each of the four simulated endodontic complication groups, no significant differences were observed with and without application of artefact reduction for any of the three CBCT units tested. Only two significant differences were detected, and both were between the ProMax 3D Max at low mode AR and ProMax 3D Max without AR: observer 2 in group 1 (P = 0.0001) and observer 4 in group 4 (P = 0.0256).
CONCLUSION: For each of the three CBCT units tested, application of artefact reduction for detecting endodontic complications is not recommended as a routine tool.
© 2018 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  artefact reduction algorithms; cone beam computed tomography; endodontic complications

Mesh:

Year:  2018        PMID: 30506961     DOI: 10.1111/iej.13051

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  1 in total

1.  Mapping the expression of beam hardening artefacts produced by metal posts positioned in different regions of the dental arch.

Authors:  Rocharles C Fontenele; Amanda Farias Gomes; Lucas P Lopes Rosado; Frederico S Neves; Deborah Q Freitas
Journal:  Clin Oral Investig       Date:  2020-08-09       Impact factor: 3.573

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.