Factors contributing to radiation dose for patients and operators during diagnostic cardiac angiography.

James A Crowhurst1,2,3, Mark Whitby1,2,4, Michael Savage1,2, Dale Murdoch1,2, Brendan Robinson1,3, Elizabeth Shaw1, Niranjan Gaikwad1, Ramkrishna Saireddy1,5, Karen Hay6, Darren L Walters1,2.   

Abstract

INTRODUCTION: Diagnostic coronary angiography (CA) uses ionising radiation with relatively high doses, which impact on both patients and staff. This study sought to identify which patient and procedural factors impact patient and operator dose the most during CA.
METHODS: Patient and procedure related variables impacting on Kerma area product (PKA ) and operator dose (OD) were collected for 16 months. Procedures were separated into 10 different procedure categories. PKA was used for patient dose and OD was measured with an instantly downloadable dosimeter (IDD) - downloaded at the end of each procedure. High and low radiation dose was defined by binary variables based on the 75th percentile of the continuous measures. Univariate and multivariate regression were used to identify predictors.
RESULTS: Of 3860 patients included, the IDD was worn for 2591 (61.7%). Obesity (BMI > 30 compared to BMI < 25) was the strongest predictor for both a PKA (odds ratio (OR) = 19.1 (95% CI 13.5-26.9) P < 0.001) and OD (OR = 3.3 (2.4-4.4) P < 0.001) above the 75th percentile. Male gender, biplane imaging, the X-ray unit used, operator experience and procedure type also predicted a high PKA . Radial access, male gender, biplane imaging and procedure type also predicted a high OD.
CONCLUSION: Radiation dose during CA is multifactorial and is dependent on patient and procedure related variables. Many factors impact on both PKA and OD but obesity is the strongest predictor for both patients and operators to receive a high radiation dose.
© 2018 The Prince Charles Hospital. Journal of Medical Radiation Sciences published by John Wiley & Sons Australia, Ltd on behalf of Australian Society of Medical Imaging and Radiation Therapy and New Zealand Institute of Medical Radiation Technology.

Entities:  

Keywords:  Body mass index; coronary angiography; fluoroscopy; quality assurance; radiation dose

Mesh:

Year:  2018        PMID: 30488575      PMCID: PMC6399189          DOI: 10.1002/jmrs.315

Source DB:  PubMed          Journal:  J Med Radiat Sci        ISSN: 2051-3895


Introduction

Cardiovascular disease is Australia's most common chronic disease and coronary artery disease is the leading cause of mortality.1 Coronary artery disease can be diagnosed and treated in the same setting with invasive cardiac angiography (CA). CA uses X‐ray fluoroscopy and carries risks to the patient undergoing these procedures.2 The side effects of radiation are both stochastic (neoplasm and heritable abnormalities) and tissue injury (deterministic).2 The associated radiation dose and relative risk of malignancy may not be conclusive3 but the deterministic effects of fluoroscopy procedures, presenting as skin injury have been reported many times4, 5, 6 and have been compiled into case review reports and recommendation documents.2, 7 The cardiologist who is closest to the X‐ray source and patient is exposed to the harmful effects of low energy scattered radiation during CA which has been highlighted in numerous studies and reports.8, 9, 10 The International Commission on Radiological Protection recommends that dosimeters should be available for all staff working in fluoroscopic laboratories. In addition, a quality assurance (QA) program should ensure the use of dosimeters with a review of abnormal dose values.11 Dosimeters, using thermo‐luminescent or more recently optically stimulated luminescent dosimetry techniques are required to be worn by catheter laboratory personnel. These are submitted for evaluation on a monthly basis. However, these do not demonstrate how radiation dose is received on a case by case basis and which factors increase a staff members’ radiation exposure more than others. Patient and operator dose have been investigated before, such as the REVERE trial12 and the RAD‐MATRIX trial,13 which investigated the impact of vascular access site on radiation dose. However, there is a paucity of data on this subject where all variables that may impact on operator dose are assessed together, and which variables are predictive of high doses. This study aimed to supplement the QA program and perform a more detailed investigation of radiation exposure to both patients and operators during contemporary diagnostic cardiac angiography procedures. This study sought to identify patient and procedural factors that may impact on patient and operator dose in addition to identifying which variables are predictive of a high patient and operator radiation dose.

Method

The study was a retrospective analysis of prospectively collected data for consecutive patients presenting to the cardiac catheterisation laboratories in this single, tertiary, teaching hospital that had diagnostic procedures performed. Data were collected from mid‐August 2014 through to mid December 2015. Approval for this study was granted by the facility human research ethics committee.

Radiation protection for operators

The examination table had lead drapes fitted to the tableside. Operators also had a lead acrylic, ceiling suspended shield. Operators all wore a protective apron of at least 0.5 mm lead equivalent at the front, thyroid shields, lead shin guards, lead eye glasses and some wore lead equivalent protective hats.

Workflow and equipment protocols

All procedures were performed in one of two identical biplane cardiac catheterisation laboratories (Siemens Axiom Artis dBc, Siemens Healthcare, Erlangen, Germany). Operators positioned (panned) the patient on the floating top tables and activated the fluoroscopy and digital acquisition. Radiographers operated the C‐arm, controlled pulse rate, dose per pulse, collimation and wedge filtration from an adjacent control room. Fluoroscopy was typically set to 7.5 pulses per second and the digital acquisition (DA) frame rate was set to 15 frames per second. A standard fluoroscopy and acquisition protocol was typically used for procedures in which the X‐ray system automatically adjusted copper filtration, kV and mA to achieve a detector dose that would produce an acceptable image. Radiation data collected were:Fluoroscopy time (FT), Kerma area product (PKA), skin entrance dose (air kerma – mGy) at the reference point (KAR) and number of DA. These were entered into an oracle database at the end of each procedure (Impax CV, Agfa Healthcare, Netherlands). PKA was the primary measure for patient dose and was calculated using the PKA meter in the X‐ray tube housing and is given in Gycm2. Both X‐ray systems were checked for accuracy of the PKA meter readings as part of their annual compliance tests. Individual elements for each procedure were entered into the database using a series of tick boxes and procedures were then grouped into 10 categories: Abdominal/femoral angiogram only (AFA). Coronary angiography only (CORS). Coronary angiography + aortogram/left ventriculogram (CORS‐LV/AO). Coronary angiography + left heart catheterisation + right heart catheterisation (CORS‐LHC/RHC). Coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR) – (CORS‐OCT/IVUS/FFR). Coronary angiography + pulmonary angiography (CORS‐PA) Coronary angiography + coronary artery bypass graft angiography (CORS‐GRAFTS). Pulmonary angiography alone (PA). Right heart catheterisation ± venography (RHC) Work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography (TAVI‐WU). The CORS group was used as the reference group for statistical analysis. Other procedural variables that could impact on radiation dose, such as catheter access route and whether biplane angiography was used were also recorded. Patient variables, including age and body mass index (BMI) were collected. BMI was grouped into three categories: <25, 25–30, and >30 kg/m2.

Operator dose analysis

The primary and secondary operators (where a second operator was present) were monitored using instantly downloadable personal dosimeters (IDD) (InstadoseTM, Mirion technologies, Georgia, USA) in addition to their usual TLD monitors. The IDD was worn on the thyroid collar, on the outside of the protective apron for the procedure duration and downloaded at the end of each procedure. The methodology for using these dosimeters has been previously described.14 The dosimeter is plugged into a personal computer (PC) to be read using the InstadoseTM software. The readout from the PC demonstrates the air kerma incident on the dosimeter with a conversion calculation to give the personal dose equivalent Hp(10). Hp(10) (the dose equivalent in soft tissue measured at a depth of 10 mm) is an accepted surrogate measurement of effective dose (E). This is inaccurate in this setting, as the operator wears a lead apron so this value was divided by 21 – in line with the methodology outlined in the NCRP 168 document15 to give an effective dose (E) whilst wearing a lead apron. Operators wearing the IDD were all cardiologists with varying experience. For analysis, they were grouped into three experience categories: Group1 = Registrar, Group 2 = Interventional fellow, Group 3 = Consultant.

Data integrity and statistical analysis

Data for procedures where an IDD was worn were compared to those procedures where the IDD was not worn to ensure a comparable data set. The distributions of variables were assessed. Means and standard deviations (SD) or medians and inter‐quartile ranges were used to describe continuous variables. Mann–Whitney U tests were used to compare continuous variables when there were only two categories and Kruskal–Wallis tests were used to compare continuous variables when there were more than two categories. The continuous outcome measures (OD and PKA) were categorised to form binary measures for use in analysis. For each variable, values below the 75th percentile were included in the reference category and values at or above the 75th percentile were included in the high dose comparison category. Categorical variables were compared using a Chi squared test. Significant categorical variables from the Chi‐square analysis with a P < 0.1 with univariate analysis were included in multivariate logistic regression modelling. Stepwise removal of variables with the highest P‐value was performed. In the final model all variables remained significantly associated with the outcome at the 0.05 level.

Results

The study population comprised 3860 patients. The mean age for patients was 66 (SD 13) years. PKA measurements were available for all 3860 procedures and the IDD was worn during 2591 (67.1%) of these procedures. Primary operator dose (OD) was measured in this subset. The cut point for the 75th percentile value for PKA was 61.4 Gycm2 (Table 1). Procedures above this were considered High PKA. The cut point for the 75th percentile for Operator 1 dose (OD) was 1.90 μSv, with dosimeter readings above this value categorised as high.
Table 1

Overall patient and radiation data and comparisons across procedure categories

Procedure category
OverallAFACORSCORS‐LV/AOCORS‐LHC/RHCCORS‐OCT/IVUS/FFRCORS‐PACORS‐GRAFTSPARHCTAVI ‐WU P‐value
N 36801110191745200268104131855121
Patient measures
Age (years)66 ± 1377 (72–82)67 (57–75)65 (56–72)69 (54–77)68 (61–75)64 (59–67)72 (66–79)52 (36–67)59 (41–72)82 (78–86)<0.001
BMI kg/m2 29.3 (25.5–33.6)30.5 (25.7–34.8)29.3 (25.4–33.8)29.6 (25.7–33.9)27.9 (24.1–33.3)29.4 (26.0–33.2)29.8 (28.3–35.3)28.9 (25.9–32.9)25.6 (22.2–30.0)27.6 (21.2–31.6)27.7 (24.9–31.4)<0.001
Radiation measures
FT (min)4.7 (2.9–8.3)4.5 (1.3–7.7)3.9 (2.5–6.4)3.7 (2.6–5.6)8.1 (5.1–11.9)9.1 (6.5–12.9)13.0 (10.4–17.9)9.8 (6.7–14.4)10.5 (8.5–14.3)4.7 (2.0–9.5)7.5 (4.5–9.7)<0.001
KAR (mGy)646 (399–980)139.0 (32.6–451)626.0 (398–934)593.0 (374–872)613.0 (372–900.5)841.5 (493.5–1242.5)1327.0 (885–1796)1018.0 (665–1476)391.5 (200–610)53.8 (19–105)721.0 (449.0–1124)<0.001
PKA (Gycm2)40.0 (24.8–61.4)12.93 (3.13–42.70)37.37 (23.96–56.41)35.76 (22.89–53.49)43.25 (26.69–65.28)49.37 (30.44–72.72)99.69 (67.66–160.07)67.11 (43.75–95.58)34.30 (17.13–57.43)5.72 (2.24–13.97)47.86 (32.09–78.38)<0.001
DA10 (9–13)4 (2–9)9 (8–11)10 (10–12)10 (8–12)12 (9–14)19 (17–22)17 (14–20)8 (5–9)0 (0–1)14 (12–16)<0.001

Table shows how the different patient characteristics and radiation measures compare across procedure category. Median and inter‐quartile ranges are shown with P‐values calculated using Kruskal–Wallis tests. BMI, body mass index; CORS, coronary angiogram; DA, digital acquisitions; FT, fluoroscopy time; KAR, air Kerma at the interventional reference point; PKA, Kerma area product; TAVI, transcatheter aortic valve implant; AFA, abdominal/femoral angiogram only; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; PA, pulmonary angiography alone; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant.

Overall patient and radiation data and comparisons across procedure categories Table shows how the different patient characteristics and radiation measures compare across procedure category. Median and inter‐quartile ranges are shown with P‐values calculated using Kruskal–Wallis tests. BMI, body mass index; CORS, coronary angiogram; DA, digital acquisitions; FT, fluoroscopy time; KAR, air Kerma at the interventional reference point; PKA, Kerma area product; TAVI, transcatheter aortic valve implant; AFA, abdominal/femoral angiogram only; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; PA, pulmonary angiography alone; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant. All patient and radiation parameters did not differ significantly between procedures where the IDD was and was not worn. The overall and procedure group comparisons for each of the 10 examination categories are given in Table 1. The CORS‐LV/AO category was the most numerous, accounting for 1745 (45%) of the procedures performed. The medians of all variables differed significantly across procedural categories. Patients in the TAVI‐WU category were the eldest (median 82 years) but the CORS‐PA category demonstrated the highest FT (13 (10.4–17.9) min), KAR (1327 (885–1796) mGy), PKA (99.69 (67.66–160.07) Gycm2) and DA (19 (17–22)). However, this category accounted for only 0.3% of procedures performed. Median OD was 0.95 μSv (Inter quartile range (IQR) 0.00–1.90) and varied significantly across procedure categories (P = 0.002) and is demonstrated graphically in Figure 1. Median patient and operator dose for each category within each variable that may impact on radiation dose is given in Table 2.
Figure 1

Median radiation dose to operators by procedure category. This graph demonstrates the differences in operator dose across procedure categories. AFA, abdominal/femoral angiogram only; BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; PA, pulmonary angiography alone; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography.

Table 2

Median radiation dose for patients and operators within variable categories

ItemPKA (Gycm2) Median (IQR) P‐valueOD (µSv) Median (IQR) P‐value
N 3860<0.00125910.002
Access
 Radial access38.37 (24.55–57.24)0.95 (0.00–1.90)
 Femoral access46.04 (26.17– 72.57)0.95 (0.00–1.90)
 Jugular access2.86 (1.43–6.06)0.72 (0.24–1.19)
Gender<0.001<0.001
 Female27.52 (16.65–44.57)0.95 (0.00–1.43)
 Male 47.45 (32.22–70.71)0.95 (0.00–2.38)
Imaging technique<0.001<0.001
 Biplane Imaging58.16 (35.55–89.74)1.43 (0.00–2.38)
 Single plane36.44 (22.88–55.27)0.95 (0.00–1.90)
Cath lab used<0.0010.372
 Cath lab 1 37.40 (23.29–59.83)0.95 (0.00–1.90)
 Cath lab 242.49 (26.40–63.97)0.95 (0.00–1.90)
BMI category<0.001<0.001
 BMI >3053.63 (37.52–76.01)1.43 (0.00–2.38)
 BMI 25–3035.96 (24.29–53.02)0.95 (0.00–1.90)
 BMI <2521.73 (13.43–33.83)0.48 (0.00–1.43)
Operator 1 experience0.0010.380
 OP1 exp. (Consultant)38.06 (23.91–60.40)0.95 (0.00–1.90)
 OP1 exp. (Fellow)41.11 (25.07–61.37)0.95 (0.00–1.90)
 OP1 exp. (Registrar)44.60 (28.38–65.41)0.95 (0.00–1.90)
Operator 2 experience0.8250.056
 OP2 exp. (Consultant)46.74 (30.65–71.07)0.95 (0.00–2.38)
 OP2 exp. (Fellow)45.68 (26.77–67.13)0.95 (0.00–1.90)
 OP2 exp. (Registrar)44.76 (29.50–69.86)1.43 (0.48–2.86)

This table demonstrates the differences in median (inter‐quartile range (IQR)) radiation dose values for the patient (Gycm2) and operator (uSv) between categories within each variable measured. BMI, body mass index; OD, primary operator dose; OP1, operator 1; OP2, operator 2.

Median radiation dose to operators by procedure category. This graph demonstrates the differences in operator dose across procedure categories. AFA, abdominal/femoral angiogram only; BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; PA, pulmonary angiography alone; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography. Median radiation dose for patients and operators within variable categories This table demonstrates the differences in median (inter‐quartile range (IQR)) radiation dose values for the patient (Gycm2) and operator (uSv) between categories within each variable measured. BMI, body mass index; OD, primary operator dose; OP1, operator 1; OP2, operator 2.

Factors contributing to a high PKA

Overall, 64% of patients were male, 33% of patients were overweight (BMI 25–30) and 45% were obese (BMI >30). Under univariate analysis, the CORS‐PA procedure category demonstrated the highest odds ratio (OR), with these patients having a 16 fold greater chance of receiving a high dose in comparison to the CORS baseline category. Under multivariate analysis, a BMI > 30 was the strongest predictor for a high PKA, demonstrating that after correcting for all other variables in the model, these patients were 19.1 times more likely to receive a high dose. Males were more likely to be in the highest quartile of radiation dose (OR: 5.3, 95% CI: 4.3–6.7), compared to females and Biplane imaging was also more likely to be associated with a high PKA (OR = 5.2 (95% CI: 4.2–6.5)) (Table 3).
Table 3

Regression analysis for variables that impact on PKA for diagnostic cardiac angiography procedures

UnivariateMultivariate
OR95% CIP valueOR95% CIP value
Gender<0.001<0.001
 FemaleRef.Cat.Ref.Cat.
 Male 4.393.61–5.32<0.0015.344.26–6.68<0.001
Access<0.001
 FemoralRef.Cat.
 Jugular0.130.02–1.000.05
 Radial0.530.46–0.62<0.001
Cath lab used<0.001<0.001
 Cath lab 1 usedRef.Cat.Ref.Cat.
 Cath lab 2 used1.201.04–1.390.011.371.14–1.640.002
Imaging technique<0.001<0.001
 Single planeRef.Cat.Ref.Cat
 Bi‐plane3.933.34–4.62<0.0015.174.14–6.47<0.001
BMI<0.001<0.001
 BMI <25Ref.Cat.Ref.Cat
 BMI 25‐303.582.61–4.90<0.0013.882.74–5.50<0.001
 BMI >309.597.13–12.90<0.00119.0813.52–26.94<0.001
Procedure type<0.001<0.001
 CORSRef.Cat.Ref.Cat
 AFA0.400.05–3.120.3860.250.03–2.260.215
 CORS‐LV/AO0.850.70–1.030.1020.880.70–1.100.254
 CORS‐OCT/IVUS/FFR2.551.91–3.41<0.0013.472.46–4.89<0.001
 CORS‐PA16.083.39–76.29<0.00110.361.96–54.770.006
 CORS‐RHC1.641.17–2.310.0042.511.66–3.79<0.001
 CORS‐GRAFTS5.154.02–6.60<0.0013.662.70–4.96<0.001
 PA1.150.37–3.530.8091.510.40–5.670.542
 RHC0.320.11–0.880.0280.450.15–1.370.157
 TAVI‐WU2.551.72–3.80<0.0011.851.13–3.010.014
Operator 1 Experience0.0770.010
 ConsultantRef.Cat.Ref.Cat.
 Fellow1.060.90–1.240.4801.060.87–1.280.588
 Registrar1.291.04–1.610.0241.511.15–1.980.003
Operator 2 Experience0.72
 ConsultantRef.Cat.
 Fellow0.970.55–1.730.923
 Registrar0.810.46–1.420.455

This table demonstrates the odds ratio and significance for all variables that impact on a high PKA in the study under univariate and multivariate logistic regression. AFA, abdominal/femoral angiogram only; BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; OR, odds ratio; PA, pulmonary angiography alone; Ref.Cat., reference category; RHC, Right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography.

Regression analysis for variables that impact on PKA for diagnostic cardiac angiography procedures This table demonstrates the odds ratio and significance for all variables that impact on a high PKA in the study under univariate and multivariate logistic regression. AFA, abdominal/femoral angiogram only; BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; OR, odds ratio; PA, pulmonary angiography alone; Ref.Cat., reference category; RHC, Right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography.

Factors contributing to a high operator 1 dose (OD)

In the subset of 2591 patients where the IDD was worn by operators, patients were predominantly male (64%), 33% were overweight and 46% were obese. Biplane imaging was used in 21% of cases and radial access was used in 69% of procedures. Under univariate analysis, a BMI > 30 was demonstrated to be the strongest predictor of a high OD, with an OR of 3.3 (95% CI: 2.5–4.4). This effect persisted in the multivariate model, where after correcting for all other variables in the model, a BMI > 30 was the strongest predictor of a high OD, (OR = 3.3 (95% CI: 2.4–4.4)). Of the 10 procedural categories, TAVI‐WU demonstrated the highest OR for predicting a high OD, (OR = 2.7 (95% CI: 1.5–4.8)). Males, biplane imaging and radial access were also associated with a high OD (Table 4).
Table 4

Regression analysis for variables that impact on operator dose for diagnostic cardiac angiography procedures

UnivariateMultivariate
OR95% CIP valueOR95% CIP value
Gender<0.001<0.001
 FemaleRef.Cat.Ref.Cat.
 Male 1.771.44–2.17<0.0011.661.33–2.06<0.001
Access0.053<0.001
 FemoralRef.Cat.Ref.Cat.
 Jugular0.830.18–3.840.8151.790.26–12.160.565
 Radial1.291.05–1.580.0172.341.69–3.17<0.001
Cath lab used0.793
 Cath lab 1 usedRef.Cat
 Cath lab 2 used0.980.81–1.170.793
Imaging technique<0.001<0.001
 Single planeRef.Cat.Ref.Cat.
 Bi‐plane1.731.40–2.14<0.0012.181.68–2.84<0.001
BMI<0.001<0.001
 BMI <25Ref.Cat.Ref.Cat.
 BMI 25–301.661.22–2.270.0011.511.09–2.080.011
 BMI >303.312.49–4.41<0.0013.252.42–4.37<0.001
Procedure type0.0110.014
 CORSRef.Cat.Ref.Cat.
 AFA1.940.35–10.680.4483.130.50–19.570.223
 CORS‐LV/AO1.030.82–1.300.7930.970.77–1.240.818
 CORS‐OCT/IVUS/FFR1.410.97–2.050.0691.440.98–2.120.067
 CORS‐PA1.940.48–7.840.3541.870.43–8.030.404
 CORS‐RHC0.830.50–1.360.4541.280.74–2.200.376
 CORS‐GRAFTS1.531.11–2.120.0101.861.24–2.790.003
 PA0.860.18–4.030.8491.490.28–7.880.641
 RHC0.520.18–1.490.2220.910.24–3.360.882
 TAVI‐WU2.051.24–3.390.0052.681.50–4.760.001
Operator 1 Experience0.094 
 ConsultantRef.Cat 
 Fellow1.230.92–1.650.168
 Registrar1.000.74–1.341.000
Operator 2 Experience0.190
 ConsultantRef.Cat
 Fellow1.350.76–2.400.310
 Registrar0.760.41–1.420.391

This table demonstrates the odds ratio and significance for all variables that impact on a high OD in the study under univariate and multivariate logistic regression. AFA, abdominal/femoral angiogram only, BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; OR, odds ratio; PA, pulmonary angiography alone; Ref.Cat., reference category; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography.

Regression analysis for variables that impact on operator dose for diagnostic cardiac angiography procedures This table demonstrates the odds ratio and significance for all variables that impact on a high OD in the study under univariate and multivariate logistic regression. AFA, abdominal/femoral angiogram only, BMI, body mass index; CORS, coronary angiography only; CORS‐LV/AO, coronary angiography + aortogram/left ventriculogram; CORS‐LHC/RHC, coronary angiography + left heart catheterisation + right heart catheterisation; CORS‐OCT/IVUS/FFR, coronary angiography + additional coronary artery lesion assessment with intravascular ultrasound (IVUS), optical coherence tomography (OCT) of fractional flow reserve (FFR); CORS‐PA, coronary angiography + pulmonary angiography; CORS‐GRAFTS, coronary angiography + coronary artery bypass graft angiography; OP1, operator 1; OP2, operator 2; OR, odds ratio; PA, pulmonary angiography alone; Ref.Cat., reference category; RHC, right heart catheterisation ± venography; TAVI‐WU, work‐up for transcatheter aortic valve implant: Includes coronary angiography, aortography, ilio‐femoral angiography.

Discussion

This study is a large and comprehensive analysis of factors that can impact on patient and operator radiation dose during 10 different diagnostic cardiac procedures performed in a tertiary cardiac catheter laboratory over 16 months period. There was no significant difference between the radiation dose variables when an IDD was or was not worn and this is important when establishing results for operator dose and whether they are relevant across the entire cohort. A local, multicentre diagnostic reference level (DRL) for CA was developed in 2013 and published in 2014.16 The 75th percentile of 61.4 Gycm2 in this study appears to be similar to that studies’ DRL for CA of 58.65 Gycm2 and importantly, the median PKA value appears to be lower than that studies’ 75th percentile DRL.16 This demonstrates ongoing compliance with the local DRL, even with the addition of more complex procedures that were not included in the DRL study of 2014.

Obesity

The multivariate analysis demonstrates that after correcting for all other variables in the model, risk of exposure to high PKA and OD values was highest in obese patients. This is consistent with previous research where obesity was demonstrated to significantly increase the radiation dose used during coronary angiography.17 However, the present study supplements these findings with the additional information that a high patient BMI is the strongest predictor for both a high patient and operator dose across a range of diagnostic procedures. A high patient BMI means a greater patient thickness which requires more X‐ray tube output. The X‐ray system boosts exposure automatically but increases patient dose from the increased absorption and scatter of X‐rays from the additional thickness. The additional scatter in turn increases OD.

Catheter access route

Radial arterial access has previously been demonstrated to be a procedural factor that increases PKA 18, 19, 20 and operator dose.21, 22 However, in the present study, under multivariate analysis, access route did not remain significantly associated with a high PKA at the 5% level. However, radial artery access was associated with a high OD. This is important as radial artery access is now the access route of choice, with lower vascular complication rates23, 24 and a higher patient preference.24 This study shows a predominantly radial approach and operators are familiar with the technique. The reason for a higher OD with radial access is likely due to the operator standing closer to the radiation source and this observation has previously been highlighted as a possible reason for a higher OD.12, 18

Biplane imaging

Unsurprisingly, biplane imaging was demonstrated to significantly increase high PKA and OD rates. Biplane imaging uses an additional C‐arm and X‐ray source to acquire images and it has previously been demonstrated to increase FT, DA25 and PKA.26 The study by Lin et al.26 also demonstrated that imaging of saphenous vein grafts and patient gender impact on PKA, a finding that is has been replicated in this study, where the CORS‐GRAFTS procedure category was demonstrated to be associated with a high PKA and OD under univariate and multivariate analysis. Biplane imaging takes more images and may increase magnification, as it requires the anatomy to be placed in the isocentre of the image in both planes, positioning the patient closer to the X‐ray source, which will increase skin dose. In addition, biplane imaging may increase the air‐gap between the patient and the detector, increasing the incident air Kerma. These factors will also impact on OD and the ceiling suspended lead shield may be more difficult to position in order to protect the operator from the additional radiation from the two C‐arms.

X‐ray system used

It is interesting that one X‐ray system delivered more PKA than the other and that this was a significant predictor under multivariate analysis for a high PKA. This did not, however, equate to a high OD. Both systems were identical models of the same age and used the same imaging protocols and exposure parameters. One explanation for this finding may be a difference in dose area product meter (DAP meter) readings between the two systems. Median PKA varied only slightly between the systems. Both systems were tested on a regular basis for dose outputs and the accuracy of the DAP meter, and there are acceptable error ranges for DAP meters. Others studies suggest that an error of ±25% is usually deemed acceptable,27, 28 which is far greater than the difference seen here.

Procedural categories

Coronary angiography (CORS) was used as the baseline/reference procedure category for this study and the other categories were compared against this. CORS is seen as the baseline procedure of cardiac angiography, with most other procedure categories building on it. CORS and CORS‐AO/LV account for 71.6% of procedures in this study. Only AFA, PA and RHC procedures did not include coronary angiography. It is unsurprising therefore that those procedure categories that is built on the CORS reference procedure, demonstrated a higher PKA. Of the procedure categories, CORS with pulmonary angiography (CORS‐PA) was the strongest procedural predictor for a high PKA. CORS‐PA had the highest median PKA and DA count. CORS‐PA routinely uses bi‐plane angiography, in which the planes are used in the posterior‐anterior projection and lateral projection for the PA component. The lateral projection is very steep and steep or more extreme angles have been shown to increase PKA.29 Intra coronary assessment of coronary artery disease with the use of pressure wires (FFR) and intra‐vascular imaging with OCT or IVUS is also shown to impact on PKA, although it did not reach significance for a high OD. The results here show that these additional assessment tools increase the FT, PKA and DA for diagnostic procedures. FFR is recommended in the assessment of intermediate lesions prior to revascularisation30 and the use of these additional tools may increase over time so it is important to measure their impact on radiation dose.

Operator experience

Operator experience is shown to impact on PKA during CA under multivariate analysis. After correcting for other variables, the least experienced operators (registrars) were statistically more likely to deliver high PKA values and these findings complement those of the REVERE trial where less experienced radial access operators delivered a higher PKA.12 The higher PKA seen is possibly due to increased fluoroscopy times when manipulating guide wires and catheters, and possibly a higher number of repeat DA from catheters disengaging mid‐acquisition. A higher number of DA and FT for inexperienced operators was also noted in the REVERE trial.12 Other studies have demonstrated a difference between individual operators in terms of FT, KAR and PKA delivered, though this was not attributed to experience level.20, 31 Operator experience did not appear to impact on OD in this study but all operators had training in the use of radiation and the effective use of protection measures, supplemented through annual refresher training.

Limitations

While not measured in this study, the effective use of the ceiling suspended shield would make a significant impact on OD and this should be analysed in future studies. Beam kV, mAs and beam geometry were not collected for each procedure, given the cohort size and this could have impacted on dose for different procedures.

Conclusions

This study demonstrates that radiation dose to patients and the operator performing the procedure is multifactorial and is affected by both patient and procedure related variables. Some variables impact on radiation dose more than others and some variables are predictive of a high PKA, some predictive of a high OD, and some both. Overall, a high patient BMI is the strongest predictor for both a high patient and operator radiation dose.

Conflict of Interest

The authors declare no conflict of interest.
  6 in total

1.  The use of digital magnification to reduce radiation dose in the cardiac catheter laboratory.

Authors:  Shailesh Dalvi; Hywel Mortimer Roberts; Christopher Bellamy; Michael Rees
Journal:  Br J Radiol       Date:  2022-01-01       Impact factor: 3.039

2.  Combined use of electro-anatomic mapping system and intracardiac echocardiography to achieve zero-fluoroscopy catheter ablation for treatment of paroxysmal atrial fibrillation: a single centre experience.

Authors:  Matevž Jan; David Žižek; Dimitrij Kuhelj; Nikola Lakič; Tine Prolič Kalinšek; Jernej Štublar; Luka Klemen; Andrej Pernat; Bor Antolič
Journal:  Int J Cardiovasc Imaging       Date:  2019-11-11       Impact factor: 2.357

3.  Comparison of radiation dose and its correlates between coronary computed tomography angiography and invasive coronary angiography in Northeastern Thailand.

Authors:  Phatraporn Aupongkaroon; Pattarapong Makarawate; Narumol Chaosuwannakit
Journal:  Egypt Heart J       Date:  2022-01-25

4.  Comparison of radiation exposure and clinical outcomes between transradial and transfemoral diagnostic cerebral approaches: a retrospective study.

Authors:  Curtis Amankwah; Lauren Lombardo; John Rutledge; Ahsan Sattar; Bree Chancellor; Dorothea Altschul
Journal:  BMJ Surg Interv Health Technol       Date:  2022-01-21

5.  Occupational dose and associated factors during transarterial chemoembolization of hepatocellular carcinoma using real-time dosimetry: A simple way to reduce radiation exposure.

Authors:  Hyoung Ook Kim; Byung Chan Lee; Chan Park; Jae Kyu Kim; Won-Ju Park; Jong Eun Lee; Hyo Soon Lim; Won Gi Jeong
Journal:  Medicine (Baltimore)       Date:  2022-01-28       Impact factor: 1.889

6.  Effect of Different Anthropometric Body Indexes on Radiation Exposure in Patients Undergoing Cardiac Catheterisation and Percutaneous Coronary Intervention.

Authors:  Youlin Koh; Sara Vogrin; Samer Noaman; Simon Lam; Raymond Pham; Andrew Clark; Leah Biffin; Laura B Hanson; Jason E Bloom; Dion Stub; Angela L Brennan; Christopher Reid; Diem T Dinh; Jeffrey Lefkovits; Nicholas Cox; William Chan
Journal:  Tomography       Date:  2022-09-11
  6 in total