Literature DB >> 30499728

Structure and motion design of a mock circulatory test rig.

Yuhui Shi1, Theodosios Korakianitis2, Zhongjian Li3, Yubing Shi4.   

Abstract

Mock circulatory test rig (MCTR) is the essential and indispensable facility in the cardiovascular in vitro studies. The system configuration and the motion profile of the MCTR design directly influence the validity, precision, and accuracy of the experimental data collected. Previous studies gave the schematic but never describe the structure and motion design details of the MCTRs used, which makes comparison of the experimental data reported by different research groups plausible but not fully convincing. This article presents the detailed structure and motion design of a sophisticated MCTR system, and examines the important issues such as the determination of the ventricular motion waveform, modelling of the physiological impedance, etc., in the MCTR designing. The study demonstrates the overall design procedures from the system conception, cardiac model devising, motion planning, to the motor and accessories selection. This can be used as a reference to aid researchers in the design and construction of their own in-house MCTRs for cardiovascular studies.

Entities:  

Keywords:  test; Mock circulatory test rig; cardiovascular dynamics; motion design; structure design

Mesh:

Year:  2018        PMID: 30499728     DOI: 10.1080/03091902.2018.1543467

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  1 in total

1.  Establishment of a Modular Hemodynamic Simulator for Accurate In Vitro Simulation of Physiological and Pathological Pressure Waveforms in Native and Bioartificial Blood Vessels.

Authors:  Florian Helms; Axel Haverich; Mathias Wilhelmi; Ulrike Böer
Journal:  Cardiovasc Eng Technol       Date:  2021-09-23       Impact factor: 2.305

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.