Literature DB >> 30499327

The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis.

T R Turgeon1, T C Gascoyne2, E K Laende3, M J Dunbar3, E R Bohm1, C G Richardson3.   

Abstract

AIMS: The introduction of a novel design of total knee arthroplasty (TKA) must achieve outcomes at least as good as existing designs. A novel design of TKA with a reducing radius of the femoral component and a modified cam-post articulation has been released and requires assessment of the fixation to bone. Radiostereometric analysis (RSA) of the components within the first two postoperative years has been shown to be predictive of medium- to long-term fixation. The aim of this study was to assess the stability of the tibial component of this system during this period of time using RSA. PATIENTS AND METHODS: A cohort of 30 patients underwent primary, cemented TKA using the novel posterior stabilized fixed-bearing (ATTUNE) design. There was an even distribution of men and women (15:15). The mean age of the patients was 64 years (sd 8) at the time of surgery; their mean body mass index (BMI) was 35.4 kg/m 2 (sd 7.9). RSA was used to assess the stability of the tibial component at 6, 12, and 24 months compared with a six-week baseline examination. Patient-reported outcome measures were also assessed.
RESULTS: The mean maximum total point motion (MTPM) of the tibial component between 12 and 24 months postoperatively was 0.08 mm (sd 0.08), which is well below the published threshold of 0.2 mm (p < 0.001). Patient-reported outcome measures consistently improved.
CONCLUSION: The tibial component of this novel design of TKA showed stability between assessment 12 and 24 months postoperatively, suggesting an acceptably low risk of medium- to long-term failure due to aseptic loosening.

Entities:  

Keywords:  Radiostereometric analysis; Stability; Tibial fixation; Total knee arthroplasty

Mesh:

Substances:

Year:  2018        PMID: 30499327     DOI: 10.1302/0301-620X.100B12.BJJ-2018-0566.R1

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  4 in total

1.  Comparison of Existing and New Total Knee Arthroplasty Implant Systems From the Same Manufacturer: A Prospective, Multicenter Study.

Authors:  William G Hamilton; Ivan J Brenkel; Steven L Barnett; Paul W Allen; Kimberly A Dwyer; James P Lesko; Stephen R Kantor; Mark G Clatworthy
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2021-12-15

2.  Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study.

Authors:  Sebastian Jaeger; Marvin Eissler; Martin Schwarze; Mareike Schonhoff; J Philippe Kretzer; Rudi G Bitsch
Journal:  Bone Joint Res       Date:  2022-04       Impact factor: 4.410

3.  Tibial Baseplate-Cement Interface Debonding in the ATTUNE Total Knee Arthroplasty System.

Authors:  Daniel Torino; Christopher Damsgaard; David J Kolessar; Daniel S Hayes; Brian Foster; Jesse Constantino; Jove Graham
Journal:  Arthroplast Today       Date:  2022-09-20

4.  A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design.

Authors:  Bart L Kaptein; Peter den Hollander; Bregje Thomassen; Martha Fiocco; Rob G H H Nelissen
Journal:  Bone Joint J       Date:  2020-09       Impact factor: 5.082

  4 in total