Literature DB >> 30481301

Ventricular substrate identification using close-coupled paced electrogram feature analysis.

Mohammad Hassan Shariat1, Divyanshu Gupta1, Enes E Gul1, Benedict Glover1, Javad Hashemi1, Hoshiar Abdollah1, Adrian Baranchuk1, Christopher Simpson1, Kevin A Michael1, Damian P Redfearn1.   

Abstract

AIMS: Substrate based catheter ablation strategies are widely employed for treatment of scar-related ventricular tachycardia (VT). We analysed intracardiac electrograms (EGMs) from close-coupled paced extrastimuli extracted from the EnSite Precision mapping system. We sought to characterize EGM responses of ventricular myocardium to varying coupling intervals from the right ventricular apex (RVA) in both healthy individuals and patients presenting with VT for catheter ablation. METHODS AND
RESULTS: Extrastimuli were delivered from the RVA after estimation of the ventricular effective refractory period. Electrograms were recorded from high-density mapping catheters in the left ventricle and exported for analysis to MATLAB. Observational data were collected from 14 patients with ischaemic VT (mean age 72.4 ± 6.3 years, one female) and five controls (mean age 59.4 ± 7.4 years, one female). These derived data were used to inform an interventional strategy on a further 10 patients (mean age 64.7 ± 10.0 years; two female). Significant differences were observed in EGM duration (ED) and latency (LT) at all coupling intervals between VT patients and controls. Significant increases in ED and LT with decreased RVA coupling interval were observed at VT isthmuses. Abnormal responses derived from control subject data were used to classify four types of ventricular EGM response. Targeting sites with abnormal LT and ED significantly reduced VT inducibility (5/14 derivation patients to 0/10 intervention patients; P = 0.03).
CONCLUSION: Paced electrogram feature analysis is a novel tool to characterize the ischaemic substrate. Association with VT isthmuses and early ablation results suggest a possible role in substrate ablation for ischaemic VT. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author(s) 2018. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  Ablation; Close-coupled pacing; Electroanatomic mapping; Electrogram; Latency; Paced electrogram feature analysis; Ventricular tachycardia

Year:  2019        PMID: 30481301     DOI: 10.1093/europace/euy265

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  How to unmask septal local abnormal ventricular activities with the new LUMIPOINTTM software.

Authors:  Francesco Solimene; Francesco Maddaluno; Maurizio Malacrida; Giuseppe Stabile
Journal:  J Arrhythm       Date:  2020-05-01

Review 2.  Ventricular Tachycardia Ablation Guided by Functional Substrate Mapping: Practices and Outcomes.

Authors:  Sara Vázquez-Calvo; Ivo Roca-Luque; Andreu Porta-Sánchez
Journal:  J Cardiovasc Dev Dis       Date:  2022-08-30