Changes in thromboinflammatory profiles across the generations of transcatheter aortic heart valves.

Travis Sexton1, Maria Alkhasova1, Marcielle de Beer2, Donald Lynch3, Susan Smyth4.   

Abstract

The transcatheter aortic valve replacement (TAVR) procedure was developed to provide patients with severe aortic stenosis an alternative to the surgical aortic valve replacement. Since the approval of the original SAPIEN the technology has rapidly evolved. While several approaches can be used for valve deployment, as delivery systems have become smaller and more flexible, the transfemoral approach has become the dominant technique for valve deployment. One hundred and forty five patients undergoing TAVR receiving one of four valve types (Sapien, Sapien XT, Sapien3 or CoreValve) via the femoral artery were included in this study. Platelet count, white blood cells count (WBC), Interleukin-6 (IL-6), and Serum Amyloid A (SAA) were determined before and after TAVR. Platelet counts declined after the procedure regardless of the valve type and were dependent upon the baseline platelet count. Use of conscious sedation blunted the decline in platelet count. With the newer generation valves, the rise in WBC post-TAVR was lower than observed with the Sapien, in keeping with less systemic inflammation. Consistent with WBC, IL-6 levels were lower following deployment of the newer generation valves. Elevations in plasma SAA, which occur following myocardial injury, were not reduced with the newer valves. Evolution of the TAVR technology has occurred rapidly over the last 5 years. The newer devices and smaller delivery systems are associated with less systemic inflammation, as reflected in WBC and plasma IL-6 levels. However, the acute phase reactant SAA remains unchanged, possibly reflecting different triggers for SAA following TAVR.

Entities:  

Keywords:  Inflammation; Platelet; Transcatheter aortic valve replacement

Mesh:

Substances:

Year:  2019        PMID: 30484011      PMCID: PMC6393181          DOI: 10.1007/s11239-018-1782-3

Source DB:  PubMed          Journal:  J Thromb Thrombolysis        ISSN: 0929-5305            Impact factor:   2.300


  4 in total

Review 1.  Transcatheter aortic valve thrombosis: a review of potential mechanisms.

Authors:  Vrishank Raghav; Prem Midha; Rahul Sharma; Vasilis Babaliaros; Ajit Yoganathan
Journal:  J R Soc Interface       Date:  2021-11-24       Impact factor: 4.118

Review 2.  Lipoprotein(a) as Orchestrator of Calcific Aortic Valve Stenosis.

Authors:  Johan G Schnitzler; Lubna Ali; Anouk G Groenen; Yannick Kaiser; Jeffrey Kroon
Journal:  Biomolecules       Date:  2019-11-21

3.  Transcatheter aortic valve replacement during the COVID-19 pandemic-A Dutch single-center analysis.

Authors:  Maxim J P Rooijakkers; Wilson W L Li; Laurens W L M Wollersheim; Guillaume S C Geuzebroek; Helmut Gehlmann; Leen A F M van Garsse; Marleen H van Wely; Michel W A Verkroost; Wim J Morshuis; Heiman Wertheim; Niels van Royen
Journal:  J Card Surg       Date:  2020-10-21       Impact factor: 1.620

4.  Thrombocytopenia Complicating Transcatheter Aortic Valve Implantation: Differences Between Two New-Generation Devices.

Authors:  Nicola Corcione; Simona Romano; Alberto Morello; Paolo Ferraro; Michele Cimmino; Michele Albanese; Martina Tufano; Daniela Capasso; Salvatore Buonpane; Salvatore Giordano; Martino Pepe; Giuseppe Biondi-Zoccai; Maria Fiammetta Romano; Arturo Giordano
Journal:  J Cardiovasc Transl Res       Date:  2021-03-15       Impact factor: 4.132

  4 in total